CPC 100 V1.41
16 - 28
CPC 100
Combined Specifications
Resistance Measurement
The accuracy of the combined measurements can be calculated from the
respective input and output specifications.
The tables below give some examples
4-wire measurement with 400A DC output and 10V VDC input
Current
Resistance
Voltage
Typ. error
Guaranteed
400A
10µ
Ω
4mV
0.70%
1.35%
400A
100µ
Ω
40mV
0.55%
1.10%
400A
1m
Ω
400mV
0.50%
0.95%
400A
10m
Ω
4V
0.50%
0.95%
4-wire measurement with 6A DC output and 10V VDC input
Current
Resistance
Voltage
Typ. error
Guaranteed
6A
100m
Ω
0.6V
0.35%
0.60%
6A
1
Ω
6V
0.35%
0.60%
1A
10
Ω
10V
0.25%
0.40%
2-wire measurement with 10V VDC input
Current
Resistance
Voltage
Typ. error
Guaranteed
< 5mA
100
Ω
0.60%
1.20%
< 5mA
1k
Ω
0.51%
1.02%
< 5mA
10k
Ω
0.50%
1.00%
Содержание CPC 100
Страница 12: ...CPC 100 V1 41 x...
Страница 28: ...CPC 100 V1 41 1 16...
Страница 90: ...CPC 100 V1 41 3 14...
Страница 194: ...CPC 100 V1 41 6 30...
Страница 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Страница 258: ...CPC 100 V1 41 9 4...
Страница 264: ...CPC 100 V1 41 10 6...
Страница 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Страница 284: ...CPC 100 V1 41 12 12...
Страница 312: ...CPC 100 V1 41 14 12...
Страница 316: ...CPC 100 V1 41 15 4...
Страница 350: ...CPC 100 V1 41 16 34...
Страница 370: ...CPC 100 V1 41 17 20 7 Finish the assembly by screwing the backnut to the housing and by tightening them with wrenches...
Страница 372: ...OMICRON Contact Addresses 22...