Thank you for purchasing the OMRON HEM-7120 Automatic Blood Pressure 

Monitor.
The OMRON HEM-7120 is a compact, fully automatic blood pressure monitor, 

operating on the oscillometric principle. It measures your blood pressure and 

pulse rate simply and quickly. For comfortable controlled inflation without the 

need of pressure pre-setting or re-inflation the device uses its advanced 

“IntelliSense” technology.
Intended Use

This product is designed to measure the blood pressure and pulse rate of 

people within the range of the designated arm cuff, following the instructions in 

this instruction manual.

It is mainly designed for general household use. Please read the Important 

Safety Information in this instruction manual before using the unit.

Please read this instruction manual thoroughly before using the unit.

Please keep for future reference.

For specific information about your own blood pressure, CONSULT YOUR 

DOCTOR.

Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia 

or arteriosclerosis.

Please read this section carefully before using the unit.

Warning: 

Indicates a potentially hazardous situation which, if not avoided, 

could result in death or serious injury.

(General Usage)

• Do not use the device on the injured arm or the arm under medical treatment.

• Do not wrap the arm cuff on the arm while being on an intravenous drip or 

blood transfusion.

(AC Adapter (optional) Usage) 

• Do not use the AC adapter if the unit or the power cord is damaged. Turn off 

the power and unplug the power cord immediately.

• Plug the AC adapter into the appropriate voltage outlet. Do not use a 

multiple-tap. 

• Never plug in or unplug the power cord from the electric outlet with wet 

hands.

Caution:

 Indicates a potentially hazardous situation which, if not avoided, 

may result in minor or moderate injury to the user or patient or 

damage to the equipment or other property.

(General Usage)

• Always consult your doctor. Self-diagnosis of measurement results and 

self-treatment are dangerous.

• People with severe blood flow problems, or blood disorders, should consult a 

doctor before using the unit, as cuff inflation can cause internal bleeding.

• If there are any abnormalities during the measurement, remove the arm cuff.

• Do not use this device on infants or persons who cannot express their 

intentions.

• Do not inflate the arm cuff more than necessary.

• Do not use the unit for any purpose other than measuring blood pressure.

• Use only the approved arm cuff for this unit. Use of other arm cuffs may result 

in incorrect measurement results.

• Do not use a mobile phone or other devices that emit electromagnetic fields, 

near the unit. This may result in incorrect operation of the unit.

• Do not disassemble the unit or arm cuff.

(Battery Usage)

• Do not insert the batteries with their polarities incorrectly aligned.

• Use only four “AA” alkaline or manganese batteries with this unit. Do not use 

other types of batteries. Do not use new and used batteries together.

• Remove the batteries if the unit will not be used for 3 months or more.

(AC Adapter (optional) Usage) 

• Fully insert the power plug.

• When disconnecting the power plug, do not pull the power cord. Be sure to 

hold the power plug.

• When handling the power cord, observe the following:

Do not damage.  

Do not break it.

Do not tamper with it. 

Do not forcibly bend or pull.

Do not twist. 

 

Do not bundle during use.

Do not pinch.   

Do not place under heavy objects.

• Wipe the dust off from the power plug.

• Disconnect the power pl

ug if the product will not be used for a long period of 

time.

• Disconnect the power plug before starting maintenance.

• Use only the original AC adapter designed for this unit. Use of unsupported 

adapters may damage and/or may be hazardous to the unit.

General Precautions

• Do not forcibly bend the arm cuff or bend the air tube excessively.

• To unplug the air plug, pull on the air plug at the connection with the main 

unit, not the tube itself.

• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.

• Do not inflate the arm cuff when it is not wrapped around your arm.

• Read and follow the “Important information regarding Electro Magnetic 

Compatibility (EMC)” in the Technical Data Section.

• Read and follow the “Correct Disposal of This Product” in the Technical Data 

Section when disposing of the device and any used accessories or optional 

parts.

Important Safety Information

Note: Subject to technical modification without prior notice. 

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile 

(cellular) telephones, medical devices in use may be susceptible to 

electromagnetic interference from other devices. Electromagnetic interference 

may result in incorrect operation of the medical device and create a potentially 

unsafe situation.

Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) 

with the aim to prevent unsafe product situations, the EN60601-1-2:2007 

standard has been implemented. This standard defines the levels of immunity to 

electromagnetic interferences as well as maximum levels of electromagnetic 

emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this 

EN60601-1-2:2007 standard for both immunity and emissions.

Nevertheless, special precautions need to be observed:

• 

Do not use mobile (cellular) telephones and other devices, which generate 

strong electrical or electromagnetic fields, near the medical device. This may 

result in incorrect operation of the unit and create a potentially unsafe situation. 

Recommendation is to keep a minimum distance of 7 m. Verify correct operation 

of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at 

OMRON HEALTHCARE EUROPE at the address mentioned in this instruction 

manual.

Documentation is also available at www.omron-healthcare.com.

This marking shown on the product or its literature, indicates that it 

should not be disposed of, with other household wastes at the end 

of its working life. To prevent possible harm to the environment or 

human health from uncontrolled waste disposal, please separate 

this product from other types of wastes and recycle it responsibly to 

promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this 

product, or their local government office, for details of where and how they can 

return this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions 

of the purchase contract. This product should not be mixed with other commercial 

wastes for disposal. 

Correct Disposal of This Product 

(Waste Electrical & Electronic Equipment)

Introduction

1.   Overview

Main unit: 

J.  Memory symbol

K.  Systolic blood pressure

L.  Diastolic blood pressure

M. Low battery symbol

N.  Cuff wrapping guide

O. Heartbeat symbol

1. Flashes during measurement.

2. If flashing after measurement 

completed or when viewing 

results stored in the memory, 

indicates blood pressure out of 

recommended range*.

2.1   Installing/Replacing the Batteries

2.   Preparation

1.
2.

Remove the battery cover.

Insert four “AA” batteries 

as indicated in the battery 

compartment and then 

replace the battery cover.

Notes:

• If the

 low battery symbol (     ) appears on the display, turn the 

monitor off then replace all batteries at the same time.

• The measurement values continue to be stored in memory even after 

the batteries are replaced.

• The supplied batteries may have a shorter life.

Disposal of used batteries should be carried out in accordance with the 

national regulations for the disposal of batteries.

The bottom edge of the arm 

cuff should be 1 to 2 cm 

above the elbow. Marker 

(arrow under the air tube) is 

centred on the middle of 

your inner arm.

3.

3.3   Taking a Reading

6.   Technical Data

5.   Optional Parts

4.3   Maintenance

3.4   Using the Memory Function

Calibration and Service

To protect your unit from damage, please observe the following:

Notes:

• When you take a measurement on the right 

arm, the air tube will be at the side of your 

elbow. Be careful not to rest your arm on 

the air tube.

Notes:

• To cancel a measurement, press the START/STOP button to release 

the air in the arm cuff.

• Remain still while taking a measurement.

Press the START/STOP button.

The arm cuff will start to inflate automatically.

• Do not subject the main unit and the cuff to extreme 

temperatures, humidity, moisture or direct sunlight.

• Do not fold the cuff or tubing tightly.

• Do not inflate the arm cuff over 299 mmHg.

• Do not disassemble the unit.

• Do not subject the unit to strong shocks or vibrations 

(for example, dropping the unit on the floor).

• Do not use volatile liquids to clean the main unit. 

• Do not wash the arm cuff or immerse it in water.

• Do not use petrol, thinners or similar solvents to clean 

the arm cuff.

• Do not carry out repairs of any kind yourself. If a defect 

occurs, consult your local OMRON representative.

• The accuracy of this blood pressure monitor has been 

carefully tested and is designed for a long service life.

• It is generally recommended to have the unit inspected 

every two years to ensure correct functioning and 

accuracy. Please consult your local OMRON 

representative.

• The unit should be cleaned with a soft, dry cloth.

• Use a soft, moistened cloth and neutral soap to clean 

the arm cuff.

Automatic Blood Pressure Monitor

HEM-7120

LCD Digital Display

Oscillometric method

Pressure: 0 to 299 mmHg 

Pulse: 40 to 180 beats/ min.

Pressure: ±3 mmHg

Pulse: ±5% of display reading

Fuzzy-logic controlled by electric pump

Automatic pressure release valve

Last Measurement

DC6V   4W

4 “AA” batteries 1.5V 

or AC adapter (optional, INPUT AC100-240V 

50/60Hz 0.12A)

Approx. 1000 measurements 

(using new alkaline batteries)

           = Type BF           

Internally powered ME equipment (When using only 

the batteries)

       = Class II ME equipment (Optional AC adapter)

+10 to +40°C / 30 to 85% RH

-20 to +60°C / 10 to 95% RH / 700-1060 hPa

Approx. 250g without batteries

Approx. 130g

Approx. 103 (w) mm 

× 

80 (h) mm × 129(l) mm

Approx. 145 mm × 466 mm 

(Cuff: arm circumference 22 to 32 cm)

Nylon, polyester, polyvinyl chloride

Main unit, arm cuff, instruction manual, battery set

3.2   How to Sit Correctly

To take a measurement, you need to be 

relaxed and comfortably seated, under 

comfortable room temperature. No 

bathing, drinking alcohol or caffeine, 

smoking, exercising or eating 30 minutes 

before taking a measurement.

• Sit on a chair with your feet flat on the floor.

• Sit upright with your back straight.

• The arm cuff should be at the same level as 

your heart.

• The blood pressure can differ between the right arm and the left arm, 

and therefore also the measured blood pressure values can be 

different. OMRON recommends to always use the same arm for 

measurement. If the values between both arms differ substantially, 

please check with your doctor which arm to use for your 

measurement.

1.

The monitor automatically stores the last measurement values (blood

pressure and pulse rate). 

Note: Please check with your local OMRON representatives for the 

appropriate optional part models.

START

INFLATE

DEFLATE

COMPLETE

1.

Unplug the air plug from the air jack.

2.

Gently fold the air tube into the arm cuff.

4.4   Storage

1.

2.

Using the Optional AC Adapter

Insert the AC adapter plug 

into the AC adapter jack on 

the rear side of the main 

unit.
Plug the AC adapter into an 

electrical outlet.

Product Description

Model

Display

Measurement Method

Measurement Range

Accuracy

Inflation

Deflation

Memory

Rating 

Power Source

Battery Life

Applied Part

Protection Against 

Electric Shock

Operating 

temperature/ Humidity

Storage temperature/ 

Humidity/ Air pressure

Console Weight

Cuff Weight

Outer Dimensions

Cuff Dimensions

Cuff/ Tube Material

Package Contents

• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device 

Directive). 

• 

This blood pressure monitor is designed according to the European Standard 

EN1060, Non-invasive sphygmomanometers Part 1: General Requirements 

and Part 3: Supplementary requirements for electromechanical blood 

pressure measuring systems.

• 

This OMRON product is produced under the strict quality system of OMRON 

HEALTHCARE Co., Ltd., Japan. The core component for OMRON blood 

pressure monitors, which is the Pressure Sensor, is produced in Japan.

Automatic Blood 

Pressure Monitor

Model HEM-7120

Instruction Manual

A. Display

B.  START/STOP button

C.  Battery compartment

D.  AC adapter jack (for optional 

AC adapter)

E.  Air jack

P.  Irregular heartbeat symbol

Q. Movement error symbol

R.  Pulse display

S.  Deflation symbol

* Note: If your systolic or diastolic 

pressure is outside the standard 

range (above 135/85 mmHg) 

the Heartbeat symbol (   ) will 

blink. Please refer to Section 

3.3.

If your systolic pressure is more than 210 mmHg

After the arm cuff starts to inflate, press and hold the START/STOP 

button until the monitor inflates 30 to 40 mmHg higher than your 

expected systolic pressure.
Notes:

• The monitor will not inflate above 299 mmHg.

• Do not apply more pressure than necessary.

• If you move during measurement, the movement 

error symbol (    ) will appear on the display. 

Keep still and repeat the measurement.

Display: 

To disconnect the AC adapter, unplug the AC adapter from the 

electrical outlet first and then remove the AC adapter plug from the 

main unit.

5342190-8A

F.  Arm cuff (Arm circumference 

22 - 

32 cm)

G. Air plug

H.  Air tube

I. Marker

Arm cuff: 

• Your blood pressure monitor includes an 

irregular heartbeat feature. Irregular heartbeats 

can influence the results of the measurement. 

The irregular heartbeat algorithm automatically 

determines if the measurement is usable or 

needs to be repeated. If the measurement 

3.   Using the Unit

1.

Insert the air plug into the air jack 

securely.

3.1   Applying the Arm Cuff

Remove tight-fitting clothing or tight rolled up sleeve from your 

upper arm. Do not place the arm cuff over thick clothes.

2.

Remove the arm cuff.

3.

Press the START/STOP button to turn the monitor off.

The monitor automatically stores the measurement in its memory.

It will automatically turn off after 2 minutes.

Note:  Wait 2-3 minutes before taking another blood pressure 

measurement. Waiting between readings allows the arteries to 

return to the condition prior to taking the blood pressure 

measurement.

2.

Put your arm through the cuff 

loop.

Close the fabric fastener 

FIRMLY.

Press and hold the START/STOP button for more than 

5 seconds.

The last measurement value is displayed along with the memory 

symbol.

1.

Press the START/STOP button to turn the monitor off.

It will automatically turn off after 2 minutes.

2.

To View the Readings Stored in Memory

To Delete All Values Stored in Memory

Press and hold the START/STOP button for more than 

15 seconds.

Notes:

• Do not bend the air tube excessively.

• Do not store the unit in the following situations:

- If the unit is wet.

- Locations exposed to extreme temperatures, humidity, direct 

sunlight, dust or corrosive vapours.

- Locations exposed to vibrations, shocks or where it will be at an 

angle.

HEM-7120-AP_A_M06_130726.pdf

results are affected by irregular heartbeats but the result is valid, the 

result is shown together with the irregular heartbeat symbol (    ). If 

the irregular heartbeats cause the measurement to be invalid, no 

result is shown. If the irregular heartbeat symbol (    ) is shown after 

you have taken a measurement, repeat the measurement. If the 

irregular heartbeat symbol (    ) is shown frequently, please make 

your doctor aware of it.

Note: 

Read and follow the “Correct Disposal of This Product” in the 

Technical Data Section when disposing of the device and any 

used accessories or optional parts.

Systolic Blood Pressure 

Above 135 mmHg 

Diastolic Blood Pressure 

Above 85 mmHg

• If your systolic or diastolic pressure is outside 

the standard range, the heartbeat symbol will 

blink when the measurement result is displayed. 

Recent research suggests that the following 

values can be used as a guide to high blood 

pressure for measurements taken at home.

Important:

This criteria is for home blood pressure measurement.

AC Adapter S

Medium Arm Cuff

Arm circumference 

22 - 32 cm

Large Arm Cuff

Arm circumference 

22 - 42 cm

Made in Vietnam

Manufacturer

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, Kyoto

617-0002 JAPAN

Production facility OMRON HEALTHCARE MANUFACTURING VIETNAM 

CO., LTD.

Binh Duong Province, VIETNAM

Asia Pacific HQ

OMRON HEALTHCARE SINGAPORE PTE LTD.

438A Alexandra Road, #05-05/08

Alexandra Technopark, Singapore 119967

www.omronhealthcare-ap.com

Note:  This reading is 

NOT

 reliable due to the incorrect wrapping of the 

cuff. Please wrap the cuff again, taking care to wrap it correctly 

and take the measurement again. When the       is displayed, the 

cuff is correctly wrapped tightly enough on the arm and the 

reading is accurate and reliable.

The Cuff Wrapping Guide is a unique feature that indicates if the cuff 

is not wrapped tightly enough around the arm. Even when the       

appears on the display, a blood pressure reading will be taken.

Cuff Wrapping Guide:

Cuff Wrapping Guide

4.   Troubleshooting and Maintenance

4.1   The Icons and Error Messages

4.2   Troubleshooting

Problem

Cause

Remedy

The reading is 

extremely low (or 

high).

Arm cuff deflates too 

soon.

Arm cuff pressure 

does not rise.

Cannot measure or 

readings are too low 

or too high.

Arm cuff not applied 

correctly.

Movement or talking 

during measurement.

Clothing is interfering 

with the arm cuff.

The air tube is not 

securely connected 

into the main unit.

Air is leaking from the 

arm cuff.

The arm cuff is loose.

The arm cuff has not 

been inflated 

sufficiently.

Apply the arm cuff 

correctly. 

Refer to section 3.1.

Remain still and do not 

talk during 

measurement.

Refer to section 3.3.

Remove any clothing 

interfering with the arm 

cuff.

Refer to section 3.1.

Make sure that the air 

tube is connected 

securely. 

Refer to section 3.1.

Replace the arm cuff 

with a new one. 

Refer to Chapter 5.

Apply the cuff correctly 

so that it is firmly 

wrapped around the arm. 

Refer to section 3.1.

Inflate the cuff so that it 

is 30 to 40 mmHg above 

your previous 

measurement result.

Refer to section 3.3.

The batteries are low.

The batteries are 

exhausted. 

Air plug disconnected.

The arm cuff was inflated 

above 299 mmHg 

when inflating the cuff 

manually.

Movement during 

measurement.

Clothing is interfering with 

the arm cuff.

Device error.

Irregular heartbeats are 

detected.

Cuff is not applied correctly.

Remove the arm cuff. Wait 

2 - 3 minutes and then take 

another measurement. 

Repeat the steps in section 

3.3. If this error continues to 

appear, contact your doctor.

Movement during 

measurement.

Carefully read and repeat 

the steps in section 3.3.

Apply the arm cuff correctly. 

Refer to section 3.1.

You should replace them 

with new ones ahead of 

time.

Refer to section 2.1.
You should replace them 

with new ones at once.

Refer to section 2.1.

Insert the plug securely.

Refer to section 3.1.

Arm cuff not applied 

correctly.

Apply the arm cuff correctly.

Refer to section 3.1.

Air is leaking from the arm 

cuff.

Replace the cuff with the 

new one.

Refer to Chapter 5.

Repeat measurement. 

Remain still and do not talk 

during measurement.

Refer to section 3.3.

If “E2” appears repeatedly, 

inflate the cuff manually 

until it is 30 to 40 mmHg 

above your previous 

measurement result.

Refer to section 3.3.

Do not inflate the cuff above 

299 mmHg.

Refer to section 3.3.

Repeat measurement. 

Remain still and do not talk 

during measurement.

Refer to section 3.3.

Remove any clothing 

interfering with the arm cuff.

Refer to section 3.1.

Contact your local OMRON 

representative.

Cause

Remedy

Error Display

• Press the START/STOP button and repeat 

measurement.

• If the problem continues, try replacing the 

batteries with new ones.

If this still does not solve the problem, contact 

your local OMRON representative.

Nothing happens 

when you press the 

buttons.

Other problems.

The batteries are 

empty.

The batteries have 

been inserted 

incorrectly.

Replace the batteries 

with new ones.

Refer to section 2.1.
Insert the batteries with 

the correct (+/-) polarity.

Refer to section 2.1.

Movement during 

measurement and the arm 

cuff has not been inflated 

sufficiently.

Always consult your doctor. Self-diagnosis of measurement 

results and self-treatment are dangerous.

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS

www.omron-healthcare.com

Notes:

• If your systolic or diastolic pressure is outside the 

standard range, the heartbeat symbol will blink 

when the measurement result is displayed.

• 

The cuff wrapping guide result appears on the 

display with the measurement values.

• If there are no measurements results stored in 

the memory, the screen to the right is displayed.

Содержание HEM-7120

Страница 1: ...on can cause internal bleeding If there are any abnormalities during the measurement remove the arm cuff Do not use this device on infants or persons who cannot express their intentions Do not inflate the arm cuff more than necessary Do not use the unit for any purpose other than measuring blood pressure Use only the approved arm cuff for this unit Use of other arm cuffs may result in incorrect me...

Страница 2: ...tteries may have a shorter life Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries The bottom edge of the arm cuff should be 1 to 2 cm above the elbow Marker arrow under the air tube is centred on the middle of your inner arm A Display B START STOP button C Battery compartment D AC adapter jack for optional AC adapter E Air ja...

Страница 3: ... systolic pressure is more than 210 mmHg After the arm cuff starts to inflate press and hold the START STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure Notes The monitor will not inflate above 299 mmHg Do not apply more pressure than necessary 2 Remove the arm cuff 3 Press the START STOP button to turn the monitor off The monitor automatically stores...

Страница 4: ...ar heartbeats but the result is valid the result is shown together with the irregular heartbeat symbol If the irregular heartbeats cause the measurement to be invalid no result is shown If the irregular heartbeat symbol is shown after you have taken a measurement repeat the measurement If the irregular heartbeat symbol is shown frequently please make your doctor aware of it Systolic Blood Pressure...

Страница 5: ...epeat the steps in section 3 3 If this error continues to appear contact your doctor Movement during measurement Carefully read and repeat the steps in section 3 3 Apply the arm cuff correctly Refer to section 3 1 You should replace them with new ones ahead of time Refer to section 2 1 You should replace them with new ones at once Refer to section 2 1 Insert the plug securely Refer to section 3 1 ...

Страница 6: ...epresentative The unit should be cleaned with a soft dry cloth Use a soft moistened cloth and neutral soap to clean the arm cuff Note Please check with your local OMRON representatives for the appropriate optional part models 1 Unplug the air plug from the air jack 2 Gently fold the air tube into the arm cuff 4 4 Storage 1 2 Using the Optional AC Adapter Insert the AC adapter plug into the AC adap...

Страница 7: ... Business users should contact their supplier and check the terms and conditions of the purchase contract This product should not be mixed with other commercial wastes for disposal Correct Disposal of This Product Waste Electrical Electronic Equipment 6 Technical Data Automatic Blood Pressure Monitor HEM 7120 LCD Digital Display Oscillometric method Pressure 0 to 299 mmHg Pulse 40 to 180 beats min...

Отзывы: