Wiring
- 19 -
optek-Manual--1004-1022-02--156-US-2017-12-11
8
Wiring
8.1
General advice on wiring
Danger!
Electrical voltage!
Before connecting, switch the converter power switch (22) to O (OFF) position!
Install electrical connections only by qualified electricians!
Fig. 10
Power switch on converter 115 / 230 V (left) and 24 V (right)
• The power switch is only protected against accidental touching. That is
why the converter has to be installed in a control cabinet (see chapter 7,
• The power supply socket is not provided with a first-to-mate last-to-break
protective ground contact. You must therefore provide an external release
device.
• The converter is designed for continuous operation and has only to be
switched voltage-free for maintenance purposes.
ALARM II
14