CMC 310 User Manual
28
OMICRON
Circuit diagram of
ext. Interf.
binary transistor outputs 11–14:
For a manufacturer description about the connection sockets
LL out
and the external interface
ext. Interf.
, visit the website
www.lemo.com.
You can order the LEMO cable directly from OMICRON.
4.4.5
Binary output relays
Manufacturer ordering information
Connector for one-guide notch and pull relief (for
ext. Interf.
).
FGG.2B.316.CLAD 72Z
Black anti-bend cable cover
GMA.2B.070 DN
4 binary output relays (BINARY OUTPUT 1–4)
Type
Potential-free contacts; software-controlled
Connection
4 mm sockets
AC loading capacity
AC breaking capacity
V
max
= 300 V, I
max
= 8 A, P
max
= 2000 VA
DC loading capacity
DC breaking capacity
→ "Load limit breaking capacity curve for binary output relays with
DC voltages" on page 29.
Inrush current
15 A (max. 4 s at 10 % duty cycle)
Carry capacity
5 A continuous at 60 °C (140 °F)
Electrical lifetime
100 000 switching cycles at 230 V
AC
/8 A and ohmic load
Operate time
Max. 10 ms (no bouncing)
Release time
Max. 5 ms (no bouncing)
Overvoltage category
II, according to IEC 61010-1
Rear side of
CMC 310
R
pullup
Inside of
CMC 310
16 V
V
extern
= 5 ... 15 V
Binary outputs 11–14
ext. Interf.