30
SPECTANO 100 User Manual
Inner conductor
Electrodes
Inner conductor
Barrier
*
same potential
Outer (ground) &
inner (guard) shield
*
To connect the adapter cable to
SPECTANO 100
:
1. Connect the standard triaxial cable according to the color code to the
SPECTANO 100
OUTPUT
and
INPUT
connectors.
2. Connect the adapter cables with the corresponding color code to the standard triaxial
cable.
9.4.1
Typical setup for a test cell with guard ring
Guard ring
Figure 9-4
Connecting
SPECTANO 100
to a test cell with guard ring
Death or severe injury caused by high voltage or current
►
Do not touch the
OUTPUT
connector when the red LED is on or blinks.
►
Always ensure that no dangerous voltages are present at any touchable
part of the device, test cell or sample.
►
Do not touch the output if the guard (inner shield of the triaxial connection)
is identical with the ground (outer shield of the triaxial connection).
The safe area is only safe if the delivered cables and adapters are used.
Otherwise, this can also be a danger zone.
DANGER
NOTICE
Outer (ground) &
inner (guard) shield
*