background image

EN

Intended use

Medical Purpose

This product is intended to be used for inhaling medication for respiratory 
disorders.

Intended User

• Legally certified medical experts, such as doctor, nurse and therapist.
• Caregiver or patient under the guidance of qualified medical experts for 

home treatment.

• The user should also be capable of understanding general operation of 

NE-C28P and the content of instruction manual.

Intended Patients

This product should not be used on patients, who are unconscious or are 
not breathing spontaneously.

Environment

This product is intended for use in a medical facility, such as Hospital, clinic 
and doctor’s office, and in a room of general household.

Durable period

Durable periods are as follows, provided the product is used to nebulize 
saline 3 times a day for 10 minutes each time at room
temperature (23°C).
Durable period may vary depending on usage environment.
 

Compressor (Main unit) 

5 years

 

Nebulizer Kit 

1 year

 

Mouthpiece 

1 year

 

Nosepiece 

1 year

 

Air Tube (PVC, 200cm) 

1 year

 

Air Filter 

60 days

 

Mask (PVC)/(SEBS) 

1 year

Precautions for use

Warnings and cautions described in the instruction manual should be 
observed.

Know your unit

Important safety instructions

Read all the information in the instruction manual and any other literature included in 
the box before using the device.

 

Warning

:

Indicates a potentially hazardous situation which, if not avoided, could result in 
serious injury.

(Usage)

•  For type, dose, and regime of medication follow the instructions of your doctor or 

respiratory therapist.

•  If you feel anything unusual during use, stop using the device immediately and 

consult your doctor.

•  Do not use only water in the nebulizer for inhaling purposes.
•  Keep the device out of the reach of unsupervised infants and children. The device 

may contain small pieces that can be swallowed.

•  Do not store the air tube while there is moisture or medication remaining inside.
•  Do not use or store the device where it may be exposed to noxious fumes or volatile 

substances.

•  Do not use the device where it may be exposed to flammable gas.
•  Do not cover the compressor with a blanket, towel, or any other type of cover during 

use.

•  Always dispose of any remaining medication after use, and use fresh medication each 

time.

•  Do not use in anaesthetic or ventilator breathing circuits.
•  Do not block the ventilation slots. Never place the device where the ventilation slots 

may be obstructed during operation.

•  Store the parts in a clean location for avoiding infection.

(Risk of electrical shock)

•  Never plug or unplug the device with wet hands.
•  The compressor and power cord are not waterproof. Do not spill water or other liquids 

on these parts. If liquid does spill on these parts, immediately unplug the power cord 
and wipe off the liquid with gauze or other absorbent material.  

•  Do not immerse the compressor in water or other liquid.
•  Do not use or store the device in humid locations.
•  Do not operate the device with a damaged power cord or plug.
•  Keep the power cord away from heated surfaces.

(Cleaning and disinfecting)

Observe the rules below when cleaning or disinfecting parts. Failure to observe these 
rules may result in damage, inefficient nebulization or infection. For the instructions, 
refer to “Cleaning and disinfecting” section.
•  Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using 

them when as follows:
- the first time after purchase
- if the device has not been used for a long period of time
- if more than one person uses the same device

•  Be sure to wash or wipe the parts after use, and ensure that they are thoroughly 

disinfected and dried, and stored in a clean location.

•  Do not leave the cleaning solution in the parts.

 

Caution:

Indicates a potentially hazardous situation which if not avoided, may result in 
minor or moderate injury, or physical damage.

(Usage)

•  Provide close supervision when this device is used by, on, or near children or invalids.
•  Make sure that the parts are attached correctly.
•  Make sure that the air filter is correctly attached.
•  Make sure that the air filter is clean. If the air filter has been used for more than 

60 days, replace it with a new one.

•  Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions or 

shake it while in use.

•  Do not use or store the device while the air tube is creased.
•  Use only original nebulizing parts, air tube, air filter and air filter cover.
•  Do not add more than 7 ml of medication to the medication tank.
•  Do not carry or leave the nebulizer kit while the medication tank contains medication.
•  Do not leave the device unattended with infants or persons who cannot express their 

consent.

•  Do not subject the device or the parts to any strong shocks such as dropping the device 

on the floor.

•  Do not distort the vaporiser head, or jab the nozzle of the medication tank with a pin 

or any sharp object.

•  Do not insert fingers or objects inside the compressor.
•  Do not disassemble or attempt to repair the compressor or power cord.
•  Do not leave the device or its parts in extreme hot or cold temperature, or under direct 

sunlight.

•  Do not block the air filter cover.
•  Do not use the device while sleeping or if drowsy.
•  Approved for human use only.
•  The compressor may become hot during operation.
•  Do not touch the compressor for other than necessary operation such as turning off 

the power while nebulizing.

•  To avoid the medication residue on the face, be sure to wipe the face after removing 

the mask.

•  To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back of 

the nose.

•  Do not block the slit between the cap and the inhalation air inlet.
•  Make sure that the vaporiser head is correctly installed before use.
•  Do not use a damaged nebulizer kit, mouthpiece or nosepiece.

(Risk of electrical shock)

•  Always unplug the device from the power outlet after use and before cleaning.
•  Plug the device into the appropriate voltage outlet. Do not overload power outlets or 

use extension cords.

•  Do not misuse the power cord.
•  Do not wind the power cord around the compressor.
•  Changes or modifications not approved by OMRON HEALTHCARE will void the user 

warranty.

(Cleaning and disinfecting)

Observe the rules below when cleaning or disinfecting parts. Failure to observe these 
rules may result in damage, inefficient nebulization or infection. For the instructions, 
refer to “Cleaning and disinfecting” section.
•  Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.
•  Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.
•  When disinfecting parts by boiling, make sure that the container does not boil dry.

Otherwise it may also result in fire.

Optional Medical Accessories

(within the scope of EC Medical Device Directive 93/42/EEC)

Name

Model

Order No.

V.V.T. Nebulizer Kit N-01 & 
Mouthpiece Set

NEB-NSET3-81E

9515573-6

Adult Mask (PVC)

NEB-MSLP-E

9956275-1

Child Mask (PVC)

NEB-MSMP-E

9956276-0

Adult Mask (SEBS)

NEB-MSLS-E

9956312-0

Child Mask (SEBS) Set

NEB-MSSS-E

9956281-6

Mouthpiece

NEB-MP-E

9956273-5

Nosepiece

NEB-NP-E

9956274-3

Air Tube (PVC, 200cm)

NEB-TP-L20E

9520606-3

Air Tube (Silicone, 100cm) 

NEB-TSC-L10E

9956269-7

Other Optional/Replacement Parts

Name

Model

Order No.

Air Filters x 5

NEB-AFR-28E

9956271-9

Thank you for purchasing the OMRON COMP AIR.
This product was developed in conjunction with respiratory therapists 
for the successful treatment of asthma, chronic bronchitis, allergies 
and other respiratory disorders. 
This is a medical device. Operate the device only as instructed by 
your doctor and/or respiratory therapist.

How the nebulizer kit works

The medication that is pumped up 
through the medication channel is 
mixed with compressed air which is 
generated by a compressor pump. 
The compressed air mixed with 
medication is turned into fine particles 
and sprayed when in contact with the 
baffle. 

Manufacturer

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN

EU-representative

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
www.omron-healthcare.com

Production facility OMRON (DALIAN) CO., LTD.

Dalian, CHINA

Subsidiary

OMRON HEALTHCARE UK LTD.

Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.

Made in China

General Safety Precautions:

•  Inspect the device and parts before using them each time, and check that there are 

no problems. In particular, be sure to check the following:
- That the nozzle or air tube are not damaged.
- That the nozzle is not blocked.
- That the compressor operates normally.

•  When using this device, there will be some noise and vibration caused by the pump in 

the compressor. There will also be some noise caused by the emission of compressed 
air from the nebulizer kit. This is normal and does not indicate a malfunction.

•  Operate the device only as intended. Do not use the device for any other purpose.
•  Do not use the device at temperatures greater than +40°C.
•  Make sure that the air tube is securely attached to the compressor and nebulizer kit, 

and does not come loose. Twist the air tube plug slightly when inserting it into the air 
tube connectors to avoid the air tube disconnecting during use.

•  To completely isolate the device from the power source unplug the plug from the 

power source.

Keep these instructions for future reference.

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile (cellular) 
telephones, medical devices in use may be susceptible to electromagnetic interference 
from other devices. Electromagnetic interference may result in incorrect operation of the 
medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with 
the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has 
been implemented. This standard defines the levels of immunity to electromagnetic 
interferences as well as maximum levels of electromagnetic emissions for medical 
devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this 
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
•  Do not use mobile (cellular) telephones and other devices, which generate strong 

electrical or electromagnetic fields, near the medical device. This may result in 
incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation of 
the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at OMRON 
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.

 

Correct Disposal of This Product

 

(Waste Electrical & Electronic Equipment)

This marking shown on the product or its literature, indicates that it should not be 
disposed of, with other household wastes at the end of its working life. To prevent 

possible harm to the environment or human health from uncontrolled waste disposal, 

please separate this product from other types of wastes and recycle it responsibly to 

promote the sustainable reuse of material resources.

Household users should contact either the retailer where they purchased this product, 

or their local government office, for details of where and how they can return this item 

for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions of the 

purchase contract. This product should not be mixed with other commercial wastes for 

disposal.

NE-C28P-UK_A_M02_130305.pdf

TM

Rear View

Power Cord

Ventilation Slots

Power Plug

Ventilation Slots

Power Switch

Air Filter Cover

Air Tube Connector

Nebulizer Kit Holder

Cap

Vaporiser Head

Compressor Nebulizer

Model 

NE-C28P

Instruction Manual

Baffle

Aerosol

Medication 

Channel

Medication 

Aerosol

Nozzle

Medication 

Compressed Air

Compressor (Main Unit) x 1

Nebulizer Kit x 1

Nosepiece x 1

Mouthpiece x 1

Adult Mask (PVC) x 1

Child Mask (PVC) x 1

Air Tube 

(PVC, 200cm) x 1

Carrying Bag x 1
Instruction Manual
Warranty card

Spare Air Filters x 5

IM-NE-C28P-UK-EN-01-03/2013

1679307-2A

Inhalation Air Inlet

Inhalation Top

Nozzle

Medication Tank

Air Tube Connector

13C0539

Отзывы: