H
J
I
Year
Month
Day
Hour
Minute
M
L
K
1 to 2 cm
D
A
B
C
F
E
D
D
G
P
Q
R
S
O
N
V
T
U
W
Y
X
Z
Thank you for purchasing the OMRON HEM-7132 Automatic Blood Pressure
Monitor.
The OMRON HEM-7132 is a compact, fully automatic blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the device uses its advanced
“IntelliSense” technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
(General Usage)
• Do not use the device on the injured arm or the arm under medical treatment.
• Do not wrap the arm cuff on the arm while being on a intravenous drip or
blood transfusion.
(AC Adapter (optional) Usage)
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
the power and unplug the power cord immediately.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
multiple-tap.
• Never plug in or unplug the power cord from the electric outlet with wet
hands.
Caution:
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
• If there are any abnormalities during the measurement, remove the arm cuff.
• Do not use this device on infants or persons who cannot express their
intentions.
• Do not inflate the arm cuff more than necessary.
• Do not use the unit for any purpose other than measuring blood pressure.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
in incorrect measurement results.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not disassemble the unit or arm cuff.
(Battery Usage)
• Do not insert the batteries with their polarities incorrectly aligned.
• Use only four “AA” alkaline or manganese batteries with this unit. Do not use
other types of batteries. Do not use new and used batteries together.
• Remove the batteries if the unit will not be used for 3 months or more.
(AC Adapter (optional) Usage)
• Fully insert the power plug.
• When disconnecting the power plug, do not pull the power cord. Be sure to
hold the power plug.
• When handling the power cord, observe the following:
Do not damage.
Do not break it.
Do not tamper with it.
Do not forcibly bend or pull.
Do not twist.
Do not bundle during use.
Do not pinch.
Do not place under heavy objects.
• Wipe the dust of the power plug.
• Disconnect the power pl
ug if the product will not be used for a long period of
time.
• Disconnect the power plug before starting maintenance.
• Use only the original AC adapter designed for this unit. Use of unsupported
adapters may damage and/or may be hazardous to the unit.
General Precautions
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• To unplug the air plug, pull on the air plug at the connection with the main unit
not the tube itself.
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Read and follow the “Important information regarding Electro Magnetic
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in
the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
Important Safety Information
Note: Subject to technical modification without prior notice.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
•
Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation
of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction
manual.
Documentation is also available at www.omron-healthcare.com.
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
Introduction
1. Overview
Main unit:
N. Average value symbol
O. Memory symbol
P. Irregular heartbeat symbol
Q. Systolic blood pressure
R. Diastolic blood pressure
S. Low battery symbol
T. Cuff wrapping guide
2.1 Installing/Replacing the Batteries
2. Preparation
1.
2.
Remove the battery cover.
2.2 Setting the Date and Time
1.
Press the button.
2.
Set the monitor to the correct date and time before
taking a measurement for the first time.
3.
Press the START/STOP button to store the setting.
Notes:
• If the batteries have been replaced, the date and time setting will
need to be reset.
• If the date and time are not set, “-:--” appears during or after
measurement.
Insert four “AA” batteries
as indicated in the battery
compartment and then
replace the battery cover.
Notes:
• If the
low battery symbol ( ) appears on the display, turn the
monitor off then replace all batteries at the same time.
• The measurement values continue to be stored in memory even after
the batteries are replaced.
• The supplied batteries may have a shorter life.
Disposal of used batteries should be carried out in accordance with the
national regulations for the disposal of batteries.
The bottom edge of the arm
cuff should be 1 to 2 cm
above the elbow. The air
tube is centred on the
middle of your inner arm.
4. Troubleshooting and Maintenance
3.
3.3 Taking a Reading
6. Technical Data
5. Optional Parts
4.1 The Icons and Error Messages
4.2 Troubleshooting
4.3 Maintenance
Problem
3.4 Using the Memory Function
Calibration and Service
To protect your unit from damage, please observe the following:
Cause
Remedy
Notes:
• When you take a measurement on the right
arm, the air tube will be at the side of your
elbow. Be careful not to rest your arm on
the air tube.
Notes:
• To cancel a measurement, press the START/STOP button to release
the air in the arm cuff.
• Remain still while taking a measurement.
Press the START/STOP button.
The arm cuff will start to inflate automatically.
Notes:
• If the previous reading was taken without setting
the date and time, the average value is not
calculated.
• If there are no measurements results stored in
the memory, the screen to the right is displayed.
• Do not subject the main unit and the cuff to extreme
temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations
(for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean
the arm cuff.
• Do not carry out repairs of any kind yourself. If a defect
occurs, consult your local OMRON representative.
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your local OMRON
representative.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and neutral soap to clean
the arm cuff.
Automatic Blood Pressure Monitor
HEM-7132
LCD Digital Display
Oscillometric method
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/ min.
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Fuzzy-logic controlled by electric pump
Automatic pressure release valve
60 Measurements with date and time
DC6V 4W
4 “AA” batteries 1.5V
or AC adapter (optional, INPUT AC100-240V
50/60Hz 0.12A)
Approx. 1000 measurements
(using new alkaline batteries)
= Type BF
Internally powered ME equipment (When using only
the batteries)
= Class II ME equipment (Optional AC adapter)
+10 to +40°C / 30 to 85% RH
-20 to +60°C / 10 to 95% RH / 700-1060 hPa
Approx. 280g without batteries
Approx. 170g
Approx. 107 (w) mm
×
79 (h) mm × 141(l) mm
Approx. 145 mm × 532 mm
(Cuff: arm circumference 22 to 42 cm)
Nylon, polyester, polyvinyl chloride
Main unit, arm cuff, instruction manual, battery set
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. No
bathing, drinking alcohol or caffeine,
smoking, exercising or eating 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The arm cuff should be at the same level as
your heart.
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. OMRON recommends to always use the same arm for
measurement. If the values between both arms differ substantially,
please check with your doctor which arm to use for your
measurement.
1.
Press and hold the button for
more than 3 seconds.
1.
Press and hold the START/STOP
button for more than 3 seconds.
The date/time display will disappear.
1.
Release the START/STOP button when the date/time
display turns off.
The arm cuff will start to inflate automatically.
2.
The monitor automatically stores the result up to 60 sets. It can also
calculate an average reading based on the measurements from the
last three readings taken within 10 minutes. If there are only two
readings in memory for that period, the average will be based on two
readings. If there is one reading in memory for that period, the average
will be based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest reading.
• When viewing the reading taken without setting the date and time,
“-:--” is displayed instead of the date and time.
The monitor stores measurement values for one user in the memory.
The guest mode can be used to take a single measurement for
another user. No measurement values are stored in the memory when
the guest mode is selected.
Note: Please check with your local OMRON representatives for the
appropriate optional part models.
START
INFLATE
DEFLATE
COMPLETE
The reading is
extremely low (or
high).
The cuff wrapping
guide lamp does not
light.
Arm cuff deflates too
soon.
Arm cuff pressure
does not rise.
Cannot measure or
readings are too low
or too high.
Arm cuff not applied
correctly.
Movement or talking
during measurement.
Clothing is interfering
with the arm cuff.
The air tube is not
securely connected
into the main unit.
Air is leaking from the
arm cuff.
The arm cuff is loose.
The arm cuff has not
been inflated
sufficiently.
Apply the arm cuff
correctly.
Refer to section 3.1.
Remain still and do not
talk during
measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.1.
Make sure that the air
tube is connected
securely.
Refer to section 3.1.
Replace the arm cuff
with a new one.
Refer to Chapter 5.
Apply the cuff correctly
so that it is firmly
wrapped around the arm.
Refer to section 3.1.
Inflate the cuff so that it
is 30 to 40 mmHg above
your previous
measurement result.
Refer to section 3.3.
1.
Unplug the air plug from the air jack.
2.
Gently fold the air tube into the arm cuff.
4.4 Storage
1.
2.
Using the Optional AC Adapter
Insert the AC adapter plug
into the AC adapter jack on
the rear side of the main
unit.
Plug the AC adapter into an
electrical outlet.
Product Description
Model
Display
Measurement Method
Measurement Range
Accuracy
Inflation
Deflation
Memory
Rating
Power Source
Battery Life
Applied Part
Protection Against
Electric Shock
Operating
temperature/ Humidity
Storage temperature/
Humidity/ Air pressure
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Cuff/ Tube Material
Package Content
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
•
This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
•
This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan.
Automatic Blood
Pressure Monitor
Model HEM-7132
Instruction Manual
A. Display
B. Memory button
C. START/STOP button
D. Date/Time setting ( ) button
E. Cuff wrapping guide lamp
F. Up/Down ( ) buttons
G. Blood pressure colour indicator
H. Battery compartment
I. AC adapter jack (for optional
AC adapter)
J. Air Jack
U. Heartbeat symbol
• Flashes during measurement.
V. Date/Time display
W. Movement error symbol
X. Blood pressure level indicator
Y. Deflation symbol
Z. Pulse display and Memory
number
Push or buttons to
change.
Push the button to confirm.
Next setting appears.
1)
2)
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Cuff Wrapping Guide Lamp
• If you move during measurement, the movement
error symbol ( ) will appear on the display.
Keep still and repeat the measurement.
Alternating date/time display
Display:
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then remove the AC adapter plug from the
main unit.
5341879-6A
K. Air plug
L. Air tube
M. Arm Cuff (Arm circumference
22 -
42 cm)
Arm cuff:
AC Adapter S
Arm Cuff
Arm circumference
22 - 42 cm
• Your blood pressure monitor includes an
irregular heartbeat feature. Irregular heartbeats
can influence the results of the measurement.
The irregular heartbeat algorithm automatically
determines if the measurement is usable or
needs to be repeated. If the measurement
The batteries are low.
The batteries are
exhausted.
Air plug disconnected.
Movement during
measurement and the arm
cuff has not been inflated
sufficiently.
The arm cuff was inflated
above 299 mmHg
when inflating the cuff
manually.
Movement during
measurement.
Clothing is interfering with
the arm cuff.
Device error.
Irregular heartbeats are
detected.
Remove the arm cuff. Wait
2 - 3 minutes and then take
another measurement.
Repeat the steps in section
3.3. If this error continues to
appear, contact your doctor.
Movement during
measurement.
Carefully read and repeat
the steps in section 3.3.
You should replace them
with new ones ahead of
time.
Refer to section 2.1.
You should replace them
with new ones at once.
Refer to section 2.1.
Insert the plug securely.
Refer to section 3.1.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.1.
Air is leaking from the arm
cuff.
Replace the cuff with the
new one.
Refer to Chapter 5.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
If “E2” appears repeatedly,
inflate the cuff manually
until it is 30 to 40 mmHg
above your previous
measurement result.
Refer to section 3.3.
Do not inflate the cuff above
299 mmHg.
Refer to section 3.3.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.1.
Contact your local OMRON
representative.
Cause
Remedy
Error Display
3. Using the Unit
1.
Insert the air plug into the air jack
securely.
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your
upper arm. Do not place the arm cuff over thick clothes.
2.
Remove the arm cuff.
3.
Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after 2 minutes.
Note: Wait 2-3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries to
return to the condition prior to taking the blood pressure
measurement.
The cuff wrapping guide lamp lights in green, the cuff is correctly
wrapped tightly enough on the arm and the reading is accurate and
reliable. If it does not light, the reading is not reliable due to the
incorrect wrapping of the cuff. Please wrap the cuff again, taking care
to wrap it correctly and take the measurement again.
Always consult your doctor. Self-diagnosis of measurement
results and self-treatment are dangerous.
2.
Apply the arm cuff to your upper
arm.
Close the fabric fastener
FIRMLY.
• Press the START/STOP button and repeat
measurement.
• If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your local OMRON representative.
Nothing happens
when you press the
buttons.
Other problems.
The batteries are
empty.
The batteries have
been inserted
incorrectly.
Replace the batteries
with new ones.
Refer to section 2.1.
Insert the batteries with
the correct (+/-) polarity.
Refer to section 2.1.
To View the Average Value
Using the Guest Mode
Press the button.
The Memory number appears for a second before the pulse rate is
displayed. The newest set is numbered “1”.
: To the older readings
: To the more recent readings
: The arm cuff was correctly wrapped tightly enough.
: The arm cuff was loose or not wrapped correctly.
1.
Press the or button to view the readings stored in
memory.
2.
To View the Readings Stored in Memory
Cuff Wrapping Guide Lamp:
To Delete All the Values Stored in Memory
Note: You cannot partially delete the stored readings.
When the memory symbol ( ) appears, first press the Memory
button. Then while holding it down, press the START/STOP
button simultaneously for more than 3 seconds.
Notes:
• Do not bend the air tube excessively.
• Do not store the unit in the following situations:
- If the unit is wet.
- Locations exposed to extreme temperatures, humidity, direct
sunlight, dust or corrosive vapours.
- Locations exposed to vibrations, shocks or where it will be at an
angle.
HEM-7132-IN_A_M03_130621.pdf
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure
Above 85 mmHg
Important:
This criteria is for home blood pressure measurement.
• If your systolic or diastolic pressure is outside the
standard range, The blood pressure colour
indicator will light in orange when the measurement
result is displayed. Recent research suggests that
the following values can be used as a guide to high
blood pressure for measurements taken at home.
results are affected by irregular heartbeats but the result is valid, the
result is shown together with the irregular heartbeat symbol ( ). If
the irregular heartbeats cause the measurement to be invalid, no
result is shown. If the irregular heartbeat symbol ( ) is shown after
you have taken a measurement, repeat the measurement. If the
irregular heartbeat symbol ( ) is shown frequently, please make
your doctor aware of it.
Note:
Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the device and any
used accessories or optional parts.
Notes:
• If your systolic or diastolic pressure is outside the
standard range, the Blood pressure colour
indicator will light when the measurement result
is displayed.
•
The cuff wrapping guide result appears on the
display with the measurement values.The cuff
wrapping guide lamp will not lignt.
Made in Vietnam
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS
www.omron-healthcare.com
Production facility OMRON HEALTHCARE MANUFACTURING VIETNAM
CO., LTD.
Binh Duong Province, VIETNAM
Asia Pacific HQ
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967
www.omron-healthcare.com.sg