SPECTANO 100 User Manual
2.2
Five safety rules
Before connecting any test leads to a test object, always observe the five safety rules for working with high
power:
►
Disconnect completely.
►
Secure against re-connection.
►
Verify that the installation is dead.
►
Carry out grounding and short-circuiting.
►
Provide protection against adjacent live parts.
2.3
Operating the measurement setup
The safety instructions given here always apply when operating the
SPECTANO 100
. They are
supplemented by further notes and warnings applicable for specific actions only. Such specific
notes and warnings are given where necessary in this user manual.
In addition to the safety instructions given in this manual, always observe the applicable
company internal safety instructions and safety documents.
In case of any problems or questions, or if you do not understand some safety instructions,
contact the OMICRON Lab Technical Support (refer to chapter “Technical support”).
►
Read the safety instructions and setup information described in this User Manual before
operating
SPECTANO 100
or its accessories. In addition, read the manuals of all devices
included in the test setup. Any maloperation can result in damage to property or persons.
►
Use
SPECTANO 100
and its accessories only in a technically sound condition and when its use
is in accordance with the safety regulations for the specific job site and application.
►
Before using
SPECTANO 100
, always ensure that neither the device nor its accessories are
damaged.
►
Connect the equipotential ground of
SPECTANO 100
with a solid connection of at least 6 mm
2
cross-section and not longer than 20 meters to a ground connection in the laboratory. Never
handle
SPECTANO 100
without grounding it.
►
Ensure that all ground connections are in good condition, clean and free of oxidation.
►
Connect all parts of the test cell which have to be connected to ground according to the related
manual.
►
Only use the delivered grounded AC power supply DRA SUPPLY to power
SPECTANO 100
.
Never use power outlets without protective ground (PE).
►
Always turn off
SPECTANO 100
(with the power switch on the back) before connecting or
disconnecting cables or changing the test cell. Be especially careful when using unsafe adapter
cables (e.g. BNC adapter).
►
Ensure that no dangerous voltages are present at touchable parts of the test setup. If this is not
possible due to, for example, unsafe adapter cables or connectors or unsafe test cells always,
1. Secure the danger zone with clearly visible barriers before performing tests.
2. Prevent the presence of other persons in the potentially dangerous area around the test
object