8
Appendixes
FZ4 User's Manual
Error Messages and Troubleshooting
571
Error Messages and Troubleshooting
This section lists error messages that display on the screen and solutions.
When a message with this symbol is displayed, the ERROR signal on the parallel interface is on.
Error message
Remedy
No value set. Please
set value.
A value must be set for this item. Please set up value.
Error of the expression.
Do the following errors exist in settings for an expression set up?
●
Different number of open/close parentheses
●
TJG/unit/value/function/operator/comma is lined next to each other.
●
Operator is set at beginning or end of expression
●
The number of function operands is insufficient or there are more than is needed
Cannot add object. Out
of range of specified
characters of
expression.
Incompliant to the conditions of expression. Please check the content.
Reference:
"Processing Item List Manual", "Setting (Calculation)" (p.452)
The problem occurred
in the camera
connection.
Is the camera cable connected to the controller?
Is camera cable disconnected?
Do [Select camera] of [Camera Image Input] and [Camera Switching] have any errors?
Please check the contents of [Camera setting], then switch off and restart.
Reference:
Setting Conditions for Camera Use (p.342)
Failed to start the
window. Memory is
insufficient.
Because memory available for use was insufficient, memory needed to open the window
could not be secured. Check the flow by adjusting the measurement region of each
processing unit and the sizes of the registered models.
●
[Shape search]
If shape setup window is opened while the RAMDisk memory is insufficient
for image logging, etc. on the RAMDisk, a warning message is displayed. In the case
that an image logging file is stored in a RAMDisk, please move the content on the
RAMDisk to a USBDisk to open up memory and reopen the set up window.
Warning range shall be
set within the range
from judgement lower
to upper limit.
Please redefine the upper and lower warning range limits such that they are within the range
between the lower and upper evaluation limits.
Reference:
"Processing Item List Manual", "Judgement conditions (Trend Monitor)" (p.510)
The "Search"
measurement result
may be "NG
(Insufficient memory)"
with the current setting.
Insufficient memory may occur during measurement. Reset the model parameters.
●
[Search]
Reference:
"Processing Item List Manual", "Judgement is NG (Insufficient Memory)" (p.66)
●
[Flexible search]
Reference:
"Processing Item List Manual", "Judgement is NG (Insufficient Memory)" (p.77)
●
[Classification]
Reference:
"Processing Item List Manual", "Judgement is NG (Insufficient Memory)" (p.148)
The problem occurred
in the system.
This is displayed when the significant abnormality occurs in the controller system. Please
contact one of our branches or regional offices.
The problem occurred
in system date. The
system battery is
insufficient.
Low battery (for the backup of date and time data) level. The batteries need to be
replaced. Please return old batteries to one of our branches or regional offices.
Failed to switch scene
Is a USBDisk set to the controller?
A USBDisk is needed to read scenes after scene group 1.
Acesst 1.0 6251194
Содержание FZ4 Series
Страница 1: ...User s Manual Cat No Z318 E1 02B Vision Sensor FZ4 Series...
Страница 12: ...1 Before Operation 10 Operation Flow FZ4 User s Manual Operation Flow Here describes the operation flow...
Страница 38: ...1 Before Operation 36 Setting Operation Mode FZ4 User s Manual 1 CPU 2 CPUs FZ4 11 H11 only...
Страница 112: ...3 Performing Test Measurement Starting Operation 110 Useful Functions for Operation FZ4 User s Manual...
Страница 326: ...4 Using Tool 324 Remotely Operating the Controller Remote Operation FZ4 User s Manual Enter the IP address...
Страница 330: ...4 Using Tool 328 Remotely Operating the Controller Remote Operation FZ4 User s Manual...
Страница 351: ...6 Changing the System Environment FZ4 User s Manual Setting the System Operation Environment 349 Communication...
Страница 566: ......
Страница 648: ...8 Appendixes 646 Manual Revision History FZ4 User s Manual...
Страница 654: ...652 FZ4 User s Manual...