Hardware overview

3.6

Emergency Stop button

Figure 3-5: Emergency Stop button

Pressing the Emergency Stop button 

immediately

 shuts off all current and voltage 

outputs except for the 

Ext. Booster

 output.

A running test is terminated, the software does not accept any more entries and/or 
commands.

Once the reason for the Emergency Stop is cleared and the Emergency Stop button 
released, the test can be re-started by pressing the 

I/O

 (test start/stop) push-button 

while in Test Card View.

DANGER

Death or severe injury caused by high voltage or current

The 

Ext. Booster 

connector is 

always

 galvanically connected to mains. 

This also applies when no external booster is selected, the green status light (0) is on, 
the outputs are turned off or the Emergency Stop button is pressed.

► Handle the 

Ext. Booster

 connector with extreme caution.

► Do not use any other booster cables than the ones supplied by OMICRON.
► Do not use booster cables that are frayed or damaged in any way.

WARNING

Death or severe injury caused by high voltage or current possible

Stopping a test does not shut off the

 CPC 80

 outputs instantaneously. First, the currently 

running test sequence finishes, then the test execution is stopped. 
Most test cards finish the running test sequence with a predefined ramp function.

► Therefore, in a hazardous situation never press "Stop Test".
► Instead, use Emergency Stop.

Содержание CPC 100

Страница 1: ...CPC Getting Started...

Страница 2: ...ta embodied in this manual represent the technical status at the time of writing and are subject to change without prior notice We have done our best to ensure that the information given in this manua...

Страница 3: ...d the regulations on site 1 2 Safety standards and rules 1 2 1 Safety standards Testing with the CPC must comply with the internal safety instructions and additional safety relevant documents In addit...

Страница 4: ...yee of the utility Turn off and disconnect the high voltage from the test object Protect yourself and your working environment against an accidental re connection of high voltage by other persons and...

Страница 5: ...o the grounding system on site This also applies to all other test devices and accessories in the test setup Make sure that the grounding clamp has a good electrical contact to the grounding system on...

Страница 6: ...the equipotential ground or PE of the generator has to be grounded properly If this is not possible measurements are not permitted and cannot be performed Additional information Instead of supplying t...

Страница 7: ...e to be connected to the CPC 100 do not carry any voltage potential During a test the only power source for a test Make sure that the ground terminal of the device under test is in good condition clea...

Страница 8: ...t amount of heat approx 300 W m at 800 A After operation wait for cables and clamps to cool down before touching them If in doubt wear gloves Never use the test cards Quick and Resistance to measure t...

Страница 9: ...nues the monitoring of the discharge status before the test object might be completely discharged The CPC displays an error message and a green light in this case This could erroneously be interpreted...

Страница 10: ...struction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be require...

Страница 11: ...100 hardware concept In contrast to the CPC 100 the CPC 80 only comes with the booster output It is not equipped with any measurement inputs or directly usable outputs and therefore intended for use w...

Страница 12: ...anel control Fuses for 6A AC and 6A DC outputs Navigation elements Test Start Stop Soft touch keyboard Key lock BIN IN Binary trigger input Emergency stop button I O Status lights INPUT Measuring inpu...

Страница 13: ...Figure 3 4 Left side view of the CPC 80 ePC with front panel control Key lock I O Status lights Soft touch keyboard Navigation elements Test Start Stop Emergency stop button External booster Groundin...

Страница 14: ...mains Table 3 1 Status lights 1 Output is switched OFF Green status light 0 on Current voltage source is inactive off 2 Output is switched ON and or measurement process is active Red status light I is...

Страница 15: ...r is always galvanically connected to mains This also applies when no external booster is selected the green status light 0 is on the outputs are turned off or the Emergency Stop button is pressed Han...

Страница 16: ...00 and CPC 80 USB connector Connector for OMICRON safety accessories RJ 45 socket for connection to a PC or a network hub Yellow LED Data are being transferred from to the network Green LED CPC connec...

Страница 17: ...test object according to your requirements 5 Set up a barrier to separate the high voltage area from the work area see Figure 1 1 Separation of work area and high voltage test area on page 9 6 Remove...

Страница 18: ...for the first time a warning message appears on the screen Only after the I O test start stop push button is pressed a second time the voltage is applied to the CPC output An error message 313 appears...

Страница 19: ...the CPC and the test object 5 Remove the barrier 4 2 Cleaning Use a cloth dampened with isopropanol alcohol to clean CPC and its accessories WARNING Death or severe injury caused by high voltage or c...

Страница 20: ...CPC Getting Started 20 OMICRON...

Страница 21: ...3 59495 4444 Additionally you can find our Service Center or Sales Partner closest to you at www omicronenergy com Customer Portal stay informed www omicronenergy com customer The Customer Portal on o...

Страница 22: ...ENU 1194 03 04...

Отзывы: