LTDVE8CH-20 – INSTRUCTION MANUAL
68
is disabled. When set to
Continuous
the light output is always active
Current range
selects the current range for the light output. It can be chosen between
Automatic
range selection (default),
Low
range (from zero up to 200 mA),
Mid
range (from
zero up to 4 A) and
High
range (from zero up to 20 A)
Current value
is the required output current expressed in A. According with the current range
selected, the output current may range from zero up to 20 A with variable resolution. Upon
entering the required current, the controller will adjust it to the closest achievable value
Ton check
enables or disables (default) the protection against excessive turn-on time
Ton max
is the maximum turn-on time for the connected light, expressed in ms. The
maximum turn-on time may range from 1 ms to 255 ms with a resolution of 1 ms
Toff check
enables or disables (default) the protection against insufficient turn-off time
Toff min
is the minimum turn-off time for the connected light, expressed in ms. The minimum
turn-off time may range from 1 ms to 255 ms with a resolution of 1 ms
Settings
Ton max
and
Toff min
are expressed in ms.
14.3.8. Setup light outputs LD5-LD8
This page allows to inspect and change all the settings related to the light outputs from 5 to 8. Please
refer to the description of page
Setup light outputs LD1-LD4
for more information.
14.3.9. Setup synch outputs SH1-SH4
This page allows to inspect and change all the settings related to the synchronization outputs from 1
to 4. The
Setup synch outputs SH1-SH4