Omron Microscan LVS-95 Series Скачать руководство пользователя страница 1

LV S - 9 5 X X

LVS -95XX Series: At a Glance

LVS -95XX: Available Symbologies

For more information on this product, visit 

www.microscan.com.

• Offline verification of barcodes to ISO/IEC standards.
• Inspects all nine of the ISO (ANSI) parameters, plus

added features of determining blemishes, opacity, and

human readable validation.

• Verifies 1D and 2D codes and reports all parameters

as specified in the applicable symbology specification.

Linear

2D

Stacked

Standard

Data  

Matrix

Micro  

PDF417

QR  

Code

Micro  

QR Code Aztec

PDF417

GS1 DataBar

Postal

LVS-9510 Desktop 

Barcode Verifier

LVS-9570 Desktop 

Barcode Verifier

LVS

®

 Barcode Verifiers

Installation Qualification 

(IQ)

The purpose of IQ is to 

confirm that the LVS-95XX 

system was installed correct

-

ly. A series of questions are 

provided pertaining to the 

installation of the system. In 

addition, a Calibrated Confor-

mance Standard Test Card for 

GS1 Symbols accompanies an 

LVS-95XX system purchase. 

The Test Card reports on 

several certified parameters. 

The LVS-95XX system must 

demonstrate its ability to stay 

/- 5 percentage points 

of the parameters listed on 

the Test Card. The IQ section 

provides a chart to document 

and verify each Test Card 

parameter. 

Operational Qualification 

(OQ)

The purpose of OQ is to 

demonstrate that LVS-95XX 

system components operate 

correctly within established 

limits and tolerances. OQ 

 

sections include: 

Testing:

 This section is

designed to certify that the

system is working according 

to factory specifications and 

according to the following 

test methods: 

Linear codes

 (1D): 

ISO/IEC 15426-1:2006(E)

Two-dimensional codes 

(2D): 

ISO/IEC 15426-2:2015(E)

To pass the OQ test, each 

test label supplied with the 

package is graded twice. 

The First Test Overall 

Grade and Second Test 

Overall Grade must mea

-

sure /- 0.3 of each 

other’s grade point aver-

age. The labels to be tested 

can be taken from the 

“LVS-95XX Series IQ and 

OQ Validation Guidelines,” 

supplied by the client, or a 

combination of both. The 

labels must have at least a 

1.4 grade point average.

Operational Status: 

This

section provides a series of

questions about the opera-

tional status of the system.

Continued on back

LVS-95XX Series  
IQ/OQ Overview

The “LVS-95XX Series Installation Qualification (IQ) 
and Operational Qualification (OQ) Guideline”  
assists in validating an LVS-95XX Series system.  
The document provides guidelines to determine if  
an LVS-95XX Series system meets IQ and OQ  
specifications and fulfills its intended purpose.  
The document can be modified to fit the  
specifications of the client’s validation protocol.

Отзывы: