background image

EN

FR

DE

IT

ES

NL

• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
•     

RU

AR

Wrist Blood Pressure Monitor

Model R2

IM-HEM-6113-E(V)-02-10/2011

5330293-3B

Содержание R2

Страница 1: ...E IT ES NL Instruction Manual Mode d emploi Gebrauchsanweisung Manuale di istruzioni Manual de instrucciones Gebruiksaanwijzing RU AR Wrist Blood Pressure Monitor Model R2 IM HEM 6113 E V 02 10 2011 5330293 3B ...

Страница 2: ...sing the Unit 11 3 1 Applying the Wrist Cuff 11 3 2 Correct Posture 13 3 3 Taking a Reading 14 3 4 Using the Memory Function 16 4 Quick Reference Guide 18 5 Handling Errors and Problems 19 5 1 Error Messages 19 5 2 Troubleshooting 20 6 Maintenance and Storage 22 7 Technical Data 24 8 Some Useful Information about Blood Pressure 27 ...

Страница 3: ...controlled inflation without the need of pressure pre setting or re inflation the device uses its advanced IntelliSense technology Clinical research has proven a direct relationship between blood pressure in the wrist and blood pressure in the arm Changes in wrist blood pressure reflect changes in arm blood pressure because the arteries in the wrist and the arm are connected Frequently measuring t...

Страница 4: ...Battery Usage If battery fluid should get in your eyes immediately rinse with plenty of clean water Consult a doctor immediately Caution Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property General Usage Do not leave the unit unattended with infants or persons who cannot express t...

Страница 5: ...be used for three months or more Do not use new and used batteries together General Precautions Do not apply strong shocks and vibrations to or drop the unit Do not take measurements after bathing drinking alcohol smoking exercising or eating Do not inflate the wrist cuff when it is not wrapped around your wrist Do not wash the wrist cuff or immerse it in water Read and follow the Important inform...

Страница 6: ...6 1 Overview Main Unit A Display B Battery compartment cover C Wrist cuff D O I START button E MEM Memory button A B C D E ...

Страница 7: ...symbol 1 Flashes during measurement 2 If flashing after measurement completed or when viewing results stored in the memory indicates blood pressure out of recommended range K Pulse L Deflation symbol Note If your systolic or diastolic pressure is outside the standard range above 135 85 mmHg the Heartbeat symbol will blink Please refer to 3 3 Taking a Reading F J L K G I H ...

Страница 8: ...8 1 Overview Package contents M Two AAA alkaline LR03 batteries N Storage case Blood pressure pass Blood Pressure Monitor unit Instruction manual Guarantee card M N ...

Страница 9: ...ling it off in the direction of the arrow 1 Push the hook on the bottom of the battery cover 2 Pull the cover off the main unit 2 Insert two identical 1 5V AAA alkaline LR03 batteries in the battery compartment Make sure their polarity is aligned with the polarity as indicated in the battery compartment 1 2 ...

Страница 10: ...tteries with new ones as soon as possible When the symbol remains lit the batteries are exhausted You should replace the batteries with new ones at once Turn the unit off before replacing the batteries Remove the batteries if the unit will not be used for three months or more Dispose of batteries according to applicable local regulations Two new identical 1 5V AAA alkaline batteries will last for ...

Страница 11: ...ent Omron recommends to always use the same arm for measurement If the values between the two arms differ substantially please check with your physician which arm to use for your measurement To ensure correct measurement apply the wrist cuff so that it fits comfortably around your wrist Do not apply over clothing 1 Place the wrist cuff over your left wrist with your left thumb facing upward Note M...

Страница 12: ...u have a slender wrist the cuff may feel slightly loose However this will not affect measurement results and you should not force the cuff to fit The remaining part of the wrist cuff can be conveniently folded back out of the way If the remaining part of the cuff is difficult to fold back measurement results will not be effected Taking a reading on the right wrist Measurements can also be made on ...

Страница 13: ...rinking alcohol or exercising Do not move or talk during measurement Position your arm so that the unit is at the same level as your heart Sit comfortably with your back straight hold your arm across your chest and relax Relax your wrist and hand Do not bend your wrist back Do not clench your fist Do not bend your wrist forward Do not use your other hand to support the wrist cuff This can result i...

Страница 14: ...on 2 After the unit has detected your blood pressure and pulse rate the cuff automatically deflates and your blood pressure and pulse rate are displayed Notes Sit still and do not talk or move until the measurement is completed Keep the unit at heart height until the measurement is completed To stop measurement press the O I START button at any time during measurement ...

Страница 15: ...diastolic pressure is outside the standard range the heartbeat symbol will blink when the measurement result is displayed Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home This criteria is for home blood pressure measurement For professional office blood pressure measurement criteria please refer to Classification of Blo...

Страница 16: ...ew set To View Previous Readings Stored in Memory 1 Press the MEM button to view readings stored in memory from the most recent to the oldest The Memory number appears for a second before the pulse rate is displayed The newest set is numbered 1 Press the MEM button repeatedly to cycle through the readings 2 Press the O I START button to turn the unit off Note If you forget to turn the unit off it ...

Страница 17: ...delete stored readings first press the MEM button Then while holding it down press the O I START button simultaneously for about 2 3 seconds All readings will then be deleted Note After the reading have been deleted the screen to the right of the illustration above appears and Memory symbol flashes Release the buttons to return to the memory display ...

Страница 18: ...you should be in a relaxed seated position 1 Align the wrist cuff with the level of your heart and gently support your left arm with your right hand Do not place your right hand on the cuff itself 2 Press the O I START button Remain quiet sit still and do not talk during the measurement Notes Always wait at least 2 3 minutes before taking another blood pressure measurement You may require more res...

Страница 19: ... a Reading Movement during measurement The wrist cuff is not fastened securely Carefully read and repeat the steps listed under 3 1 Applying the Wrist Cuff An E mark with a code number indicates the device has a hardware failure Consult your OMRON retail outlet or distributor The battery power is low Replace the batteries with two new AAA alkaline LR03 batteries Blinks or appears continuously ...

Страница 20: ... the Wrist Cuff Are your arms and shoulders tense Relax and try taking the measurement again Refer to 3 3 Taking a Reading Movement or talking during measurement Remain still and do not talk during measurement Refer to 3 3 Taking a Reading Wrist cuff pressure does not rise Air is leaking from the wrist cuff Consult your OMRON retail outlet or distributor Wrist cuff deflates too soon The wrist cuff...

Страница 21: ...rement The batteries are drained Replace the batteries with new ones Nothing happens when you press the buttons The batteries are drained Replace the batteries with new ones The batteries have been inserted incorrectly Insert the batteries with the correct polarity Other problems Press the O I START button and repeat measurement If the problem continues try replacing the batteries with new ones If...

Страница 22: ...ns Dropping the Unit Cleaning the unit with volatile liquids The unit should be cleaned with a soft dry cloth Use a soft moistened cloth and soap to clean the cuff Keep the unit in its storage case when not in use Fold the cuff into the storage case Do not store the unit in the following situations If the unit is wet Locations exposed to extreme temperatures humidity direct sunlight dust or corros...

Страница 23: ...uracy Please consult your authorised OMRON dealer or the OMRON Customer Service at the address given on the packaging or attached literature If the wrist cuff needs to be replaced have this done by an authorised expert Consult your authorised OMRON dealer or the OMRON Customer Service Do not carry out any repairs by yourself If a defect occurs or you have doubts about the correct functioning of th...

Страница 24: ... rapid deflation Power Source Two 1 5V AAA alkaline LR03 batteries Battery Life Approximately 300 measurements when used twice daily with alkaline batteries at a room temperature of 23 C Operating Temperature Humidity 10 C to 40 C 30 to 85 RH Storage Temperature Humidity 20 C to 60 C 10 to 95 RH Weight of Main Unit Approximately 117 g not including batteries Outer Dimensions 71 mm w x 41 mm h x 70...

Страница 25: ... in accordance with the national regulations for the disposal of electronic products This device fulfils the provisions of EC directive 93 42 EEC Medical Device Directive This blood pressure monitor is designed according to the European Standard EN1060 Non invasive sphygmomanometers Part 1 General Requirements and Part 3 Supplementary requirements for electromechanical blood pressure measuring sys...

Страница 26: ...his may result in incorrect operation of the unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in accordance with EN60601 1 2 2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual Documentation is also available at ...

Страница 27: ...measured by a doctor can cause anxiety which is itself a cause of high blood pressure As a variety of conditions affect blood pressure a single measurement may not be sufficient for an accurate diagnosis Many factors such as physical activity anxiety or the time of day can influence your blood pressure Thus it is best to try and measure your blood pressure at the same time each day to get an accur...

Страница 28: ...ent departments of hospitals There is no universally accepted definition of hypotension However those having the systolic pressure below 100 mmHg are assumed as hypotensive 200 150 100 50 6 12 18 24 Upper curve systolic blood pressure Lower curve diastolic blood pressure Time of day mmHg Example fluctuation within a day male 35 years old 80 180 160 140 130 120 85 90 100 110 Diastolic blood pressur...

Страница 29: ...ealthcare com Production facility OMRON HEALTHCARE MANUFACTURING VIETNAM CO LTD Binh Duong Province VIETNAM Subsidiary OMRON HEALTHCARE UK LTD Opal Drive Fox Milne Milton Keynes MK15 0DG U K OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH John Deere Str 81a 68163 Mannheim GERMANY www omron medizintechnik de OMRON SANTÉ FRANCE SAS 14 rue de Lisbonne 93561 Rosny sous Bois Cedex FRANCE Made in Vietnam ...

Страница 30: ...30 Notes ...

Отзывы: