![Omron ZW-7000 Series Скачать руководство пользователя страница 145](http://html1.mh-extra.com/html/omron/zw-7000-series/zw-7000-series_user-manual_742279145.webp)
Setting the Output Conditions
ZW-8000/7000/5000
User's Manual
143
4
Settin
gs for Function
4-3 Setting the Output Conditions
Setting Scaling
This setting is used when you want to correct any errors that are generated due to the installation status of the
Sensor Head, and display the corrected value on the main display as a measured value.
There are three types of setting: “auto scaling” (“1-point scaling”, “2-point scaling” and “Multi-point scaling”) that
automatically sets the correction value of a placed sensing object, “fixed scaling” that manually sets the
correction value and “thickness scaling” that automatically sets the correction value of a placed sensing object
(transparent object).
The "Zero reset" p.100 settings return to the default settings when scaling is set. If scaling is set, perform a zero reset
as necessary.
Performing Manual Scaling
Set the scaling by entering the correction value.
Item
Setting item
Setting value
Description
Scaling
(Manual scaling)
Scaling
OFF (default value)/ON
Set scaling ON/OFF.
Span
-10.0000 to 10.0000
(default value: 1.0000)
Set the inclination of the sensor characteristics as a
coefficient.
Offset
-999.999999 to 999.999999
(default value: 0)
A fixed value is added to/subtracted from the measured
value.
Important
-10.0
10.0
Me
asu
rement v
a
l
u
e
(mm)
Workpiece
di
s
pl
a
cement
Measurement value
(mm)
Workpiece
displacement
Содержание ZW-7000 Series
Страница 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Страница 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Страница 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Страница 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 289: ...9 APPENDICES APPENDICES 9 1 Specifications and External Dimensions 288 9 2 Laser safety 323 9 3 Firmware Update 324 ...
Страница 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Страница 335: ......