5.5 LL out 1 - 6 and LL out 7 - 12 (Low Level Outputs)
The interface connectors “LL out 1 - 6” and “LL out 7 - 12” hold a total of 12 independent
high accuracy analog signal outputs. These
Test Universe
-controlled signal sources can
serve to either drive external amplifiers or to directly provide small signal outputs.
The low level outputs are short-circuit-proof and continually monitored for overload. They
deliver calibrated signals in the range from 0 ... 7 V
eff
nominal (0 ... ± 10 V).
For technical data, → “Low Level Outputs “LL out 1 - 6” and “LL out 7 - 12”” on page 18.
Overload Warning in the Test Software
When a low level output is overloaded, the test software issues an appropriate warning. Refer to the
Test Universe
or
RelaySimTest
Help for more details.
Pin assignment
“LL out 1-3”, “LL out 4-6”, “LL out 7-9” and “LL out 10-12” each make up a selectable voltage or current
triple.
LL out 1 - 6
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
Pin
Function LL out 1-6
Function LL out 7-12
Pin 1
LL out 1
LL out 7
Pin 2
LL out 2
LL out 8
Pin 3
LL out 3
LL out 9
Pin 4
Neutral (N) connected to GND
Pin 5
LL out 4
LL out 10
Pin 6
LL out 5
LL out 11
Pin 7
LL out 6
LL out 12
Pin 8-16
For internal purposes
Housing
Screen connection