Thank you for purchasing the OMRON JPN600 Automatic Blood Pressure 

Monitor.
The OMRON JPN600 is a compact, fully automatic blood pressure monitor, 

operating on the oscillometric principle. It measures your blood pressure and 

pulse rate simply and quickly. For comfortable controlled inflation without the 

need of pressure pre-setting or re-inflation the device uses its advanced 

“IntelliSense” technology.
Intended Use

This product is designed to measure the blood pressure and pulse rate of 

people within the range of the designated arm cuff, following the instructions in 

this instruction manual.

It is mainly designed for general household use. Please read the Important 

Safety Information in this instruction manual before using the unit.

Please follow this instruction manual thoroughly.

Please keep for future reference.

For specific information about your own blood pressure, CONSULT 

YOUR DOCTOR.

Please read this section carefully before using the unit.

Warning: 

Indicates a potentially hazardous situation which, if not avoided, 

could result in death or serious injury.

(General Usage)

• DO NOT adjust medication based on measurement results from this blood

pressure monitor. Take medication as prescribed by your doctor. Only a

doctor is qualified to diagnose and treat High Blood Pressure.

• Consult your doctor before using the device for any of the following

conditions: common arrhythmias such as atrial or ventricular premature beats

or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age,

pregnancy, pre-eclampsia, renal diseases.

Note that PATIENT motion, trembling, shivering may affect the measurement

reading.

• Do not use the device on the injured arm or the arm under medical treatment.

• Do not apply the arm cuff on the arm while being on an intravenous drip or

blood transfusion.

• Consult your doctor before using the device on the arm with an arterio

-

venous (A-V) shunt.

• Do not use the device with other medical electrical (ME) equipment simulta

-

neously.

• Do not use the device in the area the HF surgical equipment, MRI, or CT

scanner exists, or in the oxygen rich environment.

• The air tube or the AC adapter cable may cause accidental strangulation in

infants.

• Contains small parts that may cause a choking hazard if swallowed by

infants.

• Stop using the unit and consult your doctor if you experience skin irritation or

other troubles.

(AC Adapter (optional) Usage)

• Do not use the AC adapter if the unit or the power cord is damaged. Turn off

the power and unplug the power cord immediately.

• Plug the AC adapter into the appropriate voltage outlet. Do not use a

multiple-tap.

• Never plug in or unplug the power cord from the electric outlet with wet

hands.

Caution:

 Indicates a potentially hazardous situation which, if not avoided,

may result in minor or moderate injury to the user or patient or 

damage to the equipment or other property.

(General Usage)

Always consult your doctor. Self-diagnosis of measurement results and

self-treatment are dangerous.

Consult your doctor before using the device for any of the following

conditions:

- If you have had a mastectomy.

- People with severe blood flow problems or blood disorders as cuff inflation

can cause bruising.

Do not take measurements more than necessary. It may cause bruising due

to blood flow interference.

If there are any abnormalities during the measurement, remove the arm cuff.

Do not use this device on infants or persons who cannot express their

intentions.

Do not inflate the arm cuff more than necessary.

Do not use the unit for any purpose other than measuring blood pressure.

Use only the approved arm cuff for this device. Use of other arm cuffs may

result in incorrect measurement results.

Do not use a mobile phone or other devices that emit electromagnetic fields,

near the device. This may result in incorrect operation of the device.

Do not disassemble the monitor or arm cuff. This may cause an inaccurate

reading.

Do not use in a location with moisture, or a location where water may splash 

on the device. This may damage the device.

Do not use the device in a moving vehicle (car, airplane).

(Battery Usage)

• Do not insert the batteries with their polarities incorrectly aligned.

• Use only four “AA” alkaline or manganese batteries with this unit. Do not use

other types of batteries. Do not use new and used batteries together.

• Remove the batteries if the unit will not be used for 3 months or more.

(AC Adapter (optional) Usage) 

• Fully insert the power plug.

• When disconnecting the power plug, do not pull the power cord. Be sure to

hold the power plug.

Important Safety Information

This marking shown on the product or its literature, indicates that it 

should not be disposed of, with other household wastes at the end 

of its working life. To prevent possible harm to the environment or 

human health from uncontrolled waste disposal, please separate 

this product from other types of wastes and recycle it responsibly to 

promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this 

product, or their local government office, for details of where and how they can 

return this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions 

of the purchase contract. This product should not be mixed with other commercial 

wastes for disposal. 

Correct Disposal of This Product 

(Waste Electrical & Electronic Equipment)

Made in Japan

Manufacturer

Introduction

1. Overview

Main unit: 

N. Average value symbol

O. Memory symbol

P. Irregular heartbeat symbol

Q. Systolic blood pressure

R. Diastolic blood pressure

S. Low battery symbol

T. Cuff wrapping guide

2.1   Installing/Replacing the Batteries

2. Preparation

1.
2.

Remove the battery cover.

2.2   Setting the Date and Time

1.

Press the      button.

2.

Set the monitor to the correct date and time before

taking a measurement for the first time.

3.

Press the START/STOP button to store the setting.

Notes:

• If the batteries have been replaced, the date and time setting will

need to be reset.

• If the date and time are not set, “-:--” appears during or after

measurement.

Insert four “AA” batteries 

as indicated in the battery 

compartment and then 

replace the battery cover.

Notes:

• If the

 low battery symbol (     ) appears on the display, turn the

monitor off then replace all batteries at the same time.

• The measurement values continue to be stored in memory even after

the batteries are replaced.

• The supplied batteries may have a shorter life.

•  Dispose of the device, components and optional accessories

according to applicable local regulations. Unlawful disposal may

cause environmental pollution.

The bottom edge of the arm 

cuff should be 1 to 2 cm 

above the elbow. Marker 

(arrow under the air tube) is 

centred on the middle of 

your inner arm.

3.

3.3   Taking a Reading

6. Technical Data

4.3   Maintenance

Calibration and Service

To protect your unit from damage, please observe the following:

Notes:

• When you take a measurement on the right

arm, the air tube will be at the side of your

elbow. Be careful not to rest your arm on

the air tube.

Notes:

• To cancel a measurement, press the START/STOP button to release

the air in the arm cuff.

• Remain still while taking a measurement.

Press the START/STOP button.

The arm cuff will start to inflate automatically.

• Do not subject the main unit and the cuff to extreme

temperatures, humidity, moisture or direct sunlight.

• Do not fold the cuff or tubing tightly.

• Do not disassemble the unit.

• Do not subject the unit to strong shocks or vibrations

(for example, dropping the unit on the floor).

• Do not use volatile liquids to clean the main unit.

• Do not wash the arm cuff or immerse it in water.

• Do not use petrol, thinners or similar solvents to clean

the arm cuff.

• Do not carry out repairs of any kind yourself. If a defect

occurs, consult your local OMRON representative.

• The accuracy of this blood pressure monitor has been

carefully tested and is designed for a long service life.

• It is generally recommended to have the unit inspected

every two years to ensure correct functioning and

accuracy. Please consult your local OMRON

representative.

• The unit should be cleaned with a soft, dry cloth.

• Use a soft, moistened cloth and neutral soap to clean

the arm cuff.

Automatic Blood Pressure Monitor

JPN600

LCD Digital Display

Oscillometric method

Pressure: 0 to 299 mmHg 

Pulse: 40 to 180 beats/ min.

Pressure: ±3 mmHg

Pulse: ±5% of display reading

Fuzzy-logic controlled by electric pump

Automatic pressure release valve

90 measurements with date and time

IP 20

DC6V   4W

4 “AA” batteries 1.5V 

or AC adapter (optional, INPUT AC100-240V 

50/60Hz 0.12A)

Approx. 1000 measurements 

(using new alkaline batteries)

Main Unit: 5 years

Cuff: 

1 year (when used 6 times a day)

           = Type BF           

Internally powered ME equipment (When using only 

the batteries)

       = Class II ME equipment (Optional AC adapter)

+10 to +40°C / 30 to 85% RH / 700 to 1060 hPa

-20 to +60°C / 10 to 95% RH / 700 to 1060 hPa

Approx. 280g without batteries

Approx. 130g

Approx. 107 (w) mm 

× 

79 (h) mm × 141(l) mm

Approx. 145 mm × 466 mm 

(Cuff: arm circumference 22 to 32 cm)

Nylon, polyester, polyvinyl chloride

Main unit, arm cuff, instruction manual, battery set, 

storage case, EMC information

3.2   How to Sit Correctly

To take a measurement, you need to be 

relaxed and comfortably seated, under 

comfortable room temperature. Avoid 

bathing, drinking alcohol or caffeine, 

smoking, exercising or eating 30 minutes 

before taking a measurement. Also avoid 

taking measurements during stressful 

time. Stress raises blood pressure.

• Sit on a chair with your legs uncrossed and

your feet flat on the floor.

• Sit upright with your back straight.

• Sit with your back and arm being supported.

• The arm cuff should be placed on your arm

at the same level as your heart.

• The blood pressure can differ between the right arm and the left arm,

and therefore also the measured blood pressure values can be

different. OMRON recommends to always use the same arm for

measurement. If the values between both arms differ substantially,

please check with your doctor which arm to use for your

measurement.

1.

Press and hold the START/STOP 

button for more than 3 seconds.

The date/time display will disappear.

1.

Release the START/STOP button when the date/time 

display turns off.

The arm cuff will start to inflate automatically.

2.

The monitor stores measurement values for one user in the memory. 

The guest mode can be used to take a single measurement for 

another user. No measurement values are stored in the memory when 

the guest mode is selected.

START

INFLATE

DEFLATE

COMPLETE

1.

Unplug the air plug from the air jack.

• Keep the unit in its storage case when not in use.

• Store the device and the components in a clean, safe location.

2.

Gently fold the air tube into the arm cuff.

4.4   Storage

Product Description

Model

Display

Measurement Method

Measurement Range

Accuracy

Inflation

Deflation

Memory

IP classification

Rating 

Power Source

Battery Life

Durable Period 

Applied Part

Protection Against 

Electric Shock

Operating 

temperature/ Humidity 

/ Air Pressure

Storage temperature/ 

Humidity/ Air pressure

Console Weight

Cuff Weight

Outer Dimensions

Cuff Dimensions

Cuff/ Tube Material

Package Contents

• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device

Directive).

This blood pressure monitor is designed according to the European Standard

EN1060, Non-invasive sphygmomanometers Part 1: General Requirements

and Part 3: Supplementary requirements for electromechanical blood

pressure measuring systems.

This OMRON product is produced under the strict quality system of OMRON

HEALTHCARE Co., Ltd., Japan. The core component for OMRON blood

pressure monitors, which is the Pressure Sensor, is produced in Japan.

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, KYOTO,

617-0002 JAPAN

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS

Production facility OMRON HEALTHCARE Co., Ltd.

Matsusaka Factory

Mie, JAPAN

Asia Pacific HQ

OMRON HEALTHCARE SINGAPORE PTE LTD.

438A Alexandra Road, #05-05/08,

Alexandra Technopark, Singapore 119967

www.omronhealthcare-ap.com

Automatic Blood 

Pressure Monitor

Model JPN600

Instruction Manual

A. Display

B. Memory button

C. START/STOP button

D. Date/Time setting (     ) button

E. Cuff wrapping guide lamp

F. Up/Down (         ) buttons

G. Battery compartment

H. AC adapter jack (for optional

AC adapter)

I. Air jack

U. Heartbeat symbol

• Flashes during measurement.

V. Date/Time display

W. Movement error symbol

X. Blood pressure level indicator

Y. Deflation symbol

Z. Pulse display and Memory

number

Push     or     buttons to 

change.
Push the      button to confirm.

Next setting appears.

1)

2)

If your systolic pressure is more than 210 mmHg

After the arm cuff starts to inflate, press and hold the START/STOP 

button until the monitor inflates 30 to 40 mmHg higher than your 

expected systolic pressure.
Notes:

• The monitor will not inflate above 299 mmHg.

• Do not apply more pressure than necessary.

Cuff Wrapping Guide Lamp

• If you move during measurement, the movement

error symbol (    ) will appear on the display.

Keep still and repeat the measurement.

Display: 

3100085-2A

J. Arm cuff (Arm circumference

22 - 

32 cm)

K. Air plug

L. Air tube

M. Marker

Arm cuff: 

• Your blood pressure monitor includes an

irregular heartbeat feature. Irregular heartbeats

can influence the results of the measurement.

The irregular heartbeat algorithm automatically

determines if the measurement is usable or

needs to be repeated. If the measurement

3. Using the Unit

1.

Insert the air plug into the air jack

securely.

3.1   Applying the Arm Cuff

Remove tight-fitting clothing or tight rolled up sleeve from your 

upper arm. Do not place the arm cuff over thick clothes.

2.

Remove the arm cuff.

3.

Press the START/STOP button to turn the monitor off.

The monitor automatically stores the measurement in its memory.

It will automatically turn off after 2 minutes.

Note:  Wait 2-3 minutes before taking another blood pressure 

measurement. Waiting between readings allows the arteries to 

return to the condition prior to taking the blood pressure 

measurement.

2.

Put your arm through the cuff

loop.

Close the fabric fastener 

FIRMLY.

Using the Guest Mode

Cuff Wrapping Guide Lamp:

Note:  Do not bend the air tube excessively.

3.

Place the arm cuff and main unit in

the storage case.

Do not store the unit in the following situations:

• If the unit is wet.

• Locations exposed to extreme temperatures,

humidity, direct sunlight, dust or corrosive 

vapors such as bleach.

• Locations exposed to vibrations, shocks or

where it will be at an angle.

Systolic Blood Pressure 

≥ 135

 mmHg

Diastolic Blood Pressure 

 85 mmHg

Important:

This criteria is for home blood pressure measurement.

• Recent research suggests that the following

values can be used as a guide to high blood

pressure for measurements taken at home.

results are affected by irregular heartbeats but the result is valid, the

result is shown together with the irregular heartbeat symbol (    ). If

the irregular heartbeats cause the measurement to be invalid, no

result is shown. If the irregular heartbeat symbol (    ) is shown after

you have taken a measurement, repeat the measurement. If the

irregular heartbeat symbol (    ) is shown frequently, please make

your doctor aware of it.

Note: Read and follow the “Correct Disposal of This Product” in the 

Technical Data Section when disposing of the device and any 

used accessories or optional parts.

The cuff wrapping guide lamp       lights in green, the cuff is correctly 

wrapped tightly enough on the arm and the reading is accurate and 

reliable. If it does not light, the reading is not reliable due to the 

incorrect wrapping of the cuff. Please wrap the cuff again, taking care 

to wrap it correctly and take the measurement again.

3.4   Using the Memory Function

Notes:

• If the previous reading was taken without setting

the date and time, the average value is not

calculated.

• If there are no measurements results stored in

the memory, the screen to the right is displayed.

The monitor automatically stores the result up to 90 sets. It can also 

calculate an average reading based on the measurements from the 

last three readings taken within 10 minutes. If there are only two 

readings in memory for that period, the average will be based on two 

readings. If there is one reading in memory for that period, the average 

will be based on one reading.
Notes:

• If the memory is full, the monitor will delete the oldest reading.

• When viewing the reading taken without setting the date and time,

“-:--” is displayed instead of the date and time.

Alternating date/time display

Press the     button.

The Memory number appears for a second before the pulse rate is 

displayed. The newest set is numbered “1”.

1.

To View the Readings Stored in Memory

4. Troubleshooting and Maintenance

4.1   Error Messages

4.2   Troubleshooting

Problem

Cause

Remedy

The reading is 

extremely low (or 

high).

Arm cuff not applied 

correctly.

Movement or talking 

during measurement.

Clothing is interfering 

with the arm cuff.

Apply the arm cuff 

correctly. 

Refer to section 3.1.

Remain still and do not 

talk during 

measurement.

Refer to section 3.3.

Remove any clothing 

interfering with the arm 

cuff.

Refer to section 3.1.

The batteries are low.

The batteries are 

exhausted. 

Air plug disconnected.

Movement during 

measurement and the arm 

cuff has not been inflated 

sufficiently.

The arm cuff was inflated 

above 299 mmHg 

when inflating the cuff 

manually.

Movement during 

measurement.

Clothing is interfering with 

the arm cuff.

Device error.

Irregular heartbeats are 

detected.

Remove the arm cuff. Wait 

2 - 3 minutes and then take 

another measurement. 

Repeat the steps in section 

3.3. If this error continues to 

appear, contact your doctor.

Movement during 

measurement.

Carefully read and repeat 

the steps in section 3.3.

You should replace them 

with new ones ahead of 

time.

Refer to section 2.1.
You should replace them 

with new ones at once.

Refer to section 2.1.

Insert the plug securely.

Refer to section 3.1.

Arm cuff not applied 

correctly.

Apply the arm cuff correctly.

Refer to section 3.1.

Air is leaking from the arm 

cuff.

Replace the cuff with the 

new one.

Refer to Chapter 5.

Repeat measurement. 

Remain still and do not talk 

during measurement.

Refer to section 3.3.

If “E2” appears repeatedly, 

inflate the cuff manually 

until it is 30 to 40 mmHg 

above your previous 

measurement result.

Refer to section 3.3.

Do not inflate the cuff above 

299 mmHg.

Refer to section 3.3.

Repeat measurement. 

Remain still and do not talk 

during measurement.

Refer to section 3.3.

Remove any clothing 

interfering with the arm cuff.

Refer to section 3.1.

Contact your local OMRON 

representative.

Cause

Remedy

Error Display

Problem

Cause

Remedy

Cannot measure or 

readings are too low 

or too high.

The arm cuff has not 

been inflated 

sufficiently.

Inflate the cuff so that it 

is 30 to 40 mmHg above 

your previous 

measurement result.

Refer to section 3.3.

• Press the START/STOP button and repeat

measurement.

• If the problem continues, try replacing the

batteries with new ones.

If this still does not solve the problem, contact 

your local OMRON representative.

Nothing happens 

when you press the 

buttons.

Other problems.

The batteries are 

empty.

The batteries have 

been inserted 

incorrectly.

Replace the batteries 

with new ones.

Refer to section 2.1.
Insert the batteries with 

the correct (+/-) polarity.

Refer to section 2.1.

Note:  The cuff wrapping guide result appears 

on the display with the measurement 

values. The cuff wrapping guide lamp will 

not light.

: To the older readings
: To the more recent readings

: The arm cuff was correctly wrapped tightly enough.
: The arm cuff was loose or not wrapped correctly.

Press the     or     button to view the readings stored in

memory.

2.

To Delete All the Values Stored in Memory

Note:  You cannot partially delete the stored readings.

When the memory symbol (    ) appears, first press the Memory 

button. Then while holding it down, press the START/STOP 

button simultaneously for more than 3 seconds.

Always consult your doctor. Self-diagnosis of measurement 

results and self-treatment are dangerous.

5. Optional Parts

Notes:

• Please use Adapter [60110HW5SW] in Taiwan, Adapter

[6V06ASW-KA] in Korea.

• Please check with your local OMRON representatives for the

appropriate optional parts.

Notes:

• Subject to technical modification without prior notice.

• This device is clinically investigated according to the requirements of ISO

81060-2:2013.

• IP classification is degrees of protection provided by enclosures in

accordance with IEC 60529. This device is protected against solid foreign

objects of diameter 12 mm such as a finger and greater.

1.

2.

Using the Optional AC Adapter

Insert the AC adapter plug 

into the AC adapter jack on 

the rear side of the main 

unit.
Plug the AC adapter into an 

electrical outlet.

To disconnect the AC adapter, unplug the AC adapter from the 

electrical outlet first and then remove the AC adapter plug from the 

main unit.

AC Adapter S

Medium Cuff 

HEM-CR24

Arm circumference 

22 - 32 cm

Small Cuff

HEM-CS24

Arm circumference

17 - 22 cm

Wide Range Soft Cuff 

HEM-RML31

Arm circumference 

22 - 42 cm

Cuff is not applied correctly. Apply the arm cuff correctly. 

Refer to section 3.1.

• When handling the power cord, observe the following:

Do not damage.  

Do not break it.

Do not tamper with it. 

Do not forcibly bend or pull.

Do not twist. 

 

Do not bundle during use.

Do not pinch.   

Do not place under heavy objects.

• Wipe the dust off from the power plug.

• Disconnect the power pl

ug if the product will not be used for a long period of

time.

• Disconnect the power plug before starting maintenance.

• Use only the original AC adapter designed for this unit. Use of unsupported

adapters may damage and/or may be hazardous to the unit.

General Precautions

• Do not forcibly bend the arm cuff or bend the air tube excessively.

• To unplug the air plug, pull on the air plug at the connection with the main

unit, not the tube itself.

• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.

• Do not inflate the arm cuff when it is not wrapped around your arm.

• Read and follow the “Important information regarding Electro Magnetic

Compatibility (EMC)” in the EMC information provided with this unit.

• Do not use the device outside the specified environment. It may cause an

inaccurate reading.

• Dispose of the device, components and optional accessories according to

applicable local regulations. Unlawful disposal may cause environmental

pollution.

Press and hold the     button for 

more than 3 seconds.

1.

To View the Average Value

Arm cuff deflates too 

soon.

Arm cuff pressure 

does not rise.

The air tube is not 

securely connected 

into the main unit.

Air is leaking from the 

arm cuff.

The arm cuff is loose.

Make sure that the air 

tube is connected 

securely. 

Refer to section 3.1.

Replace the arm cuff 

with a new one. 

Refer to Chapter 5.
Apply the cuff correctly 

so that it is firmly 

wrapped around the arm. 

Refer to section 3.1.

HEM-7131-AP3_A_M03_151208.pdf