3.5.4 Output power
Typical
Guaranteed
AC output current
1
per channel
total
280 mA
1400 mA max.
250 mA
-
DC output current
per channel
total
400 mA
1700 mA max.
350 mA
-
Supply voltage range (DC)
-
75 V ... 225 V
AC output power
1
3 x 42 W @ 150 V
6 x 42 W @ 150 V
1 x 84 W @ 300 V
3 x 37.5 W @ 150 V
6 x 37.5 W @ 150 V
1 x 75 W @ 300 V
1. Tested for resistive load and symmetric conditions (0°, 120°, 240°) at 50/60 Hz.
Continuous operation
By adjusting the supply voltage to the lowest possible setting, it is possible to minimize the power
dissipation in the amplifier in order to increase the possible continuous output power. Continuous
operation means that the
CMC 430
is able to provide a specified voltage at a specified load for at least
30 minutes.
0
10
20
30
40
50
0
25
50
75
100
125
150
Output
power
per phase
(W)
Output voltage (V)
0
10
20
30
40
150
0
20
40
60
80
100 120 140
Output
power
per phase
(W)
Output voltage (V)
+23 °C
+50 °C
Typical output power of the
CMC 430
Typical continuous output power, possible with
optimized supply voltage settings
Technical data