CRYO-LINE
Manual for installation and use
OPTIKON 2000
Code 131002 EN
2014-02-06 Rev.F
7-2
The temperature control occurs when the relative procedure (TEMP CTRL) is activated
by pressing the key next to the symbol on the display.
In this state, pressing the footswitch activates the freezing procedure of the probe until
the set temperature is reached. The temperature is maintained at the set value and
displayed on the display as long as the footswitch is pressed. When the footswitch is
released, the system activates the defrosting procedure of the tip of the cryosurgical
probe.
7.1.6
APPLICATION TIME CONTROL
The application time control occurs when the relative procedure (COUNT DOWN) is
activated by pressing the key next to the symbol on the display.
In this state, pressing the footswitch activates the freezing procedure of the cryosurgical
probe, which lasts until the end of the time period set by the operator. The expiration
of the application time is accompanied by a sound signal. At the expiration of the
application time, the cryosurgical probe is automatically defrosted.
During the freezing phase of the probe, the surgeon can stop the process at any time
and enter the defrosting phase by releasing the control footswitch of the system.
7.1.7
APPLICATION TIME AND APPLIED TEMPERATURE CONTROL
The application time and applied temperature control is activated by simultaneously
activating the temperature control procedure (TEMP CTRL) and the application time
procedure (COUNT DOWN).