8. Communication Test and Collecting Start
8-3
5) To verify the communication test result, select [Operation Monitor] in the setting menu to view.
Please wait for a while before a collecting cycle of a measurement device passes.
The communication test result appears in the [Status] field. Make sure that it should be
[Logging].
Status
Description
--
Appears before EQ100 checks the measurement device status.
Stopped
Appears after EQ100 stopped the measurement device logging.
Logging
Appears while EQ100 is performing logging/communication test of a
measurement device.
Error
Appears when communication is successful between EQ100 and a
measurement device but measured data could not be collected.
Communication
Error
Appears when communication is not successful between EQ100 and a
measurement device.
Ending Communication Test
8.2.2.
■
Steps
1) On the toolbar, click [Logger] - [Stop Test].
The communication test ends.
2) To cut the connection between EQ-Manager and EQ100, on the toolbar click [Logger] -
[Offline].