Important:

 Recent research suggests that the following 

values can be used as a guide to high blood 

pressure for measurements taken at home.

Systolic Blood Pressure  Above 135 mmHg
Diastolic Blood Pressure  Above 85 mmHg

  This criteria is for home blood pressure 

measurement.

 Your blood pressure monitor includes an irregular 

heartbeat feature. Irregular heartbeats can influence 

the results of the measurement. The irregular 

heartbeat algorithm automatically determines if the 

measurement is usable or needs to be repeated. If 

the measurement results are affected by irregular 

heartbeats but the result is valid, the result is shown 

together with the irregular heartbeat symbol (     ). If 

the irregular heartbeats cause the measurement to 

be invalid, no result is shown. If the irregular 

heartbeat symbol (     ) is shown after you have taken a measurement, 

repeat the measurement. If the irregular heartbeat symbol (     ) is 

shown frequently, please make your doctor aware of it.

Important Safety Information

5325545-5C

Introduction

1.   Overview

Main unit:

Display:

2.1   Installing/Replacing the Batteries

2.   Preparation

1.

2.

1.

2.

4.   Troubleshooting and Maintenance

1.

2.

3.

3.3   Taking a Reading

5.   Technical Data

3.2   How to Sit Correctly

Relax your wrist and hand.

4.1   The Icons and Error Messages

Error Display

4.2   Troubleshooting

Problem

3.4   Using the Memory Function

Cause

Remedy

Cause

Remedy

Do not bend 

your wrist back.

Do not clench 

your fist.

Do not bend 

your wrist forward.

Wrist Blood 

Pressure Monitor

Model HEM-6200

Instruction Manual

Thank you for purchasing the OMRON HEM-6200 Wrist Blood Pressure 

Monitor.
The OMRON HEM-6200 is a compact and easy to use blood pressure monitor, 

operating on the oscillometric principle. It measures your blood pressure and 

pulse rate simply and quickly. For comfortable controlled inflation without the 

need of pressure pre-setting or re-inflation the devices uses its advanced 

“IntelliSense” technology.
Please read this instruction manual thoroughly before using the unit.

Please keep for future reference.

For specific information about your own blood pressure, CONSULT YOUR 

DOCTOR.

Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia 

or arteriosclerosis.

Please read this section carefully before using the unit.

Warning:

 Indicates a potentially hazardous situation which, if not avoided, could result 

in death or serious injury.

(General Usage)

 Always consult your doctor. Self-diagnosis of measurement results and 

self-treatment are dangerous.

 People with severe blood flow problems, or blood disorders, should consult a 

doctor before using the unit, as cuff inflation can cause internal bleeding.

(Battery Usage)

 If battery fluid should get in your eyes, immediately rinse with plenty of clean 

water. Consult a doctor immediately.

Caution:

 Indicates a potentially hazardous situation which, if not avoided, may result in 

minor or moderate injury to the user or patient or damage to the equipment or 

other property.

(General Usage)

 Do not leave the unit unattended with infants or persons who cannot express 

their consent.

 Do not use the unit for any purpose other than measuring blood pressure.

 Do not disassemble the unit or wrist cuff.

 Do not inflate the wrist cuff over 299 mmHg.

 Do not use a mobile phone or other devices that emit electromagnetic fields, 

near the unit. This may result in incorrect operation of the unit.

 Do not operate unit in a moving vehicle (car, airplane).

(Battery Usage)

 If battery fluid should get on your skin or clothing, immediately rinse with 

plenty of clean water.

 Use only two “AAA” alkaline (LR03) batteries with this unit. Do not use other 

types of batteries.

 Do not insert the batteries with their polarities incorrectly aligned.

 Replace old batteries with new ones immediately. Replace both batteries at 

the same time.

 Remove the batteries if the unit will not be used for three months or more.

 Do not use new and used batteries together.

General Precautions

 Do not apply strong shocks and vibrations to or drop the unit.

 Do not take measurements after bathing, drinking alcohol, smoking, 

exercising or eating.

 Do not inflate the wrist cuff when it is not wrapped around your wrist.

 Do not wash the wrist cuff or immerse it in water.

 Read and follow the “Important information regarding Electro Magnetic 

Compatibility (EMC)” in the Technical Data Section.

 Read and follow the “Correct Disposal of This Product” in the Technical Data 

Section when disposing of the device and any used accessories or optional 

parts.

A. Display

B. Battery compartment cover

C. Wrist cuff

D. MEM (Memory) button

E. START/STOP button

F. SET (    ) button

G. Date/Time setting (    ) button

H. Systolic blood pressure

I.  Diastolic blood pressure

J.  Memory symbol

K. Average value symbol

L. Irregular heartbeat symbol

M. Pulse display

N. Movement error symbol 

O. Battery low symbol 

P. Date/Time display 

Q. Heartbeat symbol 

  (Flashes during measurement.)

R. Deflation symbol

S. Blood pressure level indicator

Remove the battery cover.

1) Push the hook on the bottom of 

the battery cover.

2) Pull the cover off the main unit.

Insert two 1.5V “AAA” alkaline (LR03)

batteries as indicated in the battery

compartment and then replace the

battery cover.

Notes:

 If the battery low symbol (     ) appears on the display, turn the unit off 

then replace both batteries at the same time.

 The measurement values continue to be stored in memory even after 

the batteries are replaced.

Disposal of used batteries should be carried out in accordance with the 

national regulations for the disposal of batteries.

Set the monitor to the correct date and time before

taking a measurement for the first time.
Press the

button.

Press the START/STOP button to store the setting.

Notes:

 To reset the date and time, hold down the      button while the power 

is off.

 If the batteries have been removed for 30 seconds or more, the date 

and time setting will need to be reset.

 If the date and time are not set, “-:--” appears during or after 

measurement.

1) Push the     button to change.

 Hold down to advance the 

digit rapidly.

2) Push the     button to confirm.

Next setting appears.

Year

Month

Day

Hour

Minute

To take a measurement, you need to be relaxed and comfortably seated, under 

comfortable room temperature. No eating, smoking or exercising 30 minutes 

before taking a measurement.

 Sit on a chair with your feet flat on the floor.

To delete all the values stored in memory

When the memory symbol (     ) appears, first press the MEM button. Then 

while holding it down, press the START/STOP button simultaneously for 

about 2-3 seconds.

Note:  You cannot partially delete the stored readings.

Irregular or weak pulses 

are detected.

Movement during 

measurement.

The batteries are low. 

The batteries are 

exhausted.

Remove the wrist cuff. Wait 

2-3 minutes and then take 

another measurement. 

Repeat the steps in section 

3.3. If this error continues to 

appear, contact your doctor.

Carefully read and repeat the 

steps in section 3.3.

You should replace them with 

new ones ahead of time.

Refer to section 2.1.

You should replace them with 

new ones at once.

Refer to section 2.1.

Cuff is over inflated.

Movement during 

measurement.

Carefully read and repeat the 

steps listed under section 3.3.

An Er mark with a 

code/number indicates 

the device has a 

hardware failure.

The reading is 

extremely low (or 

high).

The wrist cuff is not at 

heart level.

The cuff is wrapped 

snugly around the 

wrist.

Your arms and 

shoulders are tense.

Movement or talking 

during measurement.

Measure while in the 

correct posture. Refer 

to section 3.2.
Wrap the cuff correctly. 

Refer to section 3.1.

Relax and try taking the 

measurement again. 

Refer to section 3.3.

Wrist cuff pressure 

does not rise.

Air is leaking from the 

wrist cuff.

Consult your OMRON 

retail outlet or 

distributor.

Wrist cuff deflates too 

soon.

The wrist cuff is 

loose.

Apply the cuff correctly 

so that it is securely 

wrapped around the 

wrist. Refer to section 

3.1.

The unit loses power 

during measurement.

Nothing happens 

when you press the 

buttons.

The batteries are 

drained.

Replace the batteries 

with new ones.

The batteries are 

drained.

Replace the batteries 

with new ones.

The batteries have 

been inserted 

incorrectly.

Insert the batteries with 

the correct (+/-) polarity.

Other problems.

 Press the START/STOP button and repeat 

measurement.

 If the problem continues, try replacing the 

batteries with new ones.

If this still does not solve the problem, contact 

your OMRON retail outlet or distributor.

The blood pressure is different each time. 

The reading is extremely low (or high).

Blood pressure readings 

constantly vary with time 

of day and how relaxed 

you are. Take several 

deep breaths and try to 

remain relaxed before 

taking a measurement.

Remain still and do not 

talk during 

measurement. Refer to 

section 3.3.

Consult your OMRON

retail outlet or distributor.

Sit upright with your 

back straight.

The cuff should be at the 

same level as your heart.

First

Second

Blink

Lit

Do not use your other hand 

to support the wrist cuff. 

This can result in inaccurate 

measurement results.

Notes:

 To cancel a measurement, press the START/STOP button at any time 

during measurement.

 Remain still while taking a measurement.

Press the START/STOP button.

The wrist cuff will start to inflate automatically.

Undo the wrist cuff and remove the unit.
Press the START/STOP button to turn off the monitor.

The monitor automatically stores the measurement in its memory.

It will automatically turn off after two minutes.

The monitor automatically stores the result up to 60 sets. It can also 

calculate an average reading based on the measurements from the last 

three readings taken within 10 minutes. If there are only two readings in 

memory for that period, the average will be based on two readings. If 

there is one reading in memory for that period, the average will be 

based on one reading.
Notes:

 If the memory is full, the monitor will delete the oldest readings.

 When viewing the reading taken without setting the date and time, 

“-:--” is displayed instead of the date and time.

START

INFLATE

DEFLATE

COMPLETE

Alternating

date/time

display

Average Value

Previous Reading

Alternating

date/time display

View the Average 

Value

  If the previous reading 

was taken without 

setting the date and 

time, the average 

value is not calculated.

View Previous Reading

  The Memory number 

appears for a second 

before the pulse rate is 

displayed. The newest 

set is numbered “1”,

View Readings Stored 

in Memory

  Hold down to view 

readings rapidly.

4.3   Maintenance

To protect your unit from damage, please avoid the following:

 Subjecting your unit to extreme temperatures, humidity, 

or direct sunlight.

 Washing the cuff or exposing the cuff or unit to water. 

 Disassembling the unit.

 Subjecting the unit to strong shocks or vibrations. 

Dropping the Unit.

 Cleaning the unit with volatile liquids. THE UNIT 

SHOULD BE CLEANED WITH A SOFT, DRY CLOTH.

Use a soft, moistened cloth and soap to clean the cuff.

Keep the unit in its storage case when not in use.

Fold the cuff into the storage case.

Do not store the unit in the following situations:

 If the unit is wet.

 Locations exposed to extreme temperatures, humidity, 

direct sunlight, dust or corrosive vapours.

 Locations exposed to vibrations, shocks or where it will 

be at risk of falling.

Calibration and Service

 The accuracy of this blood pressure monitor has been 

carefully tested and is designed for a long service life.

 It is generally recommended to have the unit inspected 

every two years to ensure correct functioning and 

accuracy. Please consult your local authorised OMRON 

distributor or dealer.

Product Description

Model

Display

Measurement Method

Measurement Range

Accuracy

Inflation

Deflation

Memory

Power Source

Battery Life

Applied Part

Protection Against 

Electric Shock

Operating temperature/ 

Humidity

Storage temperature/ 

Humidity/ Air pressure

Console Weight

Outer Dimensions

Measurable

circumference of wrist

Cuff Material

Package Content

Wrist Blood Pressure Monitor

HEM-6200

LCD Digital Display

Oscillometric method

Pressure: 0 mmHg to 299 mmHg

Pulse: 40 to 180 beats/min.

Pressure: ±3 mmHg

Pulse: ±5% of reading

Automatic inflation by pump

Automatic rapid deflation

60 Measurements

2 “AAA” alkaline (LR03) batteries 1.5V

Approx. 300 measurements with new alkaline 

batteries at a room temperature of 23°C

Internally powered ME equipment

+10°C to +40°C / Maximum: 30 to 85% RH

-20°C to +60°C / Maximum: 10 to 95% RH /

700-1060 hPa

Approximately 117g without batteries

Approximately 71 (w) mm × 41 (h) mm × 70 (d) mm 

without the wrist cuff

Approximately 13.5 to 21.5 cm

Nylon and polyester

Main unit, storage case, two “AAA” alkaline (LR03) 

batteries, instruction manual

= Type B

Notes:

 Subject to technical modification without prior notice.

 Disposal of this product and used batteries should be carried out in 

accordance with the national regulations for the disposal of electronic 

products.

1.

2.

2.2   Setting the Date and Time

3.   Using the Unit

3.1   Applying the Wrist Cuff

Do not apply over clothing.

Place the wrist cuff over your wrist.

Wrap the wrist cuff around your wrist.

Your thumb should face upward.

Make sure that the wrist cuff does not cover 

the protruding part of the wrist bone (ulna) on 

the outside of the wrist.

Notes:

 You can take a measurement on either 

your left or right wrist.

 The blood pressure can differ between the right 

arm and the left arm, and therefore also the 

measured blood pressure values can be 

different. Omron recommends to always use 

the same arm for measurement. If the values 

between the two arms differ substantially, 

please check with your doctor which arm to use 

for your measurement.

 Wrap the wrist cuff securely around the 

wrist for taking accurate measurements.

ulna

Please read this instruction manual carefully before using the device.

 This device fulfils the provisions of EC directive 93/42/EEC (Medical Device

Directive).

 This blood pressure monitor is designed according to the European Standard

EN1060, Noninvasive sphygmomanometers Part 1: General Requirements 

and Part 3: Supplementary requirements for electromechanical blood

pressure measuring systems.

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile

(cellular) telephones, medical devices in use may be susceptible to 

electromagnetic interference from other devices.

Electromagnetic interference may result in incorrect operation of the medical

device and create a potentially unsafe situation.

Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)

with the aim to prevent unsafe product situations, the EN60601-1-2:2007

standard has been implemented. This standard defines the levels of immunity to 

electromagnetic interferences as well as maximum levels of electromagnetic

emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this

EN60601-1-2:2007 standard for both immunity and emissions.

Nevertheless, special precautions need to be observed:

Do not use mobile (cellular) telephones and other devices, which generate

strong electrical or electromagnetic fields, near the medical device. This may

result in incorrect operation of the unit and create a potentially unsafe situation.

Recommendation is to keep a minimum distance of 7 m. Verify correct

operation of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at OMRON 

HEALTHCARE EUROPE at the address mentioned in this instruction manual.

Documentation is also available at www.omron-healthcare.com.

This marking shown on the product or its literature, indicates that it 

should not be disposed of, with other household wastes at the end

of its working life. To prevent possible harm to the environment or

human health from uncontrolled waste disposal, please separate 

this from other types of wastes and recycle it responsibly to 

promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this

product, or their local government office, for details of where and how they can

take this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions

of the purchase contract. This product should not be mixed with other commercial

wastes for disposal.
This product does not contain any hazardous substances.

Correct Disposal of This Product

(Waste Electrical & Electronic Equipment)

Asia Pacific HQ

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, Kyoto

617-0002 JAPAN

OMRON HEALTHCARE EUROPE B.V.

Scorplus 33, 2132 LR Hoofddorp 

THE NETHERLANDS

OMRON HEALTHCARE SINGAPORE PTE LTD.

438A Alexandra Road, #05-05/08

Alexandra Technopark, Singapore 119967

www.omron-healthcare.com.sg