114
Cobra 350 User's Guide
05624-000 Rev. K
8.9 Commissioning a System with an eMB-40R
E-Stop Configuration Utility
This utility sets the E-Stop hardware delay to factory specifications.
NOTE:
Ensure that the commissioning jumper is plugged into the XBELTIO jack
on the eMB-40R before you start this procedure.
Procedure
NOTE:
In this procedure, the word “controller” means the SmartController EX,
if there is one in your system, or, if not, the eMB-40R.
From within the ACE software:
1. Open the robot object editor.
2. Select
Configure > Safety Settings > Configure ESTOP Hardware Delay
, then click
Next.
This procedure will configure Channel A and then Channel B.
It will then report the delay that it set for each.
3. Reboot the controller.
On some systems, the controller will reboot automatically.
4. Reboot the eMB-40R.
E-Stop Verification Utility
This utility verifies that the hardware E-Stop parameters are set correctly and that the hard-
ware E-Stop is working.
The hardware E-Stop must have already been configured for this wizard to run.
NOTE:
If the commissioning jumper is plugged into the XBELTIO jack on the
eMB-40R, remove it before you start this procedure.
Procedure
From within the ACE software:
1. Open the robot object editor.
2. Select
Configure > Safety Settings > Verify ESTOP Hardware Delay
, then click Next.
3. Enable high power, if not already enabled, then click Next.
4. Press an E-Stop button (on the Front Panel), then click Next.
The utility will confirm that the hardware delay has been verified for this robot, and dis-
play the delay times for channels A and B.
5. Reboot the controller.
On some systems, the controller will reboot automatically.