background image

TPV-53 power meter  User Manual

Products introduction

Power on/Auto power off

Optical power reference value setting

Optical power wavelength switching

VFL function

After short press the power button, 

the meter will turn on and automatically 

start the automatic shutdown function. 

The default automatic shutdown time is 

10 minutes. To cancel this function, press

and hold the power button when the power 

is turned on. When the power button dis-

appears, the automatic shutdown will be 

canceled.

Insert the fiber under test into the 

VFL interface, press the VFL key to enter 

the VFL test interface.

270

HZ

270

HZ

  TPV-53 is a combination of optical power meter and VFL.  Among 

them, the optical power module can be used to measure the optical 

power in mw, dB, dBm within the wavelength range of 850~1700nm, 

with high display resolution. There are seven wavelengths calibration 

points at 850nm, 980nm, 1300nm, 1310nm, 1490nm, 1550nm and 

1625nm. It can display both linear and non-linear optical power at the 

same time, which can be used for direct measurement of optical power 

and relative measurement of optical link loss. The VFL module provides 

650nm visible laser output, the test distance can reach 10km, and the 

output power can reach 1mW/10mW/20mW. It is suitable for 

single-mode or multi-mode fiber measurement and can identify 

optical jumpers, jumper boards, distribution frames, and connectors 

fault at the place.The instrument has a continuous backlight display 

function, a large LCD display, making the measurement work more 

conveniently and quickly,it is the construction of cable television 

systems, optical fiber communications and other fields Indispensable 

equipment for maintenance.

OPM interface

VFL interface

Display

Wavelength 

switch key

VFL key: start 

red light 

detection function

REF/dB key: Short 

press the dB unit 

to switch

On/off key

Wavelength

 Automatic 

shutdown

Frequency

Battery power

Unit

Numerical value

270

P absolute power = 10lgP linear power / 1mW

P relative power = P absolute power-

P reference power

Insert the fiber which is under test into the 

OPM interface.After booting, enter the 

optical power meter interface, short press 

the "REF" key to set the current power value 

as the reference power, which can achieve 

relative optical power test (insertion loss test) 

or absolute power test, enter the relative 

power test mode at the bottom of the screen 

Display insertion loss (dB).The screen displays 

the reference value; short press the "dB" key 

to switch between linear power and absolute 

power display.The units of linear power, 

absolute power, and relative power are nW, 

dBm, and dB respectively.

REF/dB

On the optical power meter interface, short 

press the "λ" key to switch the measurement 

wavelength, select 7 different test 

wavelengths, 850nm/980nm/1300nm/

1310nm/1490nm/1550nm, 850nm, 1625nm 

to ensure the accuracy of the test, select 

the wavelength should be consistentwith the 

wavelength of the measured optical signal.

λ

850

nm

980

nm

1300

nm

1310

nm

1490

nm

1550

nm

1625

nm

270

Отзывы: