CRYO-LINE
Manual for installation and use
OPTIKON 2000
Code 131002 EN
2014-02-06 Rev.F
8-2
4.
To freeze the tip of the probe (Freezing), press the footswitch connected to the
instrument. When the footswitch is pressed, the timer activates to count the time
of application.
5.
To defrost the tip of the probe (Defrosting), release the footswitch connected to
the instrument. The defrosting phase is followed by the pneumatic circuit
draining phase (Venting). In both of these states it is always possible to resume
the freezing phase (Freezing) by pressing the footswitch again.
CAUTION
Do not remove the probe before the end of the pneumatic circuit draining
phase (Venting).
6.
After a few minutes of inactivity, the system performs a new probe cleaning cycle
(Purge) in order to remove the latent humidity inside the pneumatic line of the
cryosurgical probe. To exit from this state, the operator can press the relative
Stop key on the console. To begin the freezing phase of the probe (Freezing), just
press the footswitch connected to the instrument.
8.3
APPLICATION TIME CONTROL
The important parameters are the application time, probe temperature (if equipped
with a thermocouple), pneumatic circuit pressure, activation of the application time
control and adjustment of the application time.
Operation
1.
Select a cryosurgical probe.
2.
Insert the probe connector into the relative connection on the console. If the
probe is equipped with a thermocouple, insert the relative connector into the
connection on the console.