Omicron COMPANO 100 Скачать руководство пользователя страница 20

Power up COMPANO 100

Press the 

On/Off

 key to power up 

COMPANO 100

. Grant the test set about 30 s to boot up.

The test set starts in QUICK (→ chapter 

QUICK

 in User Manual).

Power down COMPANO 100

While 

COMPANO 100

 is running, press the 

On/Off

 key again to turn off the test set.

COMPANO 100

 does not shut off instantly, but performs a controlled shutdown instead.

To abort the controlled shutdown, press any other key of your choice at the front panel control.

A system shutdown is also initiated by a battery running low, shortly before a looming break-down of
the power supply.

6.1 Testing a relay in QUICK

1. When 

COMPANO 100

 is powered up, turn the jog dial wheel until the focus is on the output

current field.

2. Press the jog dial wheel, set the current to a value of your choice (that is, a value that will make

the relay trip for sure), then press the jog dial wheel again.

3. By default, the timer is configured to start when you press the 

Start/Stop

 key, and to end when

there is a change at the dry contact connected to 

IN 1

. Therefore, for many applications it may not

be necessary to change the timer from its default.

4. Press the 

Start

/

Stop

 key. The current is issued at the current output.

5. The current output is turned off by default as soon as the binary contact of the relay switches.

6. The timer records the duration of the relay trip time, that is, the time it took the relay to trip starting

from the moment of the current injection.

Getting Started with COMPANO 100

20

OMICRON

Содержание COMPANO 100

Страница 1: ...COMPANO 100 Getting Started...

Страница 2: ...eprinting wholly or in part is not permitted The product information specifications and technical data embodied in this manual represent the technical status at the time of writing and are subject to...

Страница 3: ...O 100 test set refer to the chapters Safety instructions Battery Test set overview and Product startup These chapters are identical in Getting Started manual and User Manual Stay informed Our Customer...

Страница 4: ...or professional specialists in electronics and electrical engineering Its purpose is to familiarize you with the COMPANO 100 test set and its various application fields It contains helpful instruction...

Страница 5: ...quirements from national standards for accident prevention and environmental protection Reading the COMPANO 100 manual alone does not release you from the duty of complying with all national and inter...

Страница 6: ...Getting Started with COMPANO 100 6 OMICRON...

Страница 7: ...sconnect completely 2 Secure against re connection 3 Verify that the installation is dead 4 Carry out grounding and short circuiting 5 Provide protection against adjacent live parts 3 1 Operator quali...

Страница 8: ...ages Pay attention to the safety information provided in the documentation The mains cable cable to the OMICRON supplied battery charger must be rated for the nominal voltage and current specified in...

Страница 9: ...rea anymore now image above 6 Start the test at COMPANO 100 While testing do not enter the dangerous area Stay clear from the cable ends at the device under test Quite often these devices have connect...

Страница 10: ...rea are powerless before you work on touchable test objects connections or terminals Wiring and Safe Connection on page 9 The COMPANO 100 test set is supplied with flexible test lead adapters 2 x blac...

Страница 11: ...page 18 Ground COMPANO 100 as close as possible to the operator COMPANO 100 is of protection class II Therefore grounding is not as critical as with devices that are grounded via a power cord Neverth...

Страница 12: ...ing The test set and all of its accessories are not intended for household use At the end of its service life do not dispose of the test set with household waste For customers in EU countries incl Eur...

Страница 13: ...g the battery charging is terminated automatically when the temperature exceeds that range You can check the battery status and temperature at Setup Battery Info If the battery is fully charged the ch...

Страница 14: ...ote that these regulations change frequently Double check the current status whenever applicable or if in doubt contact OMICRON s Technical Support Support The signing the IATA Shipper s Declaration m...

Страница 15: ...umidity environment To avoid deep discharge of the battery recharge it about every 6 months Note that higher temperatures result in a drastically higher self discharge rate of the battery Do not store...

Страница 16: ...ns over time The SOH is updated with every full discharge charge cycle To enforce an update1 perform as follows 1 Read out the state of health SOH 2 Disconnect the charger 3 Output 10 A to a short cir...

Страница 17: ...Death or severe injury caused by high voltage possible Inductive loads may contain a lethal amount of energy if charged with current The amount of energy depends on the size of the inductive load the...

Страница 18: ...the picture above the connector for external safety functions to protect COMPANO 100 from unauthorized usage WARNING Death or severe injury can occur if the appropriate safety instructions are not ob...

Страница 19: ...the cord it will pull out of the test set s socket without damaging the connector or the socket Then connect COMPANO 100 to the device you would like to test For safety reasons press the emergency st...

Страница 20: ...the jog dial wheel until the focus is on the output current field 2 Press the jog dial wheel set the current to a value of your choice that is a value that will make the relay trip for sure then press...

Страница 21: ...ock 30 g 11 ms half sine 3 shocks in each axis Vibration 5 g RMS frequency range 10 2 kHz 30min in each axis Size weight Weight 9 8 kg 21 6 lb case without protection cover Dimensions w h d 360 312 21...

Страница 22: ...Getting Started with COMPANO 100 22 OMICRON...

Страница 23: ...y you can find the OMICRON Service Center or OMICRON Sales Partner closest to you at www omicronenergy com Contact Customer Portal Stay Informed www omicronenergy com customer The Customer Portal on o...

Страница 24: ...ENU 1122 03 01...

Отзывы: