6.2 Configuring network settings and accessing Web Interface via
Ethernet
1. Locate the
CMC 850
test set in
OMICRON Device Link
2. Associate the
CMC 850
test set with your computer.
a) In the device list, right-click the
CMC 850
entry to open a device-related menu.
b) Click
Associate device
and follow the on-screen instructions.
3. Access the embedded functions via the Web Interface
CMC 850
’s embedded functions are accessible via its Web Interface. Any common Web browser
can be used to access
CMC 850
and to make use of the embedded functions.
Alternative method to invoking the Web browser by the procedures described above:
a) Start a Web browser of your choice on your computer.
b) Enter
CMC 850
’s IP address (in a format like http://172.22.23.61) into the Web browser’s
address bar.
c) Hit the <Return> key.
6.3 Using a
CMIRIG-B
interface unit
To use the OMICRON-supplied
CMIRIG-B
interface unit as reference for the time synchronization of
the IEC 61850 bus system, connect the
CMIRIG-B
’s front side connector “CMC” to the
CMC 850
‘s
“external Interface” connector.