CPC 100 V1.41
5 - 24
Automatic Test vs. Manual Test
1. Option "Manual input" clear
Automatic test
– Press the I/O (test start/stop) push-button to start the test.
The test voltage increases from 0V to Vtest in a ramp characteristic within
1 second. Then Vtest is kept for a period of 2.4 seconds, and the
measurements are taken. Next, the ramp decreases again within
0.3 seconds.
– The automatic test stops by itself, it can now be assessed.
– To repeat the test, first press B
A C K
T O
T
O P
, if applicable, and then C
L E A R
R
E S U L T S
.
Manual test
Clearing "Auto" lets you set the test voltage Vtest manually with the handwheel.
– Press the I/O (test start/stop) push-button to start the test. The focus is set
onto the "V test:" entry field.
– Set the value of your choice either by turning the handwheel and/or by
entering it with the numerical keys.
– Now the measurements are taken and displayed.
– To save the measurement press K
E E P
R
E S U L T S
(pressing the I/O push-
button has the same effect). The last measured values are "frozen", the test
can now be assessed.
– Stop the manual test by pressing the I/O (test start/stop).
– To repeat the test, first press B
A C K
T O
T
O P
, if applicable, and then C
L E A R
R
E S U L T S
.
Note:
A value set with the numerical keys needs to be fully entered and
acknowledged by pressing either E
N T E R
or the handwheel before
it is applied to the output. If needed, it can then be adjusted using
the handwheel.
Values set by turning the handwheel are applied instantaneously.
Содержание CPC 100
Страница 12: ...CPC 100 V1 41 x...
Страница 28: ...CPC 100 V1 41 1 16...
Страница 90: ...CPC 100 V1 41 3 14...
Страница 194: ...CPC 100 V1 41 6 30...
Страница 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Страница 258: ...CPC 100 V1 41 9 4...
Страница 264: ...CPC 100 V1 41 10 6...
Страница 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Страница 284: ...CPC 100 V1 41 12 12...
Страница 312: ...CPC 100 V1 41 14 12...
Страница 316: ...CPC 100 V1 41 15 4...
Страница 350: ...CPC 100 V1 41 16 34...
Страница 370: ...CPC 100 V1 41 17 20 7 Finish the assembly by screwing the backnut to the housing and by tightening them with wrenches...