LBL-2008-41-v5 OptoMonitor 3 IFU_en Rev 2021-08
85
|
P a g e
In both cases, a text file will be saved on the USB key. This text files contains all the logs and gives insight
on the history of the DU or OU operations. This can help troubleshooting. Once the message “Done!”
appears in the pop-up message, the dump is completed, and the USB drive may be removed.
If the dumping failed, a pop-up window will appear with an error message.
The message can be:
“Make sure the OU is connected”: The OU and DU must be connected in order to dump the OU logs. The
connection can be wireless or via the cable.
“Make sure USB drive is connected”: A USB drive must be inserted in the DU (or through the USB
extension cable) in order to dump the logs.
Adjustable settings
The following table shows the various adjustable settings with their ranges and default values.
SETTINGS MENU
Setting
Range
Default