Bead Ruptor
TM
Cryo Cooling Unit
User Manual
Страница 1: ...Bead RuptorTM Cryo Cooling Unit User Manual...
Страница 2: ...n those specified on the plate affixed to the instrument misuse or abuse OMNI International a PerkinElmer company reserves the right to change alter modify or improve any of its instruments without an...
Страница 3: ...g Unit 1 Insulated Cryo Hose 1 Bead RuptorTM Cryo Lid 1 External Thermometer 1 Compressed Air Hose 1 Tool Kit 1 User Manual 1 Not Included Hose Barb 3 8 or equivalent 1 1 Lid 2 Airflow Control Knob 3...
Страница 4: ...nizer Lid Locate Lid Hinges 1 2 Lift Lid and remove 4 screws using the provided Allen Key Remove Lid Replace with Bead RuptorTM Cryo Lid Black 3 4 Replace the four screws and tighten using the provide...
Страница 5: ...re under the Bead Ruptor lid Place remaining Couple Clips as shown Guide the thermometer probe wire through the clips Plug the thermometer probe wire into the External Thermometer NOTE The Digital The...
Страница 6: ...n an air compressor and the Bead RuptorTM Cryo Unit inlet use a refrigerated air dryer or desiccant dryer These systems are specified in the lowest dew point that can be reached DO NOT use inlet press...
Страница 7: ...ing dry air for 30 seconds at the MAX setting Turn knob to the Off position Close the lid of the Bead RuptorTM Cryo unit and turn the air flow control knob counterclockwise to begin the flow of air in...
Страница 8: ...ecommended air pressure for use with the Bead RuptorTM Cryo Pressure above 120 PSI could damage internal components Close the Bead RuptorTM Homogenizer lid Load samples into the Bead RuptorTM Homogeni...
Страница 9: ...ounter clockwise to begin cooling Allow air to flow until the External Thermometer reads 0oC Once the temperature has reached 0oC press RUN on the Bead Ruptor to begin homogenization When the cycle ha...
Страница 10: ...MAX setting Turn knob to the Off position Close the lid of the Bead RuptorTM Cryo unit and turn the air flow control knob counterclockwise to begin the flow of air into the processing chamber Pre coo...
Страница 11: ...chamber volume as leaking may occur CAUTION 55 120 PSI is the recommended air pressure for use with the BeadRuptorTM Cryo Pressureabove120PSIcoulddamageinternalcomponents Close the Bead RuptorTM Homo...
Страница 12: ...low air to flow until the External Thermometer reads 5oC 23oF Once the temperature has reached 5oC press RUN on the Bead RuptorTM Homogenizer to begin homogenization When the cycle has ended Turn off...
Страница 13: ...The air purity is lower than recommended 1 Wait for the units to completely de frost 2 Ensure the air purity is 99 and the water content 5ppm One or several outflows do not work One or more nozzles ar...
Страница 14: ...iquid nitrogen DO NOT use compressed gases which are not specified DO NOT operate the Bead RuptorTM Cryo cooling unit with the Bead RuptorTM Homogenizer 15 mL 30 mL or 50 mL tube carriages Damage to t...
Страница 15: ...This page was intentionally left blank...
Страница 16: ...PN 19 8026 Rev H 935 Cobb Place Blvd NW Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com...