background image

 

Page 1 of 9 

 

INSTRUCTIONS FOR USE  

 

 
 

THE AIO SOLUTION

®

 

ACCESSORIES  
STORAGE CABINET 
 
 

 
 

 

 
 
AIO SOLUTION® code: 

AC-SC

 

Article No. : 

 29100

 

A.

 

GENERAL PRODUCT INFORMATION 

The product referred to in these instructions is a 
storage system for the AIO Solution

®

, a medical 

device used for patient positioning and 
immobilisation in radiation therapy. 

B.

 

PRODUCT DESCRIPTION 

This AIO cabinet is a storage system for "THE 
TOTAL AIO SOLUTION

®

 

PACKAGE”. It features 

retractable panels with dedicated spaces for all the 
items contained in the AIO Solution

® 

package. 

C.

 

PRECAUTIONS FOR USE 

Always use the correct tools for screwing, gluing 
and cutting. Wear working gloves when handling 
the fibre wood plates to avoid injuries. Wear eye 
protection and working gloves whenever using 
power tools. 
 
The cabinet will tip over if all five panels are slid 
out at the same time! Therefore it is highly 
recommended to fix the cabinet to the wall 
(fixation elements included in the package). 
 
Do not place the AIO HPL base plate in the cabinet 
as the hooks cannot carry the weight! 

D.

 

WEIGHTS AND DIMENSIONS 

D.1. AIO CABINET 

The AIO cabinet is made of fibre wood panels with 
melamine white finish. 

 

 

Overall dimensions: 

Width 1600 mm (63”)

 

Height 1980 mm (78”)

 

Depth 690 mm (27.2”)

 

Weight: 200 kg 

D.2. CASE 

The  case  is  made  of  fibre  wood  panels  which  are 
screwed together. 

Overall dimensions: 

Length 1950 mm (76.8”)

 

Width 1000 mm (39.4”)

 

Height 780 mm (30.7”)

 

Weight of packaging: 91 kg 

Total weight c case: 291 kg  

E.

 

ASSEMBLY INSTRUCTIONS 

The AIO cabinet must be assembled by two 
persons. 
 
Place the assembled AIO cabinet with the back 

against the wall. Keep ± 20 cm (± 7,9”) of free 
space on both sides and ± 120 cm (± 47,2”) of free 

space at the front to slide out the panels. 
 
The AIO cabinet requires at least an area of 1605 x 
695 mm 

(63.2” x 27.4”).

 

Minimum height = 1985 mm (78.1”).

 

F.

 

REQUIRED TOOLS 

-

 

Power tool with crosshead bit PZ2 (Pozidrive) 

-

 

Rubber hammer 

-

 

Gloves, protective glasses 

Содержание AIO SOLUTION 29100

Страница 1: ...s highly recommended to fix the cabinet to the wall fixation elements included in the package Do not place the AIO HPL base plate in the cabinet as the hooks cannot carry the weight D WEIGHTS AND DIME...

Страница 2: ...x Hook length 135 mm Art no 29100 57 4x Hook length 145 mm Art no 29100 58 4x Hook length 155 mm Art no 29100 59 4x Hook length 185 mm Art no 29100 60 4x Hook length 225 mm Art no 29100 61 List of sup...

Страница 3: ...25 mm 1 block Screw 35 mm 2 blocks Screw 45 mm 3 blocks Screw 50 mm 35 mm 45 mm 55 mm 65 mm 85 mm 95 mm 105 mm 115 mm 135 mm 145 mm 155 mm 185 mm 225 mm Screw length Hook length The 13 different hook...

Страница 4: ...ly Place the screws for both side panels Orfit logo on top in the foreseen metal inserts and tighten with a screwdriver Mount six screws on each panel 4 On the bottom panel make sure the arrows on the...

Страница 5: ...Page 5 of 9...

Страница 6: ...struction together with the following pictures to mount the correct elements The hook lengths are written on the included bags Mount the hooks with or without blocks depending on the needed supporting...

Страница 7: ...Page 7 of 9 19 20 21 22 23 Panel 1A Panel 3B Panel 4A Panel 4B Panel 5A Panel 5B...

Страница 8: ...Page 8 of 9 25 Panel 1B 26 Panel 2A 27 Panel 2B 28 Panel 3A 29 Panel 3B 30 Panel 4A...

Страница 9: ...fluid do not use Never use aerosol sprays corrosive cleaning agents solvents or abrasive detergents Periodic checks of the product should be done to insure the parts are not worn and require repair o...