3.4.2 Safe use of test leads and adapters
Test leads
WARNING
Death or severe injury caused by high voltage possible
The outputs of the
COMPANO 100
are controlled by the software running in the device.
Note that just turning off the outputs in the controlling software is not sufficiently safe.
► Do not work on connected touchable test objects, connections or teminals while the
COMPANO 100
is active (Start/stop pressed) .
► Always make sure that the
Emergency stop
button is pressed, the test set is
secured against reactivation, and all parts in the working area are powerless before
you work on touchable test objects, connections or terminals (→
The
COMPANO 100
test set is supplied with flexible test lead adapters
(2 x black, 2 x red) that have an angled connector at one end, and a
retractable sleeve at the other end.
Never directly insert one of the retractable sleeves into a
COMPANO 100
socket at the front of the test set. This does not
comply with the designated purpose of these leads and is contrary to
the safety regulations.
Retractable sleeve
Angled connector
To connect a test object to
COMPANO 100
,
always use the angled connectors for the
test set's sockets:
At the test object, use the test lead's
retractable sleeve side: