Automatic Blood Pressure

Monitor & Printer

The World Hypertension 

League recommends regular blood

pressure monitoring with clinically

validated devices.

705CP-II

IM-HEM-759-E2-02-10/2011

1628532-8F

Содержание 705CP-II

Страница 1: ...5 Mode d emploi page 67 Istruzioni per l uso pagina 89 Instrucciones de uso pagina 111 Automatic Blood Pressure Monitor Printer The World Hypertension League recommends regular blood pressure monitori...

Страница 2: ...tallation replacement 18 How to connect printer unit 19 How to load printer paper 19 How to use the printer unit 20 Troubleshooting 22 Specifications 22 Thank you for purchasing an OMRON product You c...

Страница 3: ...ies The pulse display is not suitable for monitoring the frequency of cardiac pacemakers If you suffer from disorders of heart rhythm known as arrythmia you should only use this blood pressure monitor...

Страница 4: ...d of oscillometric measurement and does not require pressure pre setting or re inflation OMRON is the world s leading manufacturer of blood pressure monitors with the goal to always fulfill the need f...

Страница 5: ...f the battery compartment as indicated Replace the battery cover NOTES If the Low Battery Indicator appears on the display replace all four batteries Long life alkaline batteries are recommended Remov...

Страница 6: ...SET button The year digits 2002 will flash Press the ADJUST button to advance the digit s one at a time If you hold down the ADJUST button the digits will advance rapidly Press SET button when desire...

Страница 7: ...er arm Sit in a chair with your feet flat on the floor and place your arm on a table so that the cuff is at the same level as your heart Put your arm through the cuff loop The bottom of the cuff shoul...

Страница 8: ...e velcro fastener FIRMLY Make certain the cuff fits snugly around your arm The cuff should make good contact with your skin You should be able to fit your index finger between the cuff and your arm ea...

Страница 9: ...t move your arm but remain still until the entire measurement completes If you want to stop the inflation or measurement press the O I button The monitor will stop inflating and start deflating rapidl...

Страница 10: ...e oldest set will be deleted to store a new set Press the O I button to turn the monitor off NOTE If you forget to turn the monitor off it will automatically shut itself off after five minutes Instruc...

Страница 11: ...pressure and pulse rate Turn on the monitor by pressing the O I button Press the button to recall stored readings while the symbol is displayed START Press these two buttons at the same time To delete...

Страница 12: ...a pillow or cushion Coldness raises your blood pressure It is recommended to measure at room temperature 20 C Relax before measurement If necessary go to the bathroom before measurement Do not measure...

Страница 13: ...are weak or exhausted The cuff pressure does not rise although the pump motor can be heard problem with mem function E Er Rectification 1 Repeat the measurement keeping perfectly still Do not move yo...

Страница 14: ...ely high above 60 C or with high humidity 85 Do not use your monitor at very low temperatures below 10 C or very high temperatures above 40 C Do not fold the cuff or air tube together too tightly and...

Страница 15: ...ements Inflation Fuzzy Logic controlled by electric pump Deflation Automatic pressure release valve Pressure Detection Capacitive pressure sensor Measurement Method Oscillometric method Power Source 4...

Страница 16: ...kaging attached literature OMRON devices are especially designed for regular blood pressure monitoring OMRON therefore has direct contact with medical specialists and takes advice on the design and fe...

Страница 17: ...ressure is more often than not a chance finding People with high blood pressure usually feel well unless the blood pressure has been high for some time and complications have occurred Such complicatio...

Страница 18: ...gnosis The blood pressure measured first thing in the morning after getting up before eating and while at rest is known as the fundamental blood pressure In practice it is rather difficult to record t...

Страница 19: ...ting Printer Cover Printer Cable Plug Power Operation Light Battery installation replacement Slide the battery cover off in the direction of the arrow Install or replace 4 LR6 batteries so the positiv...

Страница 20: ...ith a pair of scissors Press the FEED button while inserting the edge of the printing paper into the paper feeding slot When the edge of the paper protrudes sufficiently release the FEED button Close...

Страница 21: ...l look like the following example Press ALL DATA button to print a numerical list and average of all stored readings maximum depending on your blood pressure monitor The numerical print out will list...

Страница 22: ...reading was stored at 3 00 PM then the printout will indicate 15 under the T column Note If Time and Date was not initially set up before taking blood pressure measurement see How to Set Time and Date...

Страница 23: ...ew batteries Nothing is printed No printer paper comes out No printer paper installed Printer paper jammed Ran out of printer paper Specifications Power Source 4 AA batteries Battery Life Approximatel...

Страница 24: ...intechnik de OMRON SANT FRANCE SAS 14 rue de Lisbonne 93561 Rosny sous Bois Cedex FRANCE Uniquement pour le march fran ais OMRON Service Apr s Vente N Vert 0 800 91 43 14 consommateurs omron sante fr...