CMC 356 Reference Manual
78
6.11.6.1 Accuracy
of
AC
Measurements
Conditions: integration time 1 s, measurement signal sinusoidal, excitation
10 - 100 %, accuracy references the measurement full scale values.
Table 6-33:
Sampling rate 28.44 kHz;
measurement range
600 V, 100 V, 10 V, 1 V
Table 6-34:
Sampling rate 28.44 kHz;
measurement range
100 mV
Table 6-35:
Sampling rate
9.48 kHz
3.16 kHz;
measurement range
600 V, 100 V, 10 V, 1 V
Table 6-36:
Sampling rate
9.48 kHz
3.16 kHz;
measurement range
100 mV
Frequency range
Accuracy
Typical
Guaranteed
DC
± 0.15% ± 0.40%
10 Hz ... 100 Hz
± 0.06%
± 0.15%
10 Hz ... 1 kHz
+ 0.06% / -0.11%
± 0.25%
10 Hz ... 10 kHz
+ 0.06% / -0.7%
± 1.1%
Frequency range
Accuracy
Typical
Guaranteed
DC
± 0.15% ± 0.45%
10 Hz ... 100 Hz
± 0.1%
± 0.3%
10 Hz ... 1 kHz
+ 0.15% / -0.2%
± 0.5%
10 Hz ... 10 kHz
+ 0.15% / -1.0%
± 2%
Frequency range
Accuracy
Typical
Guaranteed
DC
± 0.15% ± 0.45%
10 Hz ... 100 Hz
± 0.08%
± 0.2%
10 Hz ... 1 kHz
+ 0.1% / -0.3%
± 0.5%
10 Hz ... 4 kHz
(sampling rate 9.48 kHz)
+ 0.1% / -0.5%
± 1.2%
10 Hz ... 1.4 kHz
(sampling rate 3.16 kHz)
+ 0.1% / -0.5%
± 1.0%
Frequency range
Accuracy
Typical
Guaranteed
DC
± 0.15% ± 0.5%
10 Hz ... 100 Hz
± 0.1%
± 0.35%
10 Hz ... 1 kHz
+ 0.15% / -0.35%
± 0.5%
10 Hz ... 4 kHz
(sampling rate 9.48 kHz)
+ 0.15% / -0.6%
± 1.2%
10 Hz ... 1.4 kHz
(sampling rate 3.16 kHz)
+ 0.15%/ -0.6%
± 1.2%