Digital Automatic 

Blood Pressure Monitor

Model 

M2 Classic

Instruction Manual

Introduction

Thank you for purchasing the OMRON M2 Classic Digital Automatic Blood Pressure 
Monitor.

The OMRON M2 Classic is a fully automatic blood pressure monitor, operating on the 
oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. 
For comfortable controlled inflation without the need of pressure pre-setting or 
re-inflation the devices uses its advanced “IntelliSense” technology.

The unit also stores up to 30 measurements in memory.

Please read this instruction manual thoroughly before using the unit. For specific 
information about your own blood pressure, CONSULT YOUR DOCTOR.

Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia 
or arteriosclerosis.
Please read this section carefully before using the unit.

Warning:  

• Indicates a potentially hazardous situation which, if not avoided, could result 

in death or serious injury.

(General Usage)

• Always consult your doctor. Self-diagnosis of measurement results and 

self-treatment are dangerous.

• People with severe blood flow problems, or blood disorders, should consult a 

doctor before using the unit, as cuff inflation can cause internal bleeding.

(AC Adapter Usage) 

• Never plug in or unplug the power cord from the electric outlet with wet hands. 

(Battery Usage)

• If battery fluid should get in your eyes, immediately rinse with plenty of clean 

water. Consult a doctor immediately.

Caution: 

• Indicates a potentially hazardous situation which, if not avoided, may result in minor or 

moderate injury to the user or patient or damage to the equipment or other property.

(General Usage)

• Do not leave the unit unattended with infants or persons who cannot express their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result 

in incorrect measurement results.

• Make sure that the air tube is not wrapped around other parts of your body 

when taking measurements at night. This could result in injury when the air 
pressure in the air tube is increased.

• Do not leave the cuff wrapped on the arm if taking measurements during the 

night. This could result in injury.

• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields, 

near the unit. This may result in incorrect operation of the unit.

• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to Chapter 3.3. If the cuff is over inflated, it 

can cause internal bleeding.

(AC Adapter Usage) 

• Use only the original AC adapter designed for this unit. Use of unsupported 

adapters may damage and/or may be hazardous to the unit.

• Plug the AC adapter into the appropriate voltage outlet. 
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off 

the power and unplug the power cord immediately. 

(Battery Usage)

• If battery fluid should get on your skin or clothing, immediately rinse with 

plenty of clean water.

• Use only four “AAA” alkaline batteries with this unit. Do not use other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at 

the same time.

• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.

General Precautions

• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking, 

exercising or eating.

• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the 

main unit not the tube itself.

• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic 

Compatibility (EMC)” in the Technical Data Section.

• Read and follow the “Correct Disposal of This Product” in the Technical Data 

Section when disposing of the device and any used accessories or optional parts.

What is Irregular Heartbeat?

An irregular heartbeat is a heartbeat rhythm 
that varies by more than 25% from the 
average heartbeat rhythm detected 
while the unit is measuring the systolic 
and diastolic blood pressure.
If such an irregular rhythm is detected 
more than twice during measurement, 
the irregular heartbeat symbol (    ) 
appears on the symbol when the 
measurement results are displayed. 

What is Arrhythmia?

A heartbeat is stimulated by electrical signals that cause the heart to contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in 
the bio-electrical system that drives the heartbeat. Typical symptoms are 
skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) or 
slow (bradycardia) pulse. This can be caused by heart disease, aging, physical 
predisposition, stress, lack of sleep, fatigue etc. Arrhythmia can only be 
diagnosed by a doctor through a special examination.
Whether the appearance of the irregular heartbeat symbol (    ) in the results 
indicates arrhythmia or not can only be determined by an examination and 
diagnosis by your doctor. 

      Warning: 

If the irregular heartbeat symbol (    ) is shown frequently, please make your doctor 
aware of it. Conducting self-diagnosis and treatment based on measurement 
results is dangerous. Be sure to follow the instructions of your doctor.

Important Safety Information

1.   Overview

Main unit: 

A. Display
B. O/I START button
C. M (Memory) button

D. AC adapter jack 
E. Battery compartment
F. Air jack

Arm Cuffs: 

If air starts to leak from the arm cuff, please replace with a new one.
Refer to “6. Optional Parts”.

O. Medium cuff: arm circumference 22-32 cm
P.  Large cuff: arm circumference 32-42 cm

Display: 

G

Systolic blood pressure

H

Diastolic blood pressure

I

Memory symbol

 

(Displayed when viewing

 

values stored in memory)

J

Irregular heartbeat symbol

K

Heartbeat symbol

 

1

Flashes during measurement

 

2

If flashing after measurement

  complete 

indicates 

blood 

pressure

  out 

of 

recommended 

range

L

Pulse display

M

Deflation symbol

N

.  Battery low symbol

Note: If your systolic or diastolic
          pressure is outside the 
          standard range 
          (above 135/85 mmHg), 
          the Heartbeat symbol (    ) 
          will blink. Refer to 
          “3.2 Taking a Reading”.

2.   Preparation

2.2   Installing/Replacing the Batteries

1.

Turn the main unit upside down.

2.

Slide the battery cover in the direction 
of the arrow while pressing the ribbed 
part of the cover.

3.

Install or replace four “AAA” size 
batteries so that the + (positive) and - 
(negative) polarities match the polarities 
indicated on the battery compartment.

4.

Put the battery cover back in place.
Slide the battery cover as indicated, 
until it clicks into place. 

Note: The measurement 
         values continue to be 
         stored in memory even 
         after the batteries are replaced.

2.3   Battery Life & Replacement

- When the battery low symbol (     ) starts to blink, you
  will still be able to use the unit for a short while. You 
  should replace the batteries with new ones as soon as possible.
- When the symbol (     ) remains lit, the batteries are exhausted. You should replace the 
  batteries with new ones at once. Turn the unit off before replacing the batteries.

Four new “AAA” alkaline batteries will last for approximately 300 measurements, 
when used to take two measurements a day.
Since the supplied batteries are for monitoring use only, they may have 
a shorter life and do not last for 300 measurements.

• Remove the batteries if the unit will not be used for three months or more.
• Dispose of batteries according to applicable local regulations.

If the battery low symbol (     ) appears on the display, 
replace all  four batteries at the same time.

3.   Using the Unit

Be sure to wrap the arm cuff correctly so that you get accurate results.
Measurements can be taken in light clothing. However, please remove 
thick clothes, such as sweaters, before taking a reading.

1.

Insert the air plug into the air jack 
on the left side of the main unit.

3.1   Applying the Arm Cuff

2.

Apply the arm cuff to your upper arm. The air tube should be 
centred on the inside of your arm and point down the inside of 
the arm, so that the air tube runs down the inside of your 
forearm and is in line with your middle finger.

Apply the cuff so that the air tube is 
at the side of your elbow.

Notes:
• Be careful not to rest your arm on 
  the air tube, or otherwise restrict 
  the flow of air to the cuff.
• The cuff should be 1 to 2 cm above 
  the elbow.

Taking a Reading on the Right Arm

3.

When the cuff is positioned correctly, close the fabric fastener firmly.

Note: To cancel a measurement, press the O/I START button to turn the 
           unit off and to release the air in the arm cuff.

1.

Sit on a chair with your feet flat on 
the floor and place your arm on a 
table so that the arm cuff will be 
at the same level as your heart.
Keep still and do not talk during measurement.

2.

Press the O/I START button to turn the unit on.

3.

Check the measurement results.
The unit automatically stores blood 
pressure and pulse rate into its memory.
Refer to “3.4 Using the Memory 
Function”.

Notes:
• Self-diagnosis of measured results and treatment are dangerous. 
  Please follow the instructions of your doctor.
• Wait 2-3 minutes before taking another blood pressure reading. 
  Waiting between readings allows the arteries to return to the condition prior 
  to taking the blood pressure measurement.

1 to 2 cm

Diastolic blood 

pressure

Systolic blood 

pressure

Pulse display

3.2   Taking a Reading

Important:

Important:

• If your systolic or diastolic pressure is outside 
  the standard range, the heartbeat symbol will 
  blink when the measurement result is displayed.
  Recent research suggests that the following 
  values can be used as a guide to high blood 
  pressure for measurements taken at home.

  This criteria is for home blood pressure measurement.

4.

Remove the arm cuff.

5.

Press the O/I START button to turn the unit off.

Note: If you forget to turn the unit off, it will automatically shut itself off after 
          five minutes.

Note: 

If there are no measurements results

        stored in memory, the screen to the right 
        is displayed.

Systolic Blood Pressure

Above 135 mmHg

Diastolic Blood Pressure

Above 85 mmHg

If your systolic or diastolic pressure is
outside the standard range, the heartbeat
symbol will blink when the measurement 
result is displayed. 
Refer to “3.2 Taking a Reading”.

2.

Press the Memory (M) button repeatedly to 
cycle through the previous measurement 
results.
The Memory number appears for a second 
before the pulse rate is displayed. 
The newest set is numbered “1”.
At the end of the measurement, if an irregular heartbeat is 
detected, the irregular heartbeat symbol will be displayed 
with the result.

3.

Press the O/I START button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.

1.

To delete stored readings, first press the Memory (M) button.
Then while holding it down, press the O/I START button 
simultaneously for about 2-3 seconds. All readings will 
then be deleted.

2.

Press the O/I START button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.

If your systolic pressure is known to be more than 170 mmHg, press 
and hold the O/I START button until the arm cuff inflates 30 to 
40 mmHg higher than your suspected systolic pressure.

1.

Press the O/I START button
to turn the unit on.
Measurement starts.

3.3   Instructions for Special Conditions

2.

When the cuff starts to inflate, press the O/I START button 
and keep it pressed until the pressure is 30 to 40 mmHg 
higher than your suspected systolic pressure.

3.

Release the O/I START button when the cuff has been 
inflated to the desired pressure.
The cuff starts to deflate and measurement starts.

Note: You cannot inflate the cuff above 299 mmHg. (An error will be displayed 
          if you try to inflate the cuff above 300 mmHg.)

To Delete All the Values Stored in Memory

You cannot delete individual stored readings, all the
readings in the unit will be deleted.

2

1

• Your blood pressure monitor includes an irregular 
  heartbeat feature. Irregular heartbeats can influence 
  the results of the measurement. The irregular 
  heartbeat algorithm automatically determines if the 
  measurement is usable or needs to be repeated. 
  If the measurement results are affected by irregular 
  heartbeats but the result is valid, the result is shown together with the 
  irregular heartbeat symbol (    ). 
  If the irregular heartbeats cause the measurement to be invalid, 
  no result is shown. If the Irregular heartbeat symbol (    ) is shown 
  after you have taken a measurement, repeat the measurement. 
  If the Irregular heartbeat symbol (    ) is shown frequently, please 
  make your doctor aware of it.

4.   Handling Errors and Problems

4.1   Error Messages

Error Display

Cause

Remedy

Cuff is under inflated.

Carefully read and repeat the 
steps listed under section 3.3.

Movement during 
measurement.

Repeat measurement. Remain 
still and do not talk during 
measurement.

Air plug disconnected.

Insert the air plug securely.
Refer to section 3.1.

Arm cuff not applied 
correctly.

Apply the arm cuff correctly. 
Refer to section 3.1.

Clothing is interfering 
with the arm cuff.

Remove any clothing 
interfering with the arm cuff.

Air is leaking from the 
arm cuff.

Replace cuff with new one.

Arm cuff has been 
inflated above 
299 mmHg.

Do not inflate the arm cuff 
above 299 mmHg.
Refer to section 3.3.

Battery power is low.

Replace all four “AAA” 
batteries with new ones.
Refer to section 2.2.

Device error.

Contact your OMRON retail
outlet or distributor. 

Blinks or appears
continuously 

Arm cuff deflates too
soon. 

The arm cuff is loose.

Apply the cuff correctly 
so that it is firmly 
wrapped around the 
arm. 
Refer to section 3.1.

Cannot measure or
readings are too low
or too high. 

The arm cuff has not 
been inflated
sufficiently.

Inflate the cuff so that it 
is 30 to 40 mmHg 
above your previous 
measurement result.
Refer to section 3.3.

The unit loses power 
during measurement.

The batteries are 
empty.

Replace the batteries 
with new ones.

Nothing happens 
when you press the 
buttons.

The batteries are 
empty.

Replace the batteries 
with new ones.

The batteries have 
been inserted 
incorrectly.

Insert the batteries with 
the correct (+/ -) 
polarity.

Other problems.

Press the O/I START button and repeat 
measurement.
If the problem continues, try replacing the 
batteries with new ones.
If this still does not solve the problem, contact  
your OMRON retail outlet or distributor.

Arm cuff pressure
does not rise. 

The air plug is not 
securely inserted into 
the main unit.

Make sure that the air 
jack is connected
securely. 
Refer to section 3.1.

Air is leaking from the
arm cuff. 

Replace the arm cuff 
with a new one. 

4.2   Troubleshooting

Problem

Cause

Remedy

The reading is
extremely low (or 
high). 

Arm cuff not applied 
correctly.

Apply the arm cuff 
correctly. 
Refer to section 3.1.

Movement or talking 
during measurement.

Remain still and do not 
talk during 
measurement.

Clothing is interfering 
with the arm cuff.

Remove any clothing 
interfering with the arm 
cuff.

5.   Maintenance and Storage

Maintenance

To protect your unit from damage, please observe the following:

• Do not subject the main unit and cuff to extreme 
 temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations 
 (for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit. 
 The unit should be cleaned with a soft, dry cloth.
• Do not wash the arm cuff or immerse it in water.

6.   Optional Parts

Note: Do not bend the air tube excessively.

1.

Unplug the air tube from the air jack.
Gently fold the air tube
into the arm cuff.

Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight, 
 dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at risk of 
 falling.

Medium arm cuff
Arm circumference 
22 - 32 cm

Large arm cuff
Arm circumference 
32 - 42 cm

AC adapter

CM-9997579-7

CL-9970672-9

R Adapter-9997605-0

Product Description
Model
Display
Measurement Method
Measurement Range

Memory
Accuracy

Inflation
Deflation
Pressure Detection
Power Source
Battery Life

Operating Temperature/
Humidity
Storage Temperature/
Humidity/ Air Pressure
Console Weight
Cuff Weight

Outer Dimensions
Cuff Dimensions

Cuff material
Package Content

Digital Automatic Blood Pressure Monitor
OMRON M2 Classic (HEM-7117-ALRU)
LCD Digital Display
Oscillometric method
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180/min.
30 Measurements
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Automatic by electric pump
Automatic pressure release valve
Capacitive pressure sensor
4 “AAA” batteries 1.5V or AC/ DC adapter (6V    4W)
Capacity of new alkaline batteries is approx. 300 
measurements
10°C to 40°C/ 30 to 85% RH 

-20°C to 60°C/ 10 to 95% RH/ 700-1060 hPa

Approximately 289g without batteries
Medium cuff: approximately 120g
Large cuff: approximately 180g
Approximately 114 (w) mm 

×

 84 (h) mm 

×

 129 (l) mm

Medium cuff: approximately 146 mm 

×

 446 mm

(Arm circumference 22 to 32 cm)
Large cuff: approximately 174 mm 

×

 652 mm

(Arm circumference 32 to 42 cm)
Nylon
Main unit, medium cuff, large cuff, instruction manual, 
storage case, battery set, AC adapter, guarantee card, 
blood pressure pass

7.   Technical Data

Note: Subject to technical modification without prior notice
• This OMRON product is produced under the strict quality system of OMRON 
  Healthcare Co. Ltd., Japan. The Core component for OMRON blood pressure
  monitors, which is the Pressure Sensor, is produced in Japan.
• Disposal of this product and used batteries should be carried out in accordance 
  with the national regulations for the disposal of electronic products.

This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive). 
This blood pressure monitor is designed according to the European Standard EN1060, 
Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: 
Supplementary requirements for electromechanical blood pressure measuring systems.

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile 
(cellular) telephones, medical devices in use may be susceptible to 
electromagnetic interference from other devices. Electromagnetic 
interference may result in incorrect operation of the medical device and 
create a potentially unsafe situation.
Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic 
Compatibility) with the aim to prevent unsafe product situations, the 
EN60601-1-2:2007 standard has been implemented. This standard defines 
the levels of immunity to electromagnetic interferences as well as maximum 
levels of electromagnetic emissions for medical devices.

This medical device manufactured by OMRON Healthcare conforms to this 
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate 

strong electrical or electromagnetic fields, near the medical device. This 
may result in incorrect operation of the unit and create a potentially unsafe 
situation. Recommendation is to keep a minimum distance of 7 m. Verify 
correct operation of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at 
OMRON Healthcare Europe at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.

This marking shown on the product or its literature, indicates 
that it should not be disposed of, with other household wastes 
at the end of its working life. To prevent possible harm to the 
environment or human health from uncontrolled waste disposal, 
please separate this from other types of wastes and recycle it 
responsibly to promote the sustainable reuse of material resources.

Household users should contact either the retailer where they purchased 
this product, or their local government office, for details of where and how 
they can take this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and 
conditions of the purchase contract. This product should not be mixed with 
other commercial wastes for disposal. 

This product does not contain any hazardous substances.

Correct Disposal of This Product 
(Waste Electrical & Electronic Equipment)

HEM-7117-ALRU_A_M02_090619.pdf

IM-HEM-7117-ALRU-02-11/09

                    5327736-0B

Made in China

• Do not use petrol, thinners or similar solvents to clean
 the arm cuff.

• Use a soft, moistened cloth and soap to clean the arm cuff.
• Do not carry out repairs of any kind by yourself. 
 If a defect occurs, consult your OMRON distributor.

Calibration and Service

• The accuracy of this blood pressure monitor has been
 carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected every 
  two years to ensure correct functioning and accuracy. 
  Please consult your authorised OMRON dealer.

Storage

Manufacturer

OMRON HEALTHCARE CO., LTD.

24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto,
615-0084, Japan

OMRON HEALTHCARE EUROPE B.V.

Kruisweg 577, 2132 NA Hoofddorp,The Netherlands
www.omron-healthcare.com

ComplectService Ltd.

1/1, Tishinskaya sq, Moscow, Russia,123056
www.csmedica.ru

OMRON DALIAN CO., LTD.

Economic & Technical Development Zone
Dalian 116600, China

EU-representative

Production Facility

Exclusive
distributor in
Russia & importer

Irregular Heartbeat

Irregular Heartbeat

Short

Long

Blood

pressure

Blood

pressure

Pulse

Pulse

= Type B

2.1   Using the AC Adapter

Notes:
• Never plug or unplug the power cord with wet hands.
• Use only the original AC adapter designed for this unit. Use of 
  unsupported adapters may damage the unit.

1.

Insert the AC adapter 
plug into the AC adapter 
jack on the right side of 
the main unit.

2.

Plug the AC adapter 
into an electrical outlet.

To disconnect the AC adapter, unplug the AC adapter from the electrical outlet 
first and then immediately remove the AC adapter plug from the main unit.

Note: When 30 sets of readings are stored in memory, the oldest set will be 
         deleted to store a new set.

This unit has a memory capable of storing 30 sets of readings. Every 
time you complete the measurement, the unit automatically stores the 
blood pressure and pulse rate.

1.

Press the Memory (M) button.
The result from the most recent measurement is displayed.

4.

The rest of the procedure is the same as for normal 
measurement. Refer to “3.2 Taking a Reading”, steps 3 to 5.

Note: Do not apply more pressure than necessary.

3.4   Using the Memory Function

Note: You can take a measurement on either your left or right arm. The blood 

pressure can differ between the right arm and the left arm and therefore 
also the measured blood pressure values can be different. Omron 
recommends to always use the same arm for measurement. If the 
values between the two arms differ substantially, please check with your 
physician which arm to use for your measurement.

Please read the instruction manual carefully before 
using the device.

Medium cuff

Large cuff

Q. Air Tube