![Omicron Lab SPECTANO 100 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/omicron-lab/spectano-100/spectano-100_user-manual_3978180025.webp)
24
SPECTANO 100 User Manual
Figure 8-7 Measurement circuit, guarded electrode (C
s
is stray capacitance)
8.3
Stable environment
The permittivity and dielectric losses vary with temperature and humidity. Therefore, we recommend
measuring the sample in a stable environment.
8.4
Shielding of low signal measurements
To improve the signal-to-noise ratio for low current measurements, do the following improvements on the
test setup:
►
Put the test cell and/or test device into an EMC box to shield the test system from external noise.
►
Increase the test voltage.
►
Separate guard and ground.
►
Ensure the cabling is also shielded.
For more information on sample dimension or test setup improvement, see our application notes