28
Chapter 5
Electricity
* Recommended current specifications for electrical outlet.
Grounding
The DP-100SL ground should be connected to a special ground line, which
is connected to the building ground.
A fixed installation to the ground
has to be provided before switching on the system.
Ground
connection impedance must be <0.5 Ohms.
Voltage Transients
show allowable voltage transients. Those should be
tested when all machinery and air conditioning units are operating
normally and the system is turned on.
Table 5.2 Power Requirements for the Pump
Pump, 3
Φ
+ ground
400 V
(Europe)
208 V
(USA)
200 V
(Japan)
Voltage
400 V+6% /-10%
208 V
±
10%
200 V
±
10%
Frequency 50
±
1 Hz
60 Hz
50/60 Hz
Current*
3 x 16 A
3 x 16 A
3 x 16 A
Maximum power consumption
2.5 kVA
2.5 kVA
2.5 kVA
Note
An additional ground connection must be installed in parallel to the machine’s
power connecting cable (the machine comes with an additional 10m cable, as
mentioned). The earth-connection must be done at same point as for the
earthing wire of power.