Instruction Manual

Introduction

Important Safety Information

1.   Overview

Model 

HEM-7113

The unit also stores up to 14 measurements in memory.

Main unit: 

A. Display

C. Memory button

D. Battery compartment

E. Air jack

A

B

C

D

E

5337886-7A

Thank you for purchasing the OMRON HEM-7113 Digital Automatic Blood 

Pressure Monitor.

The OMRON HEM-7113 is a fully automatic blood pressure monitor, operating 

on the oscillometric principle. It measures your blood pressure and pulse rate 

simply and quickly.

Please read this instruction manual thoroughly before using the unit. For 

specific 

information about your own blood pressure, CONSULT YOUR 

DOCTOR.

Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or 

arteriosclerosis.

Please read this section carefully before using the unit.

     Warning:

• Indicates a potentially hazardous situation which, if not avoided, could result in 

death or serious injury.

(General Usage)

• Always consult your doctor. Self-diagnosis of measurement results and 

self-treatment are dangerous.

• People with severe blood flow problems, or blood disorders, should consult a 

doctor before using the unit, as cuff inflation can cause internal bleeding.

(Battery Usage)

• If battery fluid should get in your eyes, immediately rinse with plenty of clean 

water. Consult a doctor immediately.

     Caution:

• Indicates a potentially hazardous situation which, if not avoided, may result in 

minor or moderate injury to the user or patient or damage to the equipment or 

other property.

(General Usage)

• Do not leave the unit unattended with infants or persons who cannot express 

their consent.

• Do not use the unit for any purpose other than measuring blood pressure.

• Do not disassemble the unit or arm cuff.

• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in 

incorrect measurement results.

• Make sure that the air tube is not wrapped around other parts of your body when 

taking measurements. This could result in injury when the air pressure in the air 

tube is increased.

• Do not leave the cuff wrapped on the arm if taking measurements during the 

night. This could result in injury.

• Do not inflate the arm cuff over 299 mmHg.

• Do not use a mobile phone or other devices that emit electromagnetic fields, 

near the unit. This may result in incorrect operation of the unit.

• Do not operate the unit in a moving vehicle (car, airplane).

• To inflate the cuff manually, refer to Chapter 3.3. If the cuff is over inflated, it can 

cause internal bleeding.

(Battery Usage)

• If battery fluid should get on your skin or clothing, immediately rinse with plenty of 

clean water.

• Use only four“ AAA” alkaline batteries with this unit. Do not use other types of 

batteries.

• Do not insert the batteries with their polarities incorrectly aligned.

• Replace old batteries with new ones immediately. Replace all four batteries at 

the same time.

• Remove the batteries if the unit will not be used for three months or more.

• Do not use new and used batteries together. General Precautions

• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.

• Do not take measurements after bathing, drinking alcohol, smoking, exercising 

or eating.

• Do not forcibly bend the arm cuff or bend the air tube excessively.

• When removing the air tube, pull on the air plug at the connection with the main 

unit not the tube itself.

• Do not inflate the arm cuff when it is not wrapped around your arm.

• Do not wash the arm cuff or immerse it in water.

• Read and follow the“ Important information regarding Electro Magnetic 

Compatibility (EMC)” in the Technical Data Section.

• Read and follow the“ Correct Disposal of This Product” in the Technical Data 

Section when disposing of the device and any used accessories or optional 

parts.

B. START/STOP button.

Automatic 

Blood Pressure Monitor

Be sure to wrap the arm cuff correctly so that you get accurate results.

Measurements can be taken in light clothing. However, please remove 

thick clothes, such as sweaters, before taking a reading.

2.1   Installing/Replacing the Batteries

2.   Preparation

3.   Using the Unit

2.2   Battery Life & Replacement

3.1   Applying the Arm Cuff

1.

Turn the main unit upside down.

2.

Slide the battery c

over

 in the direction 

of the arrow while pressing the ribbed 

part of the co

v

er.

3.

Install or replace four 

“AAA”

 s

ize

 

batteries so that the + (positi

ve)

 

and - (negat

iv

e) polarities match 

the polarities indicated on the battery 

compartment.

4.

Put the battery 

cover

 back in place.

Slide the battery co

v

er as indicated, 

until it clicks into place. 

- When the battery low symbol (     ) starts to blink, you will still be able to use 

  the unit for a short while. You should replace the batteries with new ones 

  ahead of time.

- When the symbol (     ) remains lit, the batteries are exhausted. You should 

  replace the batteries with new ones at once. Turn the unit off before replacing 

  the batteries.

Four new “AAA” alkaline batteries will last for approximately 300 measurements, 

when used to take two measurements a day.

Since the supplied batteries are for monitoring use only, they may have 

a shorter life and not last for 300 measurements.

• Remove the batteries if the unit will not be used for three months or more.

• Dispose of batteries according to applicable local regulations.

If the battery low symbol (     ) appears on the display, 

replace all  four batteries at the same time.

1.

1.

To delete stored readings, first press the Memory (M) 

button.

Then while holding it down, press the START/STOP 

button simultaneously for about 2-3 seconds. All readings 

will then be deleted.

2.

Press the START/STOP button to turn the unit off.

If you forget to turn the unit off, it will automatically shut

itself off after five minutes.

To Delete All the Values Stored in Memory

You cannot delete individual stored readings, all the

readings in the unit will be deleted.

2

1

4.   Handling Errors and Problems

4.1   Error Messages

Error Display

Cause

Remedy

Cuff is under inflated.

Carefully read and repeat the 

steps listed in section 3.3.

Movement during 

measurement.

Repeat measurement. Remain 

still and do not talk during 

measurement.

Air plug disconnected.

Insert the air plug securely.

Refer to section 3.1.

Arm cuff not applied 

correctly.

Apply the arm cuff correctly. 

Refer to section 3.1.

Clothing is interfering 

with the arm cuff.

Remove any clothing 

interfering with the arm cuff.

Air is leaking from the 

arm cuff.

Replace cuff with new one.

Arm cuff has been 

inflated above 

299 mmHg.

Do not inflate the arm cuff 

above 299 mmHg.

Refer to section 3.3.

Intended Use

This product is designed to measure the blood pressure and pulse rate of 

people within the range of the designated arm cuff, following the instructions in 

this instruction manual.

It is mainly designed for general household use. Please read the Important 

Safety Information in this instruction manual before using the unit.

1 - 2 cm

2.

Note: 

You can take a measurement on either your left or right arm. The blood 

pressure can differ between the right arm and the left arm and therefore 

also the measured blood pressure values can be different. OMRON 

recommends to always use the same arm for measurement. If the 

values between the two arms differ substantially, please check with your 

physician which arm to use for your measurement.

Insert the air plug into the air jack 

on the left side of the main unit.

Apply the arm cuff to your upper arm. The air tube should be 

centred on the inside of your arm and point down the inside of the 

arm, so that the air tube runs down the inside of your forearm and 

is in line with your middle finger. 

3.3   Instructions for Special Conditions

5.

Press the START/STOP button to turn the unit off.

1.

Press the START/STOP button 

to turn the unit on.

Measurement starts.

Note: If you forget to turn the unit off, it will automatically shut itself off after 

five minutes.

If your systolic pressure is known to be more than 220 mmHg, press 

and hold the START/STOP button until the arm cuff inflates 30 to 

40 mmHg higher than your suspected systolic pressure.

Arm cuff: 

F.  Arm cuff

(Medium cuff:

  arm circumference 

22-32 cm)

G. Air tube

Note: 

G

F

H

Package contents: 

Display: 

J

Systolic blood pressure

K

Diastolic blood pressure

L

Memory symbol

N

Heartbeat symbol

  1

Flashes during measurement

O

Pulse display

P

Deflation symbol

Q

. Battery low symbol

I

Instruction manual

Blood pressure monitor unit

I

J

L

N

P

K

O
Q

H. Air plug

M. Irregular Heartbeat symbol

M

If air starts to leak from the arm cuff, 

please replace with a new one.

Refer to “6. Optional Parts”.

The optional arm cuff does not 

include the air plug. Do not throw 

the air plug away in case 

purchasing the new one.

Four “AAA” alkaline (

L

R03) 

batteries

 

2

If flashing after measurement 

complete indicates blood pressure 

out of recommended range

Note: If your systolic or diastolic 

pressure is outside the 

standard range 

(above 135/85 mmHg), 

the Heartbeat symbol (    ) 

will blink. Refer to 

“3.2 Taking a Reading”.

Important:

4.

Remove the arm cuff.

Your blood pressure monitor includes an 

irregular heartbeat feature. Irregular heart-

beats can influence the results of the 

measurement. The irregular heartbeat 

algorithm automatically determines if the 

measurement is usable or needs to be 

repeated. If the measurement results are affected by 

irregular heartbeats but the result is valid, the result is 

shown together with the irregular heartbeat symbol (     ).

If the irregular heartbeats cause the measurement to be 

invalid, no result is shown. If the Irregular heartbeat 

symbol (     ) is shown after you have taken a measure

-

ment, repeat the measurement. If the Irregular heartbeat 

symbol (     ) is shown frequently, please make your doctor 

aware of it.

NormalHeartbeat

Blood

pressure

Blood

pressure

Pulse

Pulse

     

Warning:

What is Arrhythmia?

An irregular heartbeat is a heartbeat 

rhythm that varies by more than 25% 
from the average heartbeat rhythm 

detected while the unit is measuring 
the systolic and diastolic blood 
pressure. If such an irregular rhythm 
is detected more than twice during 
measurement, the irregular heartbeat 
symbol (     ) appears on the symbol 
when the measurement results are 
displayed.

A heartbeat is stimulated by electrical signals that cause the heart to contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws 

in the bio-electrical system that drives the heartbeat. Typical symptoms are 

skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) 
or slow (bradycardia) pulse. This can be caused by heart disease, aging, 
physical predisposition, stress, lack of sleep, fatigue etc. Arrhythmia can only 

be diagnosed by a doctor through a special examination. Whether the 

appearance of the irregular heartbeat symbol (     ) in the results indicates 

arrhythmia or not can only be determined by an examination and diagnosis by 

your doctor.

What is Irregular Heartbeat?

If the irregular heartbeat symbol (     ) is shown frequently, please make 

your doctor aware of it. Conducting self-diagnosis and treatment based on 

measurement results is dangerous. Be sure to follow the instructions of 

your doctor.

Irregular 

Short

Heartbeat 

Long

Note: The measurement values 

continue to be stored in 

memory even after the 

batteries are replaced.

Arm cuff pressure

does not rise. 

Make sure that the air 

jack is connected

securely. 

Refer to section 3.1.

Air is leaking from the

arm cuff. 

Replace the arm cuff 

with a new one. 

Arm cuff deflates too

soon. 

The arm cuff is loose.

Apply the cuff correctly 

so that it is firmly 

wrapped around the 

arm. 

Refer to section 3.1.

Cannot measure or

readings are too low

or too high. 

Inflate the cuff so that it 

is 30 to 40 mmHg 

above your previous 

measurement result.

Refer to section 3.3.

The unit loses power 

during measurement.

The batteries are 

empty.

Replace the batteries 

with new ones.

Nothing happens 

when you press the 

buttons.

The batteries are 

empty.

Replace the batteries 

with new ones.

The batteries have 

been inserted 

incorrectly.

Insert the batteries with 

the correct (+/ -) 

polarity.

4.2   Troubleshooting

Problem

Cause

Remedy

The reading is

extremely low (or 

high). 

Arm cuff not applied 

correctly.

Apply the arm cuff 

correctly. 

Refer to section 3.1.

Movement or talking 

during measurement.

Remain still and do not 

talk during 

measurement.

Clothing is interfering 

with the arm cuff.

Remove any clothing 

interfering with the arm 

cuff.

Other problems.

Press the START/STOP button and repeat 

measurement.

If the problem continues, try replacing the 

batteries with new ones.

If this still does not solve the problem, contact your

local authorized OMRON distributor or dealer.

Note: If the memory is full, the monitor will delete the oldest readings.

This unit has a memory capable of storing 14 sets of readings.

Press the Memory (M) button.

The result from the most recent measurement is displayed.

Note: 

If there are no measurements results

        stored in memory, the screen to the right 

        is displayed.

If your systolic or diastolic pressure is

outside the standard range, the heartbeat

symbol will blink when the measurement 

result is displayed. 

Refer to “3.2 Taking a Reading”.

Press the Memory (M) button repeatedly to 

cycle through the previous measurement 

results.

The Memory number appears for a second 

before the pulse rate is displayed. 

The newest set is numbered “1”.

Press the START/STOP button to turn the unit off.

If you forget to turn the unit off, it will automatically shut

itself off after five minutes.

3.4   Using the Memory Function

Important:

1.

2.

3.

At the end of the measurement, if an irregular heartbeat 

is detected, the irregular heartbeat symbol will be 

displayed with the result.

1 to 2 cm

3.2   Taking a Reading

Taking a Reading on the Right Arm

3.

Sit on a chair with your feet flat on 

the floor and place your arm on a 

table so that the arm cuff will be 

at the same level as your heart.

Keep still and do not talk during measurement.

2.

Press the START/STOP button to turn on the unit.

1.

3.

Check the measurement results.

The unit automatically stores blood 

pressure and pulse rate into its memory.

Refer to “3.4 Using the Memory 

Function”.

Notes:

• Self-diagnosis of measured results and treatment are dangerous. 

  Please follow the instructions of your doctor.

• Wait 2-3 minutes before taking another blood pressure reading. 

  Waiting between readings allows the arteries to return to the condition prior 

  to taking the blood pressure measurement.

Diastolic blood 

pressure

Systolic blood 

pressure

Pulse display

When the cuff is positioned correctly, close the fabric fastener 

firmly.

Apply the cuff so that the air tube is 

at the side of your elbow.

Notes:

• Be careful not to rest your arm on 

the air tube, or otherwise restrict the 

flow of air to the cuff.

• The cuff should be 1 to 2 cm above 

the elbow.

Note: To cancel a measurement, press the START/STOP button to turn off 

the unit and to release the air in the arm cuff.

Error Display

Cause

Remedy

Battery power is low.

Replace all four “AAA” 

batteries with new ones.

Refer to section 2.1.

Device error.

Blinks or appears

continuously 

Note: The irregular heartbeat symbol (     ) may also be displayed with error messages.

Note: You cannot inflate the cuff above 299 mmHg. (An error will be displayed 

          if you try to inflate the cuff above 299 mmHg.)

Note: Do not apply more pressure than necessary.

2.

When the cuff starts to inflate, press the START/STOP button 

and keep it pressed until the pressure is 30 to 40 mmHg 

higher than your suspected systolic pressure.

3.

Release the START/STOP button when the cuff has been 

inflated to the desired pressure.

The cuff starts to deflate and measurement starts.

4.

The rest of the procedure is the same as for normal 

measurement. Refer to “3.2 Taking a Reading”, steps 3 to 5.

The air plug is not 

securely inserted into 

the main unit.

The arm cuff has 

not been inflated 

sufficiently.

5.   Maintenance and Storage

Maintenance

To protect your unit from damage, please observe the following:

7.   Technical Data

6.   Optional Parts

• The unit should be cleaned with a soft, dry cloth.

• 

Use a soft, moistened cloth and soap to clean the arm cuff.

Calibration and Service

• The accuracy of this blood pressure monitor has been

carefully tested and is designed for a long service life.

Note: Do not bend the air tube excessively.

Unplug the air tube from the air jack.

Gently fold the air tube

into the arm cuff.

Storage

Do not store the unit in the following situations:

• If the unit is wet.

• Locations exposed to extreme temperatures, humidity, direct 

sunlight, 

dust or corrosive vapours.

• Locations exposed to vibrations, shocks or where it will 

be at an angle.

Medium arm cuff

Arm circumference 

22 - 32 cm

Large arm cuff

Arm circumference 

32 - 42 cm

• Do not subject the main unit and cuff to extreme 

temperatures, humidity, moisture or direct sunlight.

• Do not fold the cuff or tubing tightly.

• Do not disassemble the unit.

• Do not subject the unit to strong shocks or vibrations 

(for example, dropping the unit on the floor).

• Do not use volatile liquids to clean the main unit. 

• Do not wash the arm cuff or immerse it in water.

• Do not use petrol, thinners or similar solvents to clean 

the arm cuff.

• Do not carry out repairs of any kind by yourself.

  If a defect occurs, consult your local authorized 

OMRON distributor or dealer.

• It is generally recommended to have the unit inspected every 

two years to ensure correct functioning and accuracy. Please 

consult your local authorized OMRON distributor or dealer.

Automatic Blood Pressure Monitor

HEM-7113 (HEM-7113-AP)

LCD Digital Display

Oscillometric method

Pressure: 0 mmHg to 299 mmHg

Pulse: 40 to 180/min.

14 Measurements

Pressure: ±3 mmHg

Pulse: ±5% of display reading

Automatic by electric pump

Automatic pressure release valve

4 "AAA" batteries 1.5V

Capacity of new alkaline batteries is approximately

300 measurements

Notes: 

• Subject to technical modification without prior notice.

• 

Disposal of this product and used batteries should be carried out in accordance 

  with the national regulations for the disposal of electronic products.

Operating

temperature/ Humidity

Storage temperature/

Humidity/ Air Pressure

Console Weight

Cuff Weight

Outer Dimensions

Cuff Dimensions

Package Content

10°C to 40°C/ 30 to 85% RH

-20°C to 60°C/ 10 to 95% RH/ 700-1060 hPa

Approximately 245 g without batteries

Approximately 120 g

Approximately 104 (w) mm × 64 (h) mm × 128.5 (I) mm

Approximately 146 mm × 446 mm

(Medium cuff: arm circumference 22 to 32 cm)

Main unit, cuff, Instruction manual, Battery set

Product Description

Model

Display

Measurement Method

Measurement Range

Memory

Accuracy 

Inflation

Deflation

Power Source

Battery life

Applied Part

Protection Against

Electric Shock

= Type B

Internally powered ME equipment

Asia Pacific HQ

Production Facility

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, Kyoto,

617-0002 JAPAN

OMRON HEALTHCARE SINGAPORE PTE LTD.

OMRON HEALTHCARE MANUFACTURING

VIETNAM CO., LTD.

Binh Duong Province, VIETNAM

438A Ale

x

andra Road, 

#05-0

5/

08

, Ale

x

andra Technopark,

Singapore 

1199

67 

www.omron-healthcare.com.sg

This OMRON product is manufactured under the strict quality system of OMRON 

HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood pressure 

monitors, which is the Pressure Sensor, is produced in Japan for assembly.

Contact your local 

authorized OMRON 

distributor or dealer

Отзывы: