![OPTILAB DFB-4-B Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/optilab/dfb-4-b/dfb-4-b_user-manual_1642556012.webp)
DFB-
B Benchtop User’s Manual
Page 9 of 12
Optilab, LLC
600 E. Camelback Road, Phoenix, AZ 85012
Phone: (602) 343-1496, Fax: (602) 343-1489, Email: [email protected]
3.
Troubleshooting
S
YMPTOM
P
OSSIBLE
C
AUSE AND
S
OLUTION
O
PTICAL OUTPUT POWER
NOT HIGH ENOUGH
.
C:
Optical output connectors dirty.
S:
Disable optical output and clean optical connectors.
C:
Use of incorrect optical adapters or connectors.
S:
Use only the indicated optical adapters and connectors. If
measurement instruments accept different connector type, then
use hybrid patchcords. For shorter wavelengths (<1300nm), use
of special coupling patchcords with matching wavelength will be
necessary to couple the light efficiently.
C:
Optical output connector damaged.
S:
Measure optical output power with power meter and compare
with readout on Main Display. Return to Optilab for repair if the
difference is high (>4 dB) and cannot be corrected by cleaning or
replacing the optical connectors. Always apply dust cover plugs
to unused optical receptacles to prevent the damage of internal
optical connectors.
M
AIN
D
ISPLAY GOES
BLANK
C:
Static electrical discharge.
S:
Wait 10 to 30 seconds for the display to refresh.
C:
Insufficient ventilation.
S:
Place unit in well ventilated area or supply additional fans for
ventilation.
O
PTICAL OUTPUT POWER
UNSTABLE
.
C:
Insufficient optical output isolation.
S:
Connect isolator of corresponding wavelength to optical output
connector. The use of angled patchcords (APC) will help minimize
the light being reflected back into the laser and improve stability.
U
NIT DOES NOT POWER
UP
.
C:
Blown fuse.
S:
Contact Optilab, LLC for fuse replacement procedure.
C:
Insufficient electrical voltage.
S:
Check that the electrical supply is at least 110 VAC.
C:
Power cord is loose.
S:
Plug the power cord firmly into the unit.
U
NIT RESETS OR BLINKS
ON AND OFF
.
C:
Insufficient electrical voltage.
S:
Check that the electrical supply is at least 110 VAC.