Current value calibration
A. Function description: Change the current value according to the actual working position.
For example: The current value is 100.00mm, the actual working position should be
200.00mm, the steps are as follows:
Step 1. Check that you are in ABS (absolute) mode:
The LCD should display ABS.
Note:
1. Press
to switch between ABS and REL
modes.
2. If pressing this key is not possible, it means that the
key is locked. See "Parameter / key lock and opening".
Step 2. a) Quick parameter setting: Press and hold the
key
, the symbol
is displayed
and press the key
to display the current value
setting.
or
b) keep the button
pressed,
the icon
is displayed in the upper right corner of
the screen, after 5 seconds you will be guided to the
parameter setting, press the button
to display
the current setting.
Step 3. Press
to move between the
digits,
press
to change the value.
After entering the value, press
to complete
the setting and return to the operation screen.
Plus symbol
+ -
Portents selection,