Technical Data
51
6.3.4 Operational
Limits
in
Conjunction with Mains Supply
Principally, the maximum output power of the
CMC 356
is limited by the
mains input supply voltage.
For mains voltages of 115 V
AC
or smaller, it is also possible to supply the
CMC 356
with two phases (L-L) instead of the normal phase-neutral (L-N)
operation in order to increase the supply voltage (115 V * sqrt(3) = 200 V).
In order to limit the internal losses and to maximize the output power of the
voltage amplifier, always set the maximum test object voltage to the
minimum value possible for the test.
Beside the reduction of the available total output power of low line voltages,
no other significant degradations in the technical data of the
CMC 356
occur.
Table 6-8:
Typical total output power
at different mains voltages.
Mains
voltage
Current
Typical total output power
Currents only Currents AUX DC & voltage
230V 6 x 15 A
1600 W
1190 W + 300 W
6 x 25 A
1470 W
1060 W + 300 W
6 x 32 A
1320 W
910 W + 300 W
115V
1
1
After 15 min of continuous operation at full output power a duty cycle of 15 min
on/15 min off is required at an ambient temperature of 25°C. This does not apply to
the 6 x 32 A example because the output duration is limited by the current amplifier
(see Chapter 6.3.2, "Current Outputs" on page 43 for more details).
6 x 15 A
1120 W
710 W + 300 W
6 x 25 A
990 W
580 W + 300 W
6 x 32 A
860 W
450 W + 300 W
100V
1
6 x 15 A
910 W
500 W + 300 W
6 x 25 A
790 W
380 W + 300 W
6 x 32 A
670 W
260 W + 300 W
Содержание CMC 356
Страница 1: ...CMC 356 User Manual...
Страница 38: ...CMC 356 Reference Manual 38...
Страница 100: ...CMC 356 Reference Manual 100...
Страница 104: ...CMC 356 Reference Manual 104...
Страница 132: ...CMC 356 Reference Manual 132...
Страница 136: ...The OMICRON Bootloader Software Copyright Notices 136...
Страница 138: ...Contact Information Technical Support 138...