6
M
eth
od
s
fo
rC
on
ne
ctin
g
an
d
C
om
m
un
ica
tin
g
w
ith
E
xte
rn
al
D
ev
ice
s
184
Controlling/Outputting through Parallel Communication
FZ3 User's Manual
●
Scene/Scene Group Switch (p.189)
●
About multi-input function (p.190)
When "Output Control" Is Set to "None"
The controller outputs measurement results without synchronizing with external devices. Have the
external devices detect the GATE signal of the controller, and load measurement results during ON
status.
Inputting a measurement trigger in the STEP signal
Example) When 3 expressions are set in [Parallel Data Output]:
Output signal
Signal
Function
RUN
Turns on when the controller is available for measurement and the RUN window is displayed. It is
OFF in the ADJUST window, so switch to the RUN window during operation.
BUSY
Indicates that controller is currently measuring or switching the scene. Do not input next command
while the BUSY signal is ON. Otherwise, on-going processing or commands that are input will not
be performed correctly.
OR
Outputs overall judgement.
Selection of whether ON occurs during an OK judgement result or NG error judgement result can be
performed in the communication specifications settings window. Reference:
Communication Specifications (Parallel Interface) (p.178)
DO
Outputs the results for expressions set in the processing item [Parallel Judgement Output]/ [Parallel
Data Output].
Selection of whether ON occurs during an OK judgement result or NG error judgement result can be
performed in the communication specifications settings window. Reference:
Communication Specifications (Parallel Interface) (p.178)
Acesst 1.0 6251194
Содержание FZ3 Series
Страница 10: ...1 Before Operation 8 Operation Flow FZ3 User s Manual Operation Flow Operation flow is explained here...
Страница 28: ...1 Before Operation 26 Saving Settings and Turning Power Off FZ3 User s Manual...
Страница 60: ...2 Setting Scenes Measurement Flow 58 Editing Scene Groups FZ3 User s Manual...
Страница 106: ...4 Saving Loading Data 104 Loading Settings Data to Controller FZ3 User s Manual...
Страница 122: ...5 Changing the System Environment 120 Setting the System Operation Environment FZ3 User s Manual...