HEM-7112-AP_A_M03_110217.pdf
SPECIFICATIONS
Model:
HEM-7112
TROUBLESHOOTING TIPS
PROBLEM
No power.
No display appears on the unit.
Measurement values appear too
high or too low.
CAUSES AND SOLUTIONS
Replace all four batteries with new ones.
Check the battery installation for proper placement
of the battery polarities.
Blood pressure varies constantly. Many factors
including stress, time of day, how you wrap the
cuff, may affect your blood pressure.
Review the section "Taking a Measurement".
CARE AND MAINTENANCE
USING THE MEMORY FUNCTION
NOTE:
The monitor will automatically turn off after five minutes.
2. Press the START/STOP button to turn the unit off.
1. Press and hold the START/STOP button until the measurement value is displayed.
Then keep the button pressed down for about 10 seconds until the display appears as
illustrated below.
TO DELETE THE VALUES STORED IN MEMORY
2. Press the START/STOP button to turn the unit off.
1. To view your last measurement values, press and hold the START/STOP button.
The most recent set of measurement values appears on the display.
Every time you complete a measurement the monitor stores the blood pressure and pulse
rate in the memory. The monitor automatically stores the last set of measurement values
(systolic pressure, diastolic pressure, and pulse rate).
ERROR INDICATORS
SYMBOL
CAUSE
Monitor could not
detect pulse wave.
Cuff over-inflated.
Batteries are worn.
Device error.
Cuff under-inflated.
Cuff not applied
correctly.
Remedy
Take another measurement and
remain still until the measurement
is complete.
Replace the four batteries. Refer to
“Battery Installation”.
Contact Customer Service.
Remove the arm cuff.
Read “Taking a Measurement”.
Wait 2-3 minutes.
Take another measurement.
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
CAUTION
When the Battery low indicator appears on the display, turn the monitor off and
remove all the batteries. Replace with four new batteries at the same time. Long-life
alkaline batteries are recommended.
Battery low indicator
BATTERY REPLACEMENT
If battery fluid should get in your eyes, immediately rinse with plenty of
clean water. Contact a doctor immediately.
WARNING
TAKING A MEASUREMENT
NOTE:
The monitor will automatically turn off after five minutes.
5. Press the START/STOP button to turn the monitor off.
4. When the measurement is complete, the arm cuff completely
deflates. Your blood pressure and pulse rate are displayed.
3. As the cuff deflates, decreasing numbers appear on the display and the Heartbeat
Symbol ( ) flashes.
NOTE:
To stop the inflation or measurement, push the START/STOP button. The
monitor will stop inflating, start deflating, and will turn off.
The cuff starts to inflate automatically.
Remain still and do not move your arm until the entire measurement process is completed.
All symbols appear on the display.
2. Press the START/STOP button.
1. Sit on a chair with your feet flat on the floor. Place
your arm on a table so the cuff is level with your
heart.
Keep still and do not talk during measurement.
Be careful not to rest your arm on the air tube. This will restrict the flow of
air to the cuff.
NOTE:
Apply the cuff to your right upper arm so the
arrow and the air tube runs along the inside of
your arm. The bottom of the cuff should be
approximately
1 to 2 cm
above your elbow.
When taking a measurement using the right arm, use this instruction for Step 3.
APPLYING THE CUFF ON THE RIGHT ARM
3. Apply the cuff to your left upper arm so the
arrow is centered on the inside of your arm and
aligned with your middle finger.
The air tube runs down the inside of your arm.
The bottom of the cuff should be approximately
1 to 2 cm
above your elbow.
4. Wrap the cuff firmly in place around your arm
using the cloth fastening.
Arrow
Left Arm
Right Arm
Arrow
IMPORTANT SAFETY INFORMATION
• Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(AC Adapter Usage)
• Never plug in or unplug the power cord from the electric outlet with wet hands.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express their
consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in
incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body when
taking measurements at night. This could result in injury when the air pressure in
the air tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the night.
This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields, near
the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to "TAKING A MEASUREMENT". If the cuff
is over inflated, it can cause internal bleeding.
(AC Adapter Usage)
• Use only the original AC adapter (optional) designed for this unit. Use of
unsupported adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a multiple-tap.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off the
power and unplug the power cord immediately.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with plenty
of clean water.
• Use only four “AAA” alkaline batteries with this unit. Do not use other types of
batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at the
same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking, exercising or
eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the main
unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
Warning:
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
BATTERY INSTALLATION
3. Replace the battery cover.
2. Install 4 “AAA” size batteries so the + (positive) and -
(negative) polarities match the polarities of the battery
compartment as indicated.
1. Press the indicator on the battery cover and slide
the cover off in the direction of the arrow.
Use only 1.5V “AAA” alkaline batteries with this device.
CAUTION
UNIT DISPLAY
Systolic blood pressure
Diastolic blood pressure
Memory symbol
(Displayed when viewing
values stored in memory)
Heartbeat symbol
Pulse display
Deflation symbol
Battery low indicator
KNOW YOUR UNIT
Main Unit:
Components:
START/STOP button
AC adapter jack
Battery compartment
Air jack
Standard Adult Arm Cuff
Air tube
Air plug
Instruction
Manual
Display
APPLYING THE ARM CUFF
2. Remove tight-fitting clothing from your upper
arm.
1. Make sure the air plug is securely inserted in the
main unit.
APPLYING THE CUFF ON THE LEFT ARM
This device is intended for use in measuring blood pressure and
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
CAUTION
CAUTION
When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
NOTES:
• Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
2. When the cuff starts to inflate, press and hold the START/STOP button until the
monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
Press the START/STOP button to turn the unit on.
Measurement starts.
1.
If your systolic pressure is known to be more than 140 mmHg,
CAUTION
Inflating to a higher pressure than necessary may result in bruising where
the cuff is applied.
SPECIAL CONDITIONS
Air Tube
1 to 2 cm
Display:
Measurement Range:
Accuracy:
Inflation:
Deflation:
Rapid Air Release:
Measurement Method:
Power Source:
Battery Life:
Applied Part:
Protection against
Electric Shock:
Operating Temperature
/Humidity:
Storage Temperature
/Humidity/Air Pressure:
Main Unit Weight:
Cuff Weight:
Main Unit Dimensions:
Cuff Size:
Cuff Circumference:
Memory:
Contents:
Optional Accessories:
LCD Digital Display
Pressure: 0 to 299 mmHg, Pulse: 40 to 180 beat
s/min.
Pressure: ±3mmHg
Pulse: ±5% of reading
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Oscillometric method
1.5V 4 “AAA” batteries or AC adapter (optional)
Rating:
DC6V 4W
Approx. 300 uses with 4 new alkaline batteries
10ºC to 40ºC /30 to 85% RH
-20ºC to 60ºC /10 to 95% RH /
700-1060 hPa
Approximately 245g
not including batteries
Approximately 120g
Approximately 128 mm × 104 mm × 64 mm
Approximately 146 mm (w) × 446 mm (l)
Cuff tube (600 mm)
Fits arm circumferences 220 to 320 mm
The last reading
Main Unit, Standard Adult Arm Cuff, Battery Set,
Instruction Manual
Large Cuff 320 to 420 mm, AC adapter
HEM-7112
Notes:
•
Subject to technical modification without prior notice.
•
Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
Please read this instruction manual carefully before using the device.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
•
This blood pressure monitor is designed according to the European Standard EN1060,
Noninvasive sphygmomanometers Part 1
: G
eneral Requirements and Part 3:
Supplementary requirements for electromechanical blood pressure measuring systems.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of the medical
device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
•
Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct
operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
This product does not contain any ha
z
ardous substances.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
Asia Pacific HQ
Production Facility
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
OMRON HEALTHCARE SINGAPORE PTE LTD.
438
A Alexandra Road, #05-05/
08
, Alexandra Technopark
Singapore
11996
7
www.omron-healthcare.com.sg
OMRON (DALIAN) CO., LTD.
Dalian, China
This OMRON product is manufactured under the strict quality system of
OMRON HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan for assembly.
INSTRUCTION MANUAL
Automatic
Blood Pressure Monitor
Model
HEM-7112
INTRODUCTION
Intended Use
This product is designed to measure the blood pressure and pulse rate of people within
the range of the designated arm cuff, following the instructions in this instruction
manual. It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Your new blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood’s movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
monitor does not need a stethoscope so the monitor is simple to use.
Thank you for purchasing the OMRON HEM-7112 Automatic Blood Pressure Monitor.
5331395-1B
HEM-7112
Four “AAA” alkaline
(LR03) batteries
OPTIONAL PARTS
Maintenance
To protect your unit from damage, please observe the following:
Storage
Keep the unit in its storage case when not in use.
1 .
Unplug the air tube from the air connector.
2 .
Gently fold the air tube into the arm cuff.
Note:
Do not bend the air tube excessively.
Do not store the unit in the following situations:
1 .
2 .
Using the Optional AC Adapter
Insert the AC adapter plug
into the AC adapter jack on
the side of the main unit.
Plug the AC adapter into an
electrical outlet.
Medium Arm Cuff
Arm circumference
22 - 32 cm
AC Adapter “R”
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then remove the AC adapter plug from the
main unit.
• Do not subject the main unit and the cuff to extreme temperatures, humidity,
moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations (for example, dropping
the unit on the floor).
• Do not use volatile liquids to clean the main unit. The unit should be cleaned
with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm cuff.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean the arm cuff.
Calibration and Service
• The accuracy of this blood pressure monitor has been carefully tested and is
designed for a long service life.
• It is generally recommended to have the unit inspected every two years to
ensure correct functioning and accuracy. Please consult your local
authorised OMRON distributor or dealer.
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight, dust
or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an angle.
• Do not carry out repairs of any kind yourself. If a defect occurs, consult
your local authorised OMRON distributor or dealer.
= Type B
Internally powered ME equipment
(When using only the batteries)
= Class II ME equipment
(Optional AC adapter)
Large Arm Cuff
Arm circumference
32 - 42 cm
Note: If your systolic or diastolic
pressure is outside the standard
range (above 135/85 mmHg), the
Heartbeat symbol ( ) will blink.
Refer to “Taking a Measurement”.
1. Flashes during measurement
2. If flashing after measurement
complete indicates blood pressure
out of recommended range
• If your systolic or diastolic pressure is outside the standard range,
the heartbeat symbol will blink when the measurement result is
displayed. Recent research suggests that the following values can
be used as a guide to high blood pressure for measurements taken
at home.
Important:
Systolic Blood Pressure Above 135 mmHg
Above 85 mmHg
Diastolic Blood Pressure
INSTRUCTION MANUAL
Automatic
Blood Pressure Monitor
Model
HEM-7112
INTRODUCTION
Intended Use
This product is designed to measure the blood pressure and pulse rate of people within
the range of the designated arm cuff, following the instructions in this instruction
manual. It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Your new blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood’s movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
monitor does not need a stethoscope so the monitor is simple to use.
Thank you for purchasing the OMRON HEM-7112 Automatic Blood Pressure Monitor.
5331395-1A
HEM
-711
2