4.5.12.
AUTO POWER OFF
Selecting this item allows the AUTO POWER OFF function to be configured. It is possible to
configure any number of days from 01 to 28.
The AUTO POWER OFF timer will be ENABLED immediately once any non-zero value is set.
While ENABLED, each time the unit is powered ON the AUTO POWER OFF timer will restart and
the unit will automatically shut down after the configured number of days has elapsed.
To DISABLE the AUTO POWER OFF feature the number of days must be set to 00.
The AUTO POWER OFF feature will only operate when the unit is being powered from the internal
batteries. If the unit has an active external power connection, then it will be automatically
powered on at all times and this feature should be disabled.
4.5.13.
RELAY LOGIC
Selecting this item allows the RELAY LOGIC to be configured.
If RELAY LOGIC is configured as NORMAL then while in the LISTENING state, the Normally Closed
contacts of the relays will be connected to their respective Common contacts and the Normally
Open contacts will be open. While in the DETECTION state, the Normally Closed contacts of the
relays will be open and the Normally Open contacts will be connected to their respective Common
contacts.
If RELAY LOGIC is configured as INVERTED then then while in the LISTENING state, the Normally
Closed contacts of the relays will be open and the Normally Open contacts will be connected to
their respective Common contacts. While in the DETECTION state, the Normally Closed contacts
of the relays will be connected to their respective Common contacts and the Normally Open
contacts will be open.
4.5.14.
SENSOR SELECT
Selecting this item allows the active magnetic sensor to be selected.
If LOCAL is selected, then the INTERNAL magnetic sensor will be active.
If REMOTE is selected, then the REMOTE magnetic sensor will be active.
4.5.15.
SENSOR OUTPUT
Selecting this item allows the user to configure which components of the magnetic field are used
by the unit to detect pig passages. There are currently four possible selections.
VECTOR XYZ
–
Full 3D magnetic field strength will be used.
X-AXIS
–
Only the x component will be used.
Y-AXIS
–
Only the y component will be used.
Z-AXIS
–
Only the z component will be used.
4.5.16.
LED FLASH
Selecting this item allows the HIGH BRIGHTNESS LED INDICATORS fitted to the front of unit to
be ENABLED or DISABLED.