OMS LINEA PATAVIUM S.T.01/3X ED.0 REV.8 09/2017
64
TESTING AND PUTTING INTO OPERATION
Note:
During testing, the device's mechanical stability must be checked once all moving parts and
accessories (table, light, spittoon unit, etc.) have been set to their most unfavourable
positions. If the chair cannot be fastened to the floor, stabilizer elements are available.
1.
if an unusual tilting is identified, adjust the instrument table using the screw located inside
the front part of the arm.
2.
Check the correct setup of the chair-dental unit group; if an unusual tilting is identified,
loosen the 4 screws securing the spittoon unit and relative support and then adjust the 4
dowels until the correct position of the dental unit-chair group is achieved.
3.
Check the correct setup of the spittoon unit-instrument block: if an unusual tilting is
identified in the instrument block, adjust it using the screw located inside the front part of
the arm.
4.
Check the manual pantograph base and backrest up and down controls, and the automatic
return to exit, rinse position and return to last position control to ensure they are working
properly. Check and store correct positioning of the 3 programs (see sections 4.4.1 MANUAL
CONTROLS and 4.4.2 AUTOMATIC CONTROLS).
5.
Check that limit switches and safety systems are working properly (see section 4.1 SAFETY
SYSTEMS).
6.
Check correct calibration of the pressure regulators and general flow regulators and that of
each instrument. These devices have already been calibrated by O.M.S. in the workshop, but
they may require an inspection or re-calibration.
7.
Make sure that the water flow to the cup and spittoon is normal.
Note:
The equipment is equipped with a general water tap. During use, the tap should be open. At
the end of the work day, the tap should be closed or, alternatively, close the unit’s general
supply tap in the surgery.
DANGER
Before performing any operations in the equipment, always make sure that the power switch is
off.