Model

M-1157

ACCESSORIES Series 

INSTRUCTION MANUAL

Ver. 1.0 

2019

Содержание M-1157

Страница 1: ...Model M 1157 ACCESSORIES Series INSTRUCTION MANUAL Ver 1 0 2019...

Страница 2: ...inserts 7 5 3 Use of indicator markers 9 5 4 Installing the condenser 9 5 5 Centering the condenser 9 6 Observation 10 6 1 Brightfield observation BF 10 6 2 Darkfield obervation DF 11 6 3 Phase Contr...

Страница 3: ...this manual 2 Intended use Standard models For research and teaching use only Not intended for any animal or human therapeutic or diagnostic use IVD Models Also for diagnostic use aimed at obtaining i...

Страница 4: ...HRAGM LEVER APERTURE DIAPHRAGM SCALE OPTICAL INSERTS CENTERING SCREWS OPTICAL INSERTS HOUSINGS TOP LENS MOVEMENT LEVER POLARIZER SLIDER LOCKING SCREW ROTATING POLARIZER LOCKING SCREW POLARIZER ROTATIO...

Страница 5: ...htfield Darkfield Phase Contrast Simple Polarized Light Differential Interference Contrast Nomarski DIC The configuration of the condenser varies according to the chosen observation methods Once opene...

Страница 6: ...bjectives Phase Contrast inserts SL1 PH ring 10X 20X SL2 PH ring 40X 60 SL3 PH ring 100X SL4 PH ring 4X Darkfield inserts DF darkfield ring dry DF OIL darkfield ring oil DIC inserts U DIC10 DIC prism...

Страница 7: ...observation with oil 5 2 Installing optical inserts 1 Using the Allen key supplied with the microscope unscrew the fixing screw of the condenser cov er Fig 2 2 Remove the condenser cover to access the...

Страница 8: ...nd tighten the fixing screw Fig 2 F Fig 4 ig 4 F Fig 5 ig 5 F Fig 6 ig 6 DIC prisms can be installed also in the positions 4 and 8 Observation Mode Insert Condenser position PH SL1 SL2 SL3 SL4 1 2 3 D...

Страница 9: ...3 If necessary lower the top lens 4 Loosen the condenser locking screw Fig 8 5 Insert the condenser into the condenser holder and place the alignment pin in the groove on the back of the condenser ho...

Страница 10: ...e polarizer to remove it from the optical path Fig 11 3 Follow the indications in the table Usable objectives with different top lenses to optimize the observation When the top lens is removed the fie...

Страница 11: ...er or raise the condenser until a homogeneous illumination of the sample is obtained and therefore an optimal dark field effect is achieved 5 Using the centering screws on the condenser Fig 13 centre...

Страница 12: ...the upper part of the centering telescope until the two phase rings one dark and one bright visible in the telescope are in focus Fig 14 15 7 Using centering screws on the condenser Fig 13 center the...

Страница 13: ...from the nosepiece and insert the analyzer into the dummy slider then insert the assembly into the slot Fig 17 3 Push the lever of the polarizer to insert it in the optical path Fig 18 4 Remove the s...

Страница 14: ...ning the dedicated DIC prisms according to the objectives in use trans mitted light Analyzer DIC slider Fig 20 1 Rotate the condenser turret to enter the BF position no optical insert in the optical p...

Страница 15: ...Once the maximum darkening is achieved re move the slider from the nosepiece remove the analyzer from the dummy slider and insert it into the DIC prism Now insert the DIC slider into the slot Fig 24...

Страница 16: ...th iris Observation mode Insert Usable objectives BF IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F PH IOS U PLAN F APO DF DF DF OIL IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F APO PH SL1 SL2...

Страница 17: ...bservation modes and usable objectives Darkfield ring not well centered Center the ring correctly Image quality when observing in polar ized light is not high Polarizer is not inserted in the optical...

Страница 18: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Страница 19: ...Modello M 1157 Serie ACCESSORI MANUALE DI ISTRUZIONI Ver 1 0 2019...

Страница 20: ...o dei segnalini indicatori 27 5 4 Montaggio del condensatore 27 5 5 Centraggio del condensatore 27 6 Osservazione 28 6 1 Osservazione in Campo Chiaro BF 28 6 2 Osservazione in Campo Scuro DF 29 6 3 Os...

Страница 21: ...o nel presente manuale 2 Utilizzo previsto Modelli standard Solo per applicazioni di ricerca ed usi didattici Non indicato per utilizzo diagnostico e terapeutico umano e ve terinario Modelli IVD Anche...

Страница 22: ...URA SCALA DIAFRAMMA DI APERTURA VITI DI CENTRAGGIO INSERTI OTTICI ALLOGGIAMENTO INSERTI OTTICI LEVA MOVIMENTO LENTE FRONTALE VITE DI BLOCCO SLITTA POLARIZZATORE POLARIZZATORE GIREVOLE VITE DI BLOCCO R...

Страница 23: ...ontrasto di Fase Luce Polarizzata Semplice Contrasto Interferenziale Differen ziale Nomarski DIC La configurazione del condensatore varia in funzione delle metodiche di osservazione prescelte Una volt...

Страница 24: ...li Inserti per Contrasto di Fase SL1 anello PH 10X 20X SL2 anello PH 40X 60 SL3 anello PH 100X SL4 anello PH 4X Inserti per Campo Scuro DF anello campo scuro dry DF OIL anello campo scuro oil Inserti...

Страница 25: ...taggio degli inserti ottici 1 Utilizzando la brugola in dotazione con il micro scopio svitare la vite di fissaggio del coperchio del condensatore Fig 2 2 Rimuovere il coperchio del condensatore per ac...

Страница 26: ...zionare il coperchio e serrare la vite di fis saggio Fig 2 I prismi DIC possono essere installati anche nelle posizioni 4 e 8 F Fig 6 ig 6 Modo di Osservazione Inserto Posizione del condensatore PH SL...

Страница 27: ...nsatore 3 Se necessario abbassare la lente frontale 4 Allentare la vite di bloccaggio del condensatore Fig 8 5 Inserire il condensatore nel supporto e posizio nare il piedino di allineamento nell inta...

Страница 28: ...diaframma di campo funziona come diaframma di apertura 4 Posizionare un preparato sul tavolino 5 Mettere a fuoco 6 Iniziare l osservazione 7 Regolare il diaframma di apertura in funzione del tipo di c...

Страница 29: ...one omogenea del preparato e quindi un effetto ottimale in campo scuro 5 Utilizzando le viti di centraggio poste sul condensatore Fig 13 centrare l anello di campo scuro fino a che il campo visivo omo...

Страница 30: ...uperiore del telescopio per mettere a fuoco gli anelli uno chiaro ed uno scuro visibili nel telescopio Fig 14 15 7 Utilizzando le viti di centraggio poste sul condensatore Fig 13 centrare gli anelli i...

Страница 31: ...erire il gruppo nella fessura Fig 17 3 Spingere il comando del polarizzatore per in serirlo nel percorso ottico Fig 18 4 Rimuovere il vetrino dal tavolino 5 Allentare la vite di fissaggio della rotazi...

Страница 32: ...litta DIC Fig 20 1 Ruotare la torretta del condensatore per inserire la posizione BF nessun inserto ottico presente nel percorso ottico 2 Utilizzando la manopola inserire il polarizza tore incorporato...

Страница 33: ...olver rimuovere l analizzatore dalla slitta vuota ed inserirlo nel prisma DIC A questo punto inserire la slitta DIC nella fessura Fig 24 7 Ruotare la torretta del condensatore per in serire il prisma...

Страница 34: ...OS U PLAN F APO DF DF DF OIL IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F APO PH SL1 SL2 SL3 SL4 IOS U PLAN F PH PO IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F PH IOS U PLAN F APO DIC DIC10...

Страница 35: ...sultare il paragrafo 7 Modi di os servazione ed obiettivi utilizzabili L anello per campo scuro non ben centrato Centrare l anello in modo corretto La qualit dell immagine osservando in luce polarizza...

Страница 36: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Страница 37: ...Modelo M 1157 Serie ACCESORIOS MANUAL DE INSTRUCCIONES Ver 1 0 2019...

Страница 38: ...5 3 Uso de se ales indicadoras 45 5 4 Montaje del condensador 45 5 5 Centrado del condensador 45 6 Observaci n 46 6 1 Observaci n en Campo Claro BF 46 6 2 Observaci n en Campo Oscuro DF 47 6 3 Observa...

Страница 39: ...ellos que utilizan dicho instrumento 2 Utilizaci n Modelos est ndar Para uso exclusivo de investigaci n y docencia No est destinado a ning n uso terap utico o diagn stico animal o humano Modelos IVD T...

Страница 40: ...URA TORNILLOS CENTRADORES DE INSERTOS PTICOS CARCASA PARA INSERTOS PTICOS PALANCA DE MOVIMIENTO DE LA LENTE FRONTAL TORNILLO DE BLOQUEO DE LA CORREDERA DEL POLARIZADOR POLARIZADOR GIRATORIO TORNILLO D...

Страница 41: ...Simple Contraste Diferencial Interferencial Nomarski DIC La configuraci n del condensador var a seg n los m todos de observaci n elegidos Estas son las piezas que pertenecen al dispositivo y que enco...

Страница 42: ...sertos de Contraste de Fase SL1 anillo PH 10X 20X SL2 anillo PH 40X 60 SL3 anillo PH 100X SL4 anillo PH 4X Insertos de Campo Oscuro DF anillo campo oscuro dry DF OIL anillo campo oscuro oil Insertos d...

Страница 43: ...a ci n S lo para observaci n en aceite 5 2 Montaje de insertos pticos 1 Con la llave Allen suministrada con el microsco pio desenroscar el tornillo de fijaci n de la tapa del condensador Fig 2 2 Retir...

Страница 44: ...te el tornillo de fijaci n Fig 2 F Fig 4 ig 4 F Fig 5 ig 5 F Fig 6 ig 6 Los prismas DIC tambi n se pueden instalar en las posiciones 4 y 8 Modo de Observaci n Inserto Posici n del condensador PH SL1 S...

Страница 45: ...ecesario baje la lente frontal 4 Afloje el tornillo de bloqueo del condensador Fig 8 5 Inserte el condensador en el soporte del conden sador y coloque el pie de alineaci n en la ranura de la parte pos...

Страница 46: ...optimizar la observaci n Si se excluye la lente frontal el diafragma de campo funciona como diafragma de apertura 4 Poner una muestra sobre la platina 5 Enfoque la imagen 6 Comience la observaci n 7...

Страница 47: ...scuro 5 Con los tornillos de centrado del condensador Fig 13 centre el anillo de campo oscuro hasta que el campo de visi n se ilumine uniformemente La observaci n en campo oscuro requiere mucha ilumin...

Страница 48: ...llos uno claro y otro oscuro visibles en el telescopio Fig 14 15 7 Con los tornillos para centrar el condensador de fases Fig 13 intente centrar los anillos de modo que el aro brillante quede sobre pu...

Страница 49: ...tar la corredera vac a del rev lver e insertar el analizador en la carcasa vac a de la corredera luego insertar el ensamblaje en la ranura Fig 17 3 Pulsar el control del polarizador para insertarlo en...

Страница 50: ...z transmitida corredera DIC Fig 20 1 Gire la torreta del condensador para entrar en la posici n BF sin inserto ptico en la trayectoria ptica 2 Usando la perilla inserte el polarizador incor porado en...

Страница 51: ...ire la corredera del rev lver retire el analiza dor de la corredera vac a e ins rtelo en el prisma DIC Ahora inserte la corredera DIC en la ra nura Fig 24 7 Gire la torreta del condensador para insert...

Страница 52: ...con iris Modo de observaci n Inserto Objetivos utilizables BF IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F PH IOS U PLAN F APO DF DF DF OIL IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F APO...

Страница 53: ...e el p rrafo 7 Modos de obser vaci n y objetivos utilizables El anillo de campo oscuro no est bien centrado Centrar el anillo correctamente La calidad de la imagen cuando se ob serva con luz polarizad...

Страница 54: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Страница 55: ...Mod le M 1157 S rie ACCESSOIRES MANUEL D UTILISATION Ver 1 0 2019...

Страница 56: ...5 3 Utilisation de marqueurs indicateurs 63 5 4 Montage du condensateur 63 5 5 R glage du condensateur 63 6 Observation 64 6 1 Observation en Fond Clair BF 64 6 2 Observation en Fond Noir DF 65 6 3 Ob...

Страница 57: ...ument non indiqu e dans ce manuel 2 Emploi pr vu Mod les standard R serv la recherche et l enseignement Ne pas utiliser des fins th rapeutiques ou diagnostiques animales ou humaines Mod les de DIV gal...

Страница 58: ...GME D OUVERTURE VIS DE CENTRAGE DES INSERTS OPTIQUES BO TIER POUR INSERTS OPTIQUES LEVIER DE MOUVEMENT DE LA LENTILLE FRONTALE VIS DE BLOCAGE DU COULISSEAU DU POLARISEUR POLARISEUR ORIENTABLE VIS DE B...

Страница 59: ...ple Contraste Interf rentiel Diff rentiel No marski DIC La configuration du condensateur varie selon les m thodes d observation choisies Une fois l emballage ouvert les pi ces de l appareil sont les s...

Страница 60: ...les Inserts de Contraste de Phase SL1 anneau PH 10X 20X SL2 anneau PH 40X 60 SL3 anneau PH 100X SL4 anneau PH 4X Inserts de Fond Noir DF anneau fond noir dry DF OIL anneau fond noir oil Inserts de DIC...

Страница 61: ...uile 5 2 Montage des inserts optiques 1 A l aide de la cl Allen fournie avec le micros cope d vissez la vis de fixation du couvercle du condensateur Fig 2 2 Retirez le couvercle du condensateur pour a...

Страница 62: ...rrer la vis de fixation Fig 2 F Fig 4 ig 4 F Fig 5 ig 5 F Fig 6 ig 6 Les prismes DIC peuvent galement tre install s aux positions 4 et 8 Mode de Observation Inserton Position de condensateur PH SL1 SL...

Страница 63: ...lentille frontale 4 Desserrer la vis de blocage du condensateur Fig 8 5 Ins rez le condensateur dans le support de condensateur et placez le pied d alignement dans la fente l arri re du support de co...

Страница 64: ...tion Lorsque la lentille frontale est exclue le diaphragme de champ fonctionne comme diaphragme d ouverture 4 Poser un chantillon sur la platine 5 Faire la mise au point 6 Commencez l observation 7 Aj...

Страница 65: ...entrage du condensateur Fig 13 centrer la bague de fond noir jusqu ce que le champ de vision soit uniform ment clair Le fond noir n cessite une grande quan tit de lumi re En passant de la m thode d ob...

Страница 66: ...14 6 Tourner la partie sup rieure du t lescope pour faire la mise au point des anneaux un anneau clair et un anneau sombre visibles dans le t lescope Fig 14 15 7 Tourner les deux vis de centrage du c...

Страница 67: ...nsemble dans la fente Fig 17 3 Poussez la commande du polariseur pour l in s rer dans le chemin optique Fig 18 4 Retirer la lame de la platine 5 Desserrer la vis de fixation de la rotation du pola ris...

Страница 68: ...e lumi re transmise curseur DIC Fig 20 1 Tourner la tourelle du condenseur pour entrer en position BF pas d insert optique dans le che min optique 2 A l aide du bouton ins rez le polariseur int gr dan...

Страница 69: ...maximale trouv e retirez le curseur du revolver retirez l analyseur de le curseur vide et ins rez le dans le prisme DIC Ins rez maintenant le curseur DIC dans la fente Fig 24 7 Tourner la tourelle du...

Страница 70: ...f 100X avec iris Mode de observation Insert Objectifs utilisables BF IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F PH IOS U PLAN F APO DF DF DF OIL IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN...

Страница 71: ...dapt Voir paragraphe 7 Modes d observation et objectifs utilisables L anneau de fond noir n est pas bien centr Centrer l anneau correctement La qualit de l image vue en lumi re polaris e n est pas lev...

Страница 72: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Страница 73: ...Modell M 1157 ZUBEH R Serie BEDIENUNGSANLEITUNG Ver 1 0 2019...

Страница 74: ...wendung von Indikator Markern 81 5 4 Montage des Kondensators 81 5 5 Zentrierung des Kondensators 81 6 Beobachtung 82 6 1 Beobachtung im Hellfeld BF 82 6 2 Beobachtung im Dunkelfeld DF 83 6 3 Beobacht...

Страница 75: ...ben ist 2 Verwendung Standardmodelle Nur f r Forschung und Lehre verwenden Nicht f r therapeutische oder diagnostische Zwecke bei Tieren oder Menschen bestimmt IVD Modelle Auch f r diagnostische Zweck...

Страница 76: ...APERTURBLENDE SKALA SCHRAUBEN MIT ZENTRIERENDEN OPTISCHEN EINS TZEN GEH USE F R OPTISCHE EINS TZE FRONTLINSEN BEWEGUNGSHEBEL POLARISATOR SCHIEBESCHRAUBE SCHWENKBARER POLARISATOR POLARISATOR ROTATIONSF...

Страница 77: ...es Polarisiertes Licht Nomarski Differential Interferential Contrast DIC Die Konfiguration des Kondensators variiert je nach den gew hlten Beobachtungsmethoden Nach dem ffnen der Verpackung sind die T...

Страница 78: ...Objektive Phasenkontrast Eins tze SL1 PH Ring 10X 20X SL2 PH Ring 40X 60 SL3 PH Ring 100X SL4 PH Ring 4X Dunkelfeld Eins tze DF Dunkelfeldring dry DF OIL Dunkelfeldring oil Eins tze f r DIC U DIC10 DI...

Страница 79: ...l 5 2 Montage von optischen Eins tzen 1 L sen Sie mit dem im Lieferumfang des Mikro skops enthaltenen Inbusschl ssel die Befesti gungsschraube der Kondensatorabdeckung Fig 2 2 Entfernen Sie die Konde...

Страница 80: ...ziehen Sie die Befestigungsschraube an Fig 2 DIC Prismen k nnen auch in den Positionen 4 und 8 installiert werden F Fig 4 ig 4 F Fig 5 ig 5 F Fig 6 ig 6 Beobach tungsmodus Einsetzen Position des Konde...

Страница 81: ...ch senken Sie die Frontlinse ab 4 L sen Sie die Sicherungsschraube des Konden sators Fig 8 5 Setzen Sie den Kondensator in den Kondensa torhalter ein und setzen Sie den Justierfu in den Schlitz auf de...

Страница 82: ...t werden k nnen um die Beobachtung zu optimieren Wenn die Frontlinse ausgeschlossen ist fungiert die Feldblende als Aperturblende 4 Eine Zubereitung auf den Tisch legen 5 Fokussierung 6 Starten Sie di...

Страница 83: ...und damit eine optimale Wirkung im Dunkelfeld erreicht ist 5 Zentrierschrauben am Kondensator Fig 13 verwenden um den Dunkelfeldring zu zentrieren bis das Sichtfeld gleichm ig ausgeleuchtet ist Das Du...

Страница 84: ...eskop ein Fig 14 6 Drehen Sie die Oberseite des Teleskops um sich auf die im Teleskop sichtbaren Ringe einer hell und einer dunkel zu fokussieren Fig 14 15 7 ZentrierenSiedieRingemitdenZentrierschraub...

Страница 85: ...die Baugruppe in den Schlitz ein Fig 17 3 Dr cken Sie die Steuerung des Polarisators um ihn in den optischen Pfad einzuf gen Fig 18 4 Entfernen Sie den Schlitten vom Tisch 5 L sen Sie die Befestigungs...

Страница 86: ...Dur chlicht Schlitten DIC Fig 20 1 Drehen Sie den Kondensersturm um in die Position BF zu gelangen kein optischer Einsatz im optischen Pfad 2 Stecken Sie den eingebauten Polarisator mit dem Drehknopf...

Страница 87: ...Sie den Analysator aus dem leeren Schieber und setzen Sie ihn in das DIC Prisma ein Stecken Sie nun den DIC Schieber in den Steckplatz ein Fig 24 7 Drehen Sie den Kondensatosturm um das DIC Prisma ent...

Страница 88: ...it Irisblende Beobach tungsmodus Einsetzen Verwendbare Objektive BF IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F PH IOS U PLAN F APO DF DF DF OIL IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F...

Страница 89: ...rt Den Ring richtig zentrieren Die Bildqualit t bei Betrachtung im po larisierten Licht ist nicht hoch Der Polarisator ist nicht in den optisch en Weg eingef hrt Setzen Sie den Polarisator in den op t...

Страница 90: ...92 info optikamicroscopes com www optikamicroscopes com OPTIKA Spain spain optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optika...

Страница 91: ...Modelo M 1157 ACESS RIOS S rie MANUAL DE INSTRU ES Ver 1 0 2019...

Страница 92: ...3 Utiliza o de sinais indicadores 99 5 4 Montagem do condensador 99 5 5 Centragem do condensador 99 6 Observa o 100 6 1 Observa o em Campo Claro BF 100 6 2 Observa o em Campo Oscuro DF 101 6 3 Observa...

Страница 93: ...neste manual 2 Utiliza o prevista Modelos padr o Apenas para uso em pesquisa e ensino N o se destina a qualquer uso terap utico ou diagn stico animal ou humano Modelos IVD Tamb m para uso diagn stico...

Страница 94: ...DE ABERTURA PARAFUSOS DE CENTRALIZA O DE INSERTOS PTICOS ENCAIXE PARA INSERTOS PTICOS ALAVANCA DE MOVIMENTO LENTE FRONTAL PARAFUSO DE TRAVA DE CORREDI A DO POLARIZADOR POLARIZADOR GIRAT RIO PARAFUSO D...

Страница 95: ...izada Simples Contraste Interferencial Diferencial Nomarski DIC A configura o do condensador varia de acordo com os m todos de observa o escolhidos Uma vez aberta a embalagem as partes do dispositivo...

Страница 96: ...sertos de Contraste de Fase SL1 anel PH 10X 20X SL2 anel PH 40X 60 SL3 anel PH 100X SL4 anel PH 4X Insertos de Campo Escuro DF anel de campo escuro dry DF OIL anel de campo escuro oil Insertos de DIC...

Страница 97: ...icos 1 Utilizando a chave Allen fornecida com o micros c pio desaparafuse o parafuso de fixa o da tampa do condensador Fig 2 2 Remova a tampa do condensador para acessar as caixas dos insertos pticos...

Страница 98: ...aperte o parafuso de fixa o Fig 2 Os prismas DIC tamb m podem ser instalados nas posi es 4 e 8 F Fig 4 ig 4 F Fig 5 ig 5 F Fig 6 ig 6 Modo de Observa o Inserto Posi o do condensador PH SL1 SL2 SL3 SL4...

Страница 99: ...3 Se necess rio baixe a lente frontal 4 Solte o parafuso de bloqueio do condensador Fig 8 5 Insira o condensador no suporte do condensa dor e coloque o p de alinhamento na ranhura na parte de tr s do...

Страница 100: ...a o Quando a lente frontal exclu da o diafragma de campo funciona como um diafragma de abertura 4 Colocar uma prepara o sobre a platina 5 Focalize em 6 Iniciar a observa o 7 Ajuste o diafragma de aber...

Страница 101: ...ulares levantar ou abaixar o condensador at que uma ilumina o homog nea da amostra possa ser alcan ada obtendo se assim um efeito de campo escuro adequado 5 Usando os parafusos de centraliza o no cond...

Страница 102: ...esc pio de centragem at que os dois an is de fase um escuro e um brilhante vis veis no telesc pio estejam focados Fig 14 15 7 Usando parafusos de centraliza o no condensador Fig 13 centre os an is de...

Страница 103: ...o conjunto no slot Fig 17 3 Pressione o controle do polarizador para inse ri lo no caminho ptico Fig 18 4 Retire a amostra da platina 5 Desaperte o parafuso de fixa o da rota o do polarizador Fig 18 6...

Страница 104: ...lisador para luz transmitida slide DIC Fig 20 1 Gire a torre do condensador para entrar na posi o BF sem inser o ptica no caminho ptico 2 Utilizando o bot o insira o polarizador incor porado no conden...

Страница 105: ...mo retire a l mina do rev lver retire o analisador da l mina vazia e insira a no prisma do DIC Ago ra insira o slide DIC no slot Fig 24 7 Gire a torre do condensador para inserir o prisma DIC corresp...

Страница 106: ...o com ris Modo de observa o Inserto Objectivos utiliz veis BF IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F PH IOS U PLAN F APO DF DF DF OIL IOS W PLAN IOS W PLAN F IOS U PLAN F IOS U PLAN F APO P...

Страница 107: ...quada Ver ponto 7 Modos de observa o e objectivos utiliz veis Anel de campo escuro n o bem cen trado Centrar o anel corretamente A qualidade da imagem quando visu alizada em luz polarizada n o alta O...

Страница 108: ...optikamicroscopes com OPTIKA USA usa optikamicroscopes com OPTIKA China china optikamicroscopes com OPTIKA India india optikamicroscopes com OPTIKA Central America camerica optikamicroscopes com 100...