background image

Automatic Blood Pressure Monitor

with ComFit

TM

Cuff

Model

HEM-790ITCAN

INSTRUCTION MANUAL

ENGLISH 

  

  

 

  

 

Содержание IntelliSense HEM-790ITCAN

Страница 1: ...Automatic Blood Pressure Monitor with ComFitTM Cuff Model HEM 790ITCAN INSTRUCTION MANUAL ENGLISH ...

Страница 2: ...tructions Know Your Unit 9 Unit Display 11 Display Symbols 12 Irregular Heartbeat Symbol 12 Movement Error Symbol 12 USER ID Symbol 12 Morning Hypertension Symbol 13 Heartbeat Symbol 14 Average Value Symbol 14 Morning Average Symbol 14 Evening Average Symbol 14 Battery Installation 15 Using the AC Adapter 17 Setting the Date and Time 20 Applying the Arm Cuff 24 TABLE OF CONTENTS ...

Страница 3: ...orning and Evening Averages 41 Morning Averages 41 Evening Averages 41 To Display Morning and Evening Averages 42 Display Combinations 43 To Delete All Values Stored in the Memory 44 Downloading Instructions Installing the Software 45 Using the Software 56 Transferring Data to a Personal Computer 58 Software Trouble Shooting Tips 60 Care and Maintenance Care and Maintenance 61 Error Indicators and...

Страница 4: ... into a digital reading The monitor stores measurement results for two people and features Morning and Evening Averages TruReadTM allows for three consecutive measurements with timed intervals for an average reading Omron Health Management Software included with this unit allows you to view process and print data about blood pressure measured by the HEM 790ITCAN The HEM 790ITCAN comes with the fol...

Страница 5: ... other property OPERATING THE DEVICE Contact your physician for specific information about your blood pressure Self diagnosis and treatment using measured results may be dangerous Follow the instructions of your physician or licensed healthcare provider If battery fluid should get in your eyes immediately rinse with plenty of clean water Contact a physician immediately Read all of the information ...

Страница 6: ...able local regulations Unlawful disposal may cause environmental pollution Do not use a cellular phone near the device It may result in an operational failure Use only Omron authorized parts and accessories Parts and accessories not approved for use with the device may damage the unit Use only 1 5V manganese or alkaline batteries with this device Do not use other types of batteries This may damage...

Страница 7: ...E AND MAINTENANCE Do not subject the monitor to strong shocks such as dropping the unit on the floor Do not submerge the device or any of the components in water Store the device and the components in a clean safe location Changes or modification not approved by Omron Healthcare will void the user warranty Do not disassemble or attempt to repair the unit or components SAFETY INFORMATION 7 ...

Страница 8: ...ken in a quiet place 4 Remove tight fitting clothing from your arm 5 Sit in a chair with your feet flat on the floor Rest your arm on a table so that the cuff is at the same level as your heart 6 Remain still and do not talk during the measurement 7 Keep a record of your blood pressure and pulse readings for your physician A single measurement may not provide an accurate indication of your true bl...

Страница 9: ...9 KNOW YOUR UNIT Main Unit Air Jack Battery Compartment Start Stop Button USER ID Selection Switch Morning Evening Average Buttons Memory Button SET Button Display AC Adapter Jack USB Port ...

Страница 10: ...f Air Tube Air Plug Connects to Air Jack AC Adapter Storage Case USB Cable CD ROM Four AA Batteries Instruction Manual Housing Cord Automatic Blood Pressure Monitor w th ComFit M Cuff Model HEM 790ITCAN INSTRUCTION MANUAL FRANÇA S ENGL SH ...

Страница 11: ...bol Pulse Display Date Time Display Low Battery Indicator Deflation Symbol Morning Average Symbol Heartbeat Symbol Irregular Heartbeat Symbol Evening Average Symbol Average Value Symbol Memory Symbol USER ID Symbol A or B Morning Hypertension Symbol Week Display TruReadTM Symbol ...

Страница 12: ...at symbol displays with your measurement results we recommend you consult your physician Follow the directions of your doctor MOVEMENT ERROR SYMBOL The Movement Error Symbol is displayed if you move your body during the measurement Please remove the arm cuff and wait 2 3 minutes Take another measurement USER ID SYMBOL The USER ID Symbol indicates USER A or USER B as selected by the user to take a ...

Страница 13: ...or specific information about your blood pressure Self diagnosis and treatment using measured results may be dangerous Follow the instructions of your physician or licensed healthcare provider Omron Healthcare recommends you contact your physician if the MORNING HYPERTENSION SYMBOL and or the HEARTBEAT SYMBOL appear on the display with your measurement values MORNING HYPERTENSION SYMBOL The Mornin...

Страница 14: ... FUNCTION The Heartbeat Symbol flashes when the blood pressure reading stored in the memory is above 135 for the Systolic Blood Pressure value and or 85 for the Diastolic Blood Pressure value AVERAGE VALUE SYMBOL The Average Value Symbol is displayed when you press the memory button The most recent average reading appears on the display screen MORNING AVERAGE SYMBOL The Morning Average symbol is d...

Страница 15: ...he cover off in the direction of the arrow 2 Install 4 AA size batteries so the positive and negative polarities match the polarities of the battery compartment as indicated 3 Replace the battery cover BATTERY INSTALLATION 15 Å Å Å Å Å Å Å Å Battery Cover ...

Страница 16: ...the batteries If the batteries are removed while the unit is still on the date and time will be reset to that of the previous use However the measurement results are not deleted When the batteries are replaced you may need to reset the date and time If the year is flashing on the display screen refer to page 22 RESETTING THE DATE AND TIME The supplied batteries may have a shorter life CAUTION Disp...

Страница 17: ...ter is used If no batteries are installed you may need to reset the date and time if the AC Adapter is disconnected The measurement results will not be deleted CAUTION Use only Omron authorized parts and accessories Parts and accessories not approved for use with the device may damage the unit USING THE AC ADAPTER 17 ...

Страница 18: ...tended to be correctly oriented in a vertical or floor mount position NOTES Use only the authorized Omron AC Adapter that came with this monitor We recommend you install the batteries even if the AC Adapter is used If no batteries are installed you may need to reset the date and time if the AC Adapter is disconnected The measurement results will not be deleted CAUTION Use only authorized parts and...

Страница 19: ...60 cycles TO DISCONNECT THE AC ADAPTER Hold the Housing to remove the AC Adapter from the electrical outlet Remove the AC Adapter plug from the monitor WARNING Do not plug or unplug the power cord into the electrical outlet with wet hands CAUTION Do not overload power outlets Plug the device into the appropriate voltage outlet USING THE AC ADAPTER 19 1 2 Housing Cord ...

Страница 20: ... morning and in the evening Morning and Evening averages are stored for this week and the last seven weeks The monitor can also calculate an average reading based on the three most recent measurements taken within 10 minutes of the last reading NOTE If only two readings are stored in the memory for the 10 minute period the average is based on the two readings If one reading is stored in the memory...

Страница 21: ...button to advance by increments of one year Press the SET button to set the current year The month flashes on the display NOTE Press and hold the Memory button to advance the date and time values faster 3 SETTING THE MONTH Press the Memory button to advance by increments of one month Press the SET button to set the current month The day flashes on the display ...

Страница 22: ...on to advance by increments of one hour Press the SET button to set the current hour The minute flashes on the display 6 SETTING THE MINUTE Press the Memory button to advance by increments of one minute Press the SET button to set the minute setting 7 TO TURN THE MONITOR OFF Press the START STOP button NOTE The monitor will automatically turn off after the minute has been set after battery replace...

Страница 23: ...ay for the USER ID selected To change the on or oFF setting press the Memory button then press the SET button The interval time flashes on the display Press the Memory button to select the desired interval time then press the SET button The year flashes on the display 2 Refer to page 19 SETTING THE DATE AND TIME NOTE If you need to change the Date and Time always adjust the Date and Time setting b...

Страница 24: ...is intended for use in measuring blood pressure and pulse rate in the adult population Do not use this device on infants or persons who cannot express their intentions APPLYING THE CUFF ON THE LEFT ARM 1 Make sure the air plug is securely inserted in the main unit 2 Remove tight fitting clothing from your upper left arm APPLYING THE ARM CUFF 24 ...

Страница 25: ...or Place your left arm on a table so the cuff is level with your heart 4 Hold the grip on the cuff securely with your right hand placing your thumb on the thumb grip 5 Turn the palm of your left hand upward APPLYING THE ARM CUFF Thumb Grip Cloth Strip ...

Страница 26: ...arm and aligned with your middle finger The air tube runs down the inside of your arm The bottom of the cuff should be approximately 1 2 above your elbow 7 Wrap the cuff firmly in place around your arm using the cloth strip 26 APPLYING THE ARM CUFF 1 2 inch 1 2 cm Air Tube Blue Strip ...

Страница 27: ...on page 25 Apply the cuff to your right upper arm so the thumb grip is centered on the inside of your inner arm The bottom of the cuff should be approximately 1 2 above your elbow NOTE Be careful not to rest your arm on the air tube This will restrict the flow of air to the cuff APPLYING THE ARM CUFF 27 Thumb Grip ...

Страница 28: ...surements taken using the GUEST Mode are not stored in the memory CAUTION Operate the device only as intended Do not use the device for any other purpose USING THE GUEST MODE 1 Press and hold the START STOP button The A and B Symbols disappear from the display 2 Release the START STOP button The cuff starts to inflate automatically NOTE To stop the inflation press the START STOP button The monitor...

Страница 29: ...s and Evening Averages Slide the USER ID Selection Switch up or down to select USER A or B USING THE USER ID When you select a USER ID the monitor can be set to take one measurement using the Single Mode or three consecutive measurements using the TruReadTM Mode when you press the START STOP button NOTE The default value is set to the Single Mode to take one measurement The TruReadTM Mode takes th...

Страница 30: ...your USER ID A or B 2 Press and hold the SET button The TruReadTM symbol and the oFF setting appears on the display 3 Press the Memory button to select on NOTE Press the Memory button to alternate between on and oFF TAKING A MEASUREMENT 30 ...

Страница 31: ... 60 seconds 5 Press the MEMORY button to change the interval of time between each measurement The interval can be set to 15 seconds 30 seconds 60 seconds or 120 seconds Continue to press the MEMORY button until the desired interval appears on the display 6 Press the SET button to confirm the setting The year flashes on the display TAKING A MEASUREMENT 31 ...

Страница 32: ...n to turn the monitor off To change the date and time refer to page 19 SETTING THE DATE AND TIME CAUTION Do not use a cellular phone near the device It may result in an operational failure USING THE SINGLE MODE 1 Select your USER ID A or B TAKING A MEASUREMENT 32 ...

Страница 33: ...r detects the pulse during inflation Do not move your arm and remain still until the entire measurement process is completed NOTE To stop the inflation or measurement press the START STOP button The monitor will stop inflating start deflating and will turn off 3 Inflation stops automatically and the measurement is started As the cuff deflates decreasing numbers appear on the display The Heartbeat ...

Страница 34: ...plete the arm cuf f completely deflates Your blood pressure and pulse rate are displayed 5 Press the START STOP button to turn the monitor off NOTE The monitor will automatically turn off after five minutes TAKING A MEASUREMENT 34 ...

Страница 35: ... cuff deflates decreasing numbers appear on the display The Heartbeat Symbol flashes at every heartbeat When the measurement is complete the arm cuff completely deflates NOTE Do not move your arm and remain still until the entire measurement process is completed 4 Check the progress of the consecutive measurements NOTE If a measurement fails the monitor will repeat taking a measurement up to five ...

Страница 36: ... are complete The average for three measurements appears on the display 5 Press the SET button to display the blood pressure and pulse rate for the first measurement Press the SET button again to display the next measurement 6 Press the START STOP button to turn the monitor off TAKING A MEASUREMENT 36 ...

Страница 37: ...ssure is known to be more than 220 mmHg press and hold the START STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure NOTE Do not apply more pressure than necessary The monitor will not inflate above 299 mmHg When the measurement is complete the arm cuf f completely deflates Your blood pressure and pulse rate are displayed ...

Страница 38: ...ave the most recent values The monitor also stores 8 weeks of Morning Averages and 8 weeks of Evening Averages for each user A and B AVERAGING FUNCTION The monitor calculates the average reading based on the three most recent sets of measurement values taken within 10 minutes of the most recent reading NOTE If two sets of measurement values are stored in the memory for the 10 minutes period the av...

Страница 39: ...TO DISPLAY THE MEASUREMENT VALUES 1 Select your USER ID A or B 2 Press the Memory button The Average Symbol appears on the display when the average reading is shown USING THE MEMORY FUNCTION 39 ...

Страница 40: ...ess the Memory button to display the next set of values Press and hold the Memory button to display the values faster TruReadTM Mode The average for the consecutive measurements appears on the display with the TruReadTM symbol To display the values for the individual measurements press the SET button while the average is displayed 4 Press the START STOP button to turn the power off USING THE MEMOR...

Страница 41: ...eek begins Sunday at 4 00 AM MORNING AVERAGES Morning averages are based on the first average reading using the TruReadTM Mode or the first reading using the Single Mode Measurement times must be between 4 00 AM and 9 59 AM EVENING AVERAGES Evening averages are based on the last average reading using the TruReadTM Mode or the last reading using the Single Mode Measurement times must be between 7 0...

Страница 42: ...s on the display 3 Continue to press the Morning Average button or the Evening Average button to display the previous weeks The monitor displays 1 WEEK for the previous week to 7 WEEK for the oldest set of averages NOTE If no measurement is taken in a given week the display will appear as illustrated 4 Press the START STOP button to turn the monitor off USING THE MEMORY FUNCTION 42 ...

Страница 43: ... be displayed in the following combinations Display Average above CHS guidelines Morning Hypertension Average within CHS guidelines Morning Average Evening Average Morning average from This Week with Morning Average Symbol Morning average from 7 weeks ago with Morning Average Symbol blinking Heartbeat Symbol Morning Hypertension Symbol Evening average from This Week with Evening Average Symbol Eve...

Страница 44: ...he values for the Evening Average TO DELETE ALL VALUES STORED IN THE MEMORY The values stored in the memory are deleted by USER ID You cannot partially delete values stored in the memory All values for the user you select will be deleted 1 Select your USER ID A or B 2 While holding the Memory button and the START STOP button simultaneously for more than 2 seconds all values will be deleted USING T...

Страница 45: ...ometer NOTE This software is exclusive to Omron s blood pressure monitors and pedometers Visit http software omronhealthcare com to register your customer information for online support CAUTION This is a CD ROM disc Do not play this disc on general audio players DVD players or personal game devices It may damage the speaker of the device and may harm your eardrums INSTALLING THE SOFTWARE 45 ...

Страница 46: ...32MB or more Display Super VGA 800 x 600 pixel or higher resolution video adapter and monitor Others A CD ROM or DVD drive USB connector USB Revision 1 1 compliant A printer Microsoft Internet Explorer 5 0 or higher Adobe Acrobat Reader 5 0 or higher Trademark and Registered trademark Microsoft and Windows are registered trademarks in the United States and other countries Adobe Adobe logo Acrobat ...

Страница 47: ... Management Software You can ignore the C of the C1 20 AFTER INSTALLATION A B NOTE You can ignore the C of the C1 20 INSTALLING THE SOFTWARE NOTE FOR USING THE OMRON HEALTH MANAGEMENT SOFTWARE For using the Omron Health Management Software with the HEM 790ITCAN the version needs to be 1 20 or higher If you use the included CD ROM you can stop reading this NOTE because its version is 1 20 or higher...

Страница 48: ... Select Uninstall Omron Health Management Software 4 Follow the directions on the message boxes 2 3 Install Omron Health Management Software from either a CD ROM or the Omron Website A From a CD ROM version C1 20 or higher 1 Follow the directions in the included instruction manual B From http software omronhealthcare com 1 Follow the UPDATE SOFTWARE directions on the screen 2 4 Transfer the data f...

Страница 49: ...ftware Enable the anti virus software again after installation is complete 1 Insert the installation CD ROM disc into the CD drive of your PC If no screen appears Windows 2000 i Click My Computer and double click the CD drive in which the CD ROM disc is inserted ii When folder names of the CD ROM are listed double click the setup exe icon Windows XP i Click My Computer from the start menu ii From ...

Страница 50: ...2 The installation screen of MDAC may appear Click OK If this screen does not appear skip to Step 5 3 Select the checkbox I accept and click Next INSTALLING THE SOFTWARE 50 ...

Страница 51: ...ars when your PC must be restarted Select the checkbox Let setup restart the system now and click Finish Your PC will restart After restart start the CD ROM Refer to Step 1 for how to start the CD ROM When restart is unnecessary Setup is complete appears Click Close ...

Страница 52: ...ework 1 1 License Agreement appears Read the License Agreement thoroughly and if you agree with it select I agree and click Install NOTE The program will not be installed if you disagree with the License Agreement 7 The dialog box appears that informs the Microsoft NET Framework 1 1 installation is complete Click OK INSTALLING THE SOFTWARE 52 ...

Страница 53: ...ontinue installation 9 The License Agreement of Omron Health Management Software appears Read the License Agreement thoroughly and if you agree with it select I accept the terms and click Next NOTE The program will not be installed if you disagree with the License Agreement INSTALLING THE SOFTWARE 53 ...

Страница 54: ...ck the installation destination folder and click Next To change the installation destination click Change and select a folder for installation 11 The installation confirmation screen appears To continue installation click Install INSTALLING THE SOFTWARE 54 ...

Страница 55: ...when your PC must be restarted Click Yes to restart your PC 13 The Omron Health Management Software shortcut icon is created on the desktop 14 Remove the CD ROM disc from the CD drive 15 Enable anti virus software if it had been disabled before you installed this software INSTALLING THE SOFTWARE 55 ...

Страница 56: ...he Omron Health Management Software shortcut icon on the desktop If no icon appears on the desktop From the Start menu select All programs or Programs Omron Health Management Software and click the Omron Health Management Software shortcut 2 If you are using Omron Health Management Software for the first time click Add User on the User Management screen USING THE SOFTWARE 56 ...

Страница 57: ... manage multiple users data Data is controlled by user name The number of users who can register for this software varies depending on the amount of free disc space in your PC The user name can be up to 28 characters 4 Click the user name and click Graph Screen 5 Click the Blood Pressure Monitor icon USING THE SOFTWARE ...

Страница 58: ...ssories Parts and accessories not approved for use with the device may damage the unit 1 Open the USB Port cover 2 Plug the normal USB Connector Large Connector into the personal computer 3 Plug the mini USB connector Small Connector into the USB Port TRANSFERRING DATATO A PERSONAL COMPUTER 58 Normal USB Connector Mini USB Connector ...

Страница 59: ... sure your cable is connected then click Yes 6 Select the blood pressure monitor being used and click OK 7 Start importing data Click Cancel if you want to cancel 8 The dialog appears when complete Click OK and view your data on the graph TRANSFERRING DATATO A PERSONAL COMPUTER 59 ...

Страница 60: ...or installation ERROR MESSAGES The operating system currently installed in your PC is neither Windows 2000 nor Windows XP You cannot run this program on your PC Install Omron Health Management Software onto a PC on which the corresponding operating system is running as listed in Operating environment requirement Out of Disc Space Disc space required for the installation exceeds available disc spac...

Страница 61: ...ly Clean the monitor with a soft dry cloth Do not use any abrasive or volatile cleaners Do not attempt to clean the cuf f CAUTION Do not submerge the device or any of the components in water Do not subject the monitor to extreme hot or cold temperatures humidity or direct sunlight CAUTION Store the device and the components in a clean safe location CAUTION Do not subject the monitor to strong shoc...

Страница 62: ...ification not approved by Omron Healthcare will void the user warranty Do not disassemble or attempt to repair the unit or components CAUTION Use only Omron authorized parts and accessories Parts and accessories not approved for use with the device may damage the unit CARE AND MAINTENANCE 62 ...

Страница 63: ...t TruReadTM Mode The monitor will repeat the process up to 5 times Cuff over inflated Take another measurement and remain still until the measurement is complete Monitor could not detect pulse wave Air plug is not connected Batteries are worn Replace the four batteries Refer to Battery Installation ERROR INDICATORS ERROR INDICATORS ANDTROUBLESHOOTINGTIPS ...

Страница 64: ...es Check the battery installation for proper placement of the battery polarities Measurement values appear too high or too low Blood pressure varies constantly Many factors including stress time of day and how you wrap the cuff may affect your blood pressure Review the sections Before Taking a Measurement and Taking a Measurement TROUBLESHOOTING TIPS ...

Страница 65: ...y one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the product and the receiver Connect the product into an outlet on a circuit dif ferent from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help POTENTIAL FOR RADIO TELEVISION INTERFERENCE for Canada only This digital apparatus doe...

Страница 66: ...ies Repair or replacement is our only responsibility and your only remedy under the above warranties To obtain warranty service contact Customer Service by calling 1 800 634 4350 for the address of the repair location and the return shipping and handling fee Enclose the Proof of Purchase Include a letter with your name address phone number and description of the specific problem Pack the product c...

Страница 67: ...n used 6 times a day with 4 new alkaline batteries Operating Temperature 50 F to 104 F 10 C to 40 C 30 to 85 RH Humidity Storage Temperature 4 F to 140 F 20 C to 60 C 10 to 95 RH Humidity Air Pressure 700 1060 hPa Main Unit Weight Approximately 14 7 8 oz 420 g not including batteries Main Unit Dimensions Approx 6 3 32 l 5 5 32 w 3 5 16 h 155 mm 131 mm 84 mm Cuff Size Approximately 6 w 152 mm Cuff ...

Страница 68: ...Made in China OMRON HEALTHCARE INC 1200 Lakeside Drive Bannockburn Illinois 60015 www omronhealthcare com Copyright 2009 Omron Healthcare Inc ...