UV-Photometer  

FlowMissio-C 

 

 

manual

 

 

Version: 3.0 

 
 

 

Содержание FlowMissio-C

Страница 1: ...Opsytec Dr Gr bel GmbH Goethestra e 17 76275 Ettlingen Tel 07243 94 783 50 Fax 07243 94 783 65 info uv groebel de UV Photometer FlowMissio C manual Version 3 0...

Страница 2: ...age 9 8 Calculations 9 9 Maintenance and service 10 9 1 Exchange of the UV lamp 10 9 2 Cleaning of the cuvette 10 9 3 Exchange of the silica gel 10 10 Technical data 12 10 1 Device 12 10 2 Lamp 12 10...

Страница 3: ...entioned instructions can result in an injury of the user Meaning Failure to observe the mentioned instructions can result in a damage of the device Meaning Instructions are to be observed for the reg...

Страница 4: ...g output is menu driven and controlled via touch switches For pro cess controlling two variable switch thresholds are available The photometer is designed for an absorption path length of 50 mm The gr...

Страница 5: ...has to be disconnected from the power supply and the absence of voltage is to be checked The system is equipped with UV C lamp There is a risk of photochem ical damage of the eye erythema and skin can...

Страница 6: ...lead to damages at the equipment Connect water inlet and water outlet Then connect clamps and mains After the power on of the equipment the following start logo appears Afterwards the transmission me...

Страница 7: ...thresholds In the upper right corner the threshold are indicated with a two line pictogram Again upper line indicates threshold 1 lower line indicates threshold 2 In this example the threshold 2 is r...

Страница 8: ...tion The FlowMissio determines a calibration factor for the actual measure ment made to achieve the reference value This calibration factor is stored and you are back in the main menu Press for 2 seco...

Страница 9: ...two values the transmission T 50 is calculated Lampe Trans I I c T 50 The constant c is a calibration factor which is determined in the menu reference value see Calibration formerly reference value Du...

Страница 10: ...vacuum during removing the water from the cuvette Always keep the water inlet open Never open the cuvette with water pressure pre sent The cuvette is not vacuum resistant Disconnect the plug from the...

Страница 11: ...bags into the housing Now close the housing again Be sure the housing is closed and no cables are jammed Also note that the transparent front cover should always be closed Close the cable inlets at t...

Страница 12: ...kg Operating temperature 0 to 40 C Housing IP65 Signal output 2 voltage free commutators 250V 1A Analog output 4 20 mA 20mA 110 Power supply 100 240 V 50 60 Hz 500 mA 10 2 Lamp Lamp low pressure Mercu...

Страница 13: ...imilar alarm 4 Alarm 5 NO theshold 2 similar Pre alarm 6 NC theshold 2 similar Pre alarm 7 Pre alarm X3 1 PE 2 4 20 mA High 0 110 3 4 20 mA Low X4 RS232 1200Baud 8 Data 1 Stop None 1 Transmit Data TxD...

Страница 14: ...10 Technical data 14 10 4 Spare parts Description Order number Lamp insert 93070101 Sealing set cuvette 93070102 Silica gel 93070103 UVCS Sensor 810312 Subject to changes...

Страница 15: ...10 Technical data 15 10 5 drawing 215 0 330 0 305 0 401 5 167 241 0 220 0 285 0 Schlauch DN11...