background image

Instructions and 

for Opera

®

 Relieve Overlay Mattress System

                     Opera Care, Azure House, Connaught Road, Hull, HU7 3AP

                 0333 222 8584 | [email protected] | opera-care.co.uk

  

Содержание Relieve Overlay

Страница 1: ...Instructions and for Opera Relieve Overlay Mattress System Opera Care Azure House Connaught Road Hull HU7 3AP 0333 222 8584 support opera care co uk opera care co uk...

Страница 2: ...ervice 11 14 Warranty 11 15 12 Thank you for purchasing this anti decubitus mattress replacement system PLEASE READ THESE INSTRUCTIONS CAREFULLY BEFORE SETTING UP AND USING THE DEVICE Pay special atte...

Страница 3: ...e ulcers by facilitating blood circulation and decreasing pressure of each tissue s contact area Always consult a physician or healthprofessional before using this mattress system This anti decubitus...

Страница 4: ...Hazard exists This device draws air from the surrounding environment Thus cigarette smoking may damage internal components 2 4 Disinfection Between Patients This system should be disinfected thorough...

Страница 5: ...l make sure to follow national state and or facility requirements for disposal of the unit 3 6 Maximum Operating Temperature The air mattress operates at a max of 100 F 37 7 C 3 7 Sharp Protrusions CA...

Страница 6: ...e internal pressure of the mattress according to patient weight Simply turn the knob to adjust the pressure setting if the mattress is too soft or firm to suit each patient s needs Caregivers should a...

Страница 7: ...ower Cord provided with a molded on anti kink bushing held in place by integral slot in bottom enclosure 5 Hanging Hooks Hanging hooks are designed to hang the power unit on almost any foot board 5 7...

Страница 8: ...EN 60601 2 52 2010 6 2 Secure Power Unit Secure the power unit onto the foot board using hanging hooks 6 4 Connect Hose Couplers Firmly connect the air hose couplers to the couplers on the power unit...

Страница 9: ...TUCKING THE BEDSHEET SHEET IN TOO TIGHTLY MAY REDUCE THE EFFECTIVENESS OF THE SYSTEM 8 Patient Transfer and Transport 8 1 Transfer It is recommended to have the mattress system fully inflated during...

Страница 10: ...ying with a clean dry cloth Use only mild detergents to clean the coverlet and the mattress Any appropriate NON PHENOLIC cleaning agent may be used for heavy and coverlet are completely dry before let...

Страница 11: ...ment is not and or Bottoming out is occurring and or Patient leaves a deep indentation at the contact area which does not return back to its original shape Check CPR Pull Cord Unzip mattress cover and...

Страница 12: ...e production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by our company or if in the sole opinion of our company that the system sh...

Страница 13: ...insulated IPX0 do not immerse power unit in any liquid or spray any liquids directly on the power unit This system is not AP APG Protected Continuous operation Power Cord Operating Transportation Sto...

Страница 14: ...in which radiated RF disturbances are controlled The customer or the user of the Air Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF...

Страница 15: ...1000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrupt...

Страница 16: ...Opera Care Azure House Connaught Road Hull HU7 3AP 0333 222 8584 support opera care co uk opera care co uk U 2S2301 30831 G V1 1...