Omni Q55 Ultrasonic Скачать руководство пользователя страница 1

Q55 Ultrasonic

Homogenizer

User Manual

Содержание Q55 Ultrasonic

Страница 1: ...Q55 Ultrasonic Homogenizer User Manual...

Страница 2: ...ty of either merchantability or fitness for a particular pur pose No person or company is authorized to change modify or amend the terms of this warranty in any manner or fashion whatsoever Under no c...

Страница 3: ...R immerse the converter in liquids of any kind or let condensed moisture or liquid drip into the converter NEVER grasp an activated horn or touch the tip of a vibrating probe It can cause severe burns...

Страница 4: ...ensions W 8 20 0 cm D 7 5 19 0 cm H 5 75 14 6 cm Output Voltage 250 VRMS Output Frequency 20 KHz AC Adapter Input Voltage 90 to 264 VAC 50 60 Hz 1 5A Output Voltage 24V DC 3 75A Rated Power 90 Watts C...

Страница 5: ...er cord or adapter Trans formers are not recommended Environmental Pollution Degree 2 Installation Category II Operating Limits Shipping Storage Temperature 41 104 F 5 40 C Relative Humidity 10 95 Non...

Страница 6: ...iquid the rapid vibration of the tip causes cavitation the formation and violent col lapse of microscopic bubbles The collapse of thousands of cavitation bubbles releases tremendous energy in the cavi...

Страница 7: ...RIPTIONS OF COMPONENTS FUNCTIONS OF CONTROLS 5 1 Q55 FRONT PANEL 5 2 Q55 REAR PANEL Power Supply Convertor Cable Convertor Probe Lab Stand with Clamp Convertor Cable Connector Footswitch Jack DC Power...

Страница 8: ...itch Continuous Position Probe tip is energized when the amplitude is set above the red area on the Amplitude Control Knob Thumb switch Position Probe tip is energized once the thumb switch or footswi...

Страница 9: ...4 CONVERTER CLAMPING Improper clamping can damage the system and void the warranty Using a sound en closure part 060 12 or stand part 060 10 will ensure a proper fit Proper Clamping Improper Clamping...

Страница 10: ...ersal 98 264 V 47 63Hz AC to DC switch mode power supply capable of delivering 24 volts at 3 75 amperes is supplied with your system If it is necessary to convert the unit for different voltage operat...

Страница 11: ...electrical outlet 3 If the optional foot switch is used insert the plug into the jack located on the rear panel Do not insert footswitch plug while the system is powered on 4 Probes must be properly t...

Страница 12: ...mode 1 Ensure that the amplitude control knob is set to 0 2 Set the switch on the front panel to thumb switch 3 If needed connect the foot switch cable to the designated jack on the back of the power...

Страница 13: ...NTENANCE Ultrasonic homogenizers create high intensity vibration which puts stress on the con verter and horn assembly The sides and end of the probe must never be allowed to come in contact with anyt...

Страница 14: ...debris on the threading of the connecting stud Allow the alcohol to dry completely 3 Clean threading in converter Use alcohol and a cotton swab to remove any debris on the threading Do not allow liqu...

Страница 15: ...turn the power on for a few seconds to remove residue The tip also can be sterilized using alco hol with the power on 9 TROUBLESHOOTING The most probable causes for malfunction are listed below and sh...

Страница 16: ...volumes are application specific Tip Diameter Processing Volume Range 5 64 1 6 mm 200 L to 5 mL 1 8 3 2 mm 500 L to 15 mL 1 4 6 4 mm 10 mL to 50 mL Vessel shape and size A narrow vessel is preferable...

Страница 17: ...low amplitude setting you will deliver low intensity sonication If you operate at a high amplitude set ting you will have high intensity sonication In order to be able to reproduce results the amplit...

Страница 18: ...935 Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com PN 03 2021 55 Rev A...