CRYO-LINE
Manual for installation and use
OPTIKON 2000
Code 131002 EN
2014-02-06 Rev.F
8-6
8.6
SHUTDOWN PROCEDURE
NOTE
If the Cryo-Line surgical system is not used for an extended period, follow the procedure
described below.
1)
Press the “POWER” switch to turn off the console.
2)
Close the valve on the cylinder or the centralized system outlet.
3)
Detach the power cord and the gas supply hose from the centralized system outlet or
high-pressure gas cylinder and from the rear panel of the console. Grip the plug, not
the cable.
4)
Disconnect the footswitch control from the rear panel of the console.
5)
Disconnect the connectors of the probe and thermocouple.
See the chapter
“Cleaning, sterilization and maintenance”.