background image

 
 

 

 
 
 

 

 
 
 
 

Optilab EDFA-PA-L-R 

User’s Manual 

 

Erbium Doped Fiber Amplifier 

– Pre-Amp Type 

Rackmount 

 
 
 
 
 

 

CAUTION –  
USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER 
THAN THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION 
EXPOSURE. 

 

THE USER MUST READ THIS MANUAL BEFORE OPERATING THE PRODUCT. 
OPERATIONS OTHER THAN THOSE DESCRIBED IN THIS MANUAL MAY RESULT IN 
PERSONAL INJURY AND DAMAGE TO THE PRODUCT. 

 
 

Note that any attempt to open or fix the equipment without prior approval by Optilab, LLC voids the 
warranty. 

 
 

Ver. 1.3 

Jan 4

th

, 2021

 

Revision History 

Содержание EDFA-PA-L-R

Страница 1: ...ECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION EXPOSURE THE USER MUST READ THIS MANUAL BEFORE OPERATING THE PRODUCT OPERATIONS OTHER THAN THOSE DESCRIBED IN THIS MANUAL MAY RESULT IN PERSONAL INJURY...

Страница 2: ...2021 Updated Logo Copyright 2016 by Optilab LLC All rights reserved This document is copyrighted property of Optilab LLC It may not be used in whole or in part for manufacture sale or design of items...

Страница 3: ...___________________________________________________________ 2 2 2 Initial Inspection___________________________________________________________ 2 2 3 Controls__________________________________________...

Страница 4: ...ility and safety of the device Controllable either through the front panel LCD interface or through USB 2 0 control the EDFA PA L R offers superior integration backed by a three year warranty for long...

Страница 5: ...controls and connectors 2 2 Initial Inspection Your EDFA PA L R was carefully inspected before it left the manufacturer It should be in proper working order upon receipt You should however inspect th...

Страница 6: ...with main power switch enabling disabling electrical power to the rackmount EDFA system USB 2 0 Port This connection is used for remote access of the EDFA system this port is 2 0 USB protocol type Ple...

Страница 7: ...ctor patchcords 3 Connect the optical output port using the indicated connector patchcords to the appropriate signal destination to utilize the amplified optical output signal s 4 After checking all p...

Страница 8: ...rds to the optical input output receptacles cleaning them as necessary 3 Turn the EDFA Enable Key switch back to the ON position normal operation will resume after a few seconds 2 5 PC Connection Mode...

Страница 9: ...ds to the LR module Command Description Example Response READ Reads the equipment basic information READ CR LF Optilab LLC Model Type EDFA PA L R SN XXXXXXX Version V2 0 0 Input Low Output 12 52 dBm R...

Страница 10: ...th power meter and compare with readout on PC connection READ command Return to Optilab for repair if the difference is high 4 dB and cannot be corrected by cleaning or replacing the optical connector...

Страница 11: ...or to fix the equipment by the user without prior approval of Optilab LLC 4 2 Service and Calibration Your EDFA PA L R unit has been designed to provide years of trouble free operation No internal mai...

Страница 12: ...s and then repeat on the dry wipe 5 Visually inspect the end face of the optical connector with an optical microscope to verify cleanliness Repeat steps 2 to 5 as needed Cleaning Optical Connector Sid...